# K033801: Kyphx HV-R, Model C01A

- Pathway: 510(k) (Traditional)
- Applicant: Kyphon, Inc.
- Decided: Apr 1, 2004 after 115 days
- Product code: NDN (Cement, Bone, Vertebroplasty)
- Device class: 2
- Regulation: 21 CFR 888.3027
- Recalls of this device: 2 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 1 generation back to K002652 (2001). 1 ancestor in all.
212 later devices descend from it; 6 cite it directly.
0 of 214 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K002652: Surgical Simplex P Radiopaque Bone Cement and Acm and Mixevacii, Stryker Corp. (2001) Summary page 1.

## Cited in the summary but kept off the family tree
- K0017004: Howmedica Osteonics Surgical Simplex P Radiopaque Bone Cement. Number could not be read

## Most influential descendants
- K041584: Kyphx HV-R Bone Cement, Model C01A (2004), 209 descendants of its own
- K042691: Vertefix Vertebroplasty Procedure Set (2005), 189 descendants of its own
- K060300: Confidence High Viscosity Bone Cement (2006), 176 descendants of its own
- K093828: Kyph X HV-R Bone Cement (2010), 167 descendants of its own
- K102397: Kyphon Xpede Bone Cement (2011), 165 descendants of its own

## Recalls of this device
- Nov 10, 2009 · Not classified · Z-1625-2011: Product may not reach proper viscosity (dough state) and inconsistency or lack of homogeneity during cement preparation and during delivery period.
- Nov 10, 2009 · Not classified · Z-1624-2011: Product may not reach proper viscosity (dough state) and inconsistency or lack of homogeneity during cement preparation and during delivery period.

## Indications for use
The KyphX® HV-R Bone Cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis using a kyphoplasty procedure.

Source page: https://510kreport.com/device/K033801