# K040133: Urine Amphetamine/Methamphetamine Screen Flex Reagent Cartridge

- Pathway: 510(k) (Traditional)
- Applicant: Dade Behring, Inc.
- Decided: Apr 8, 2004 after 78 days
- Product code: DKZ (Enzyme Immunoassay, Amphetamine)
- Device class: 2
- Regulation: 21 CFR 862.3100
- Recalls of this device: none
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 2 generations back to K993982 (2000). 2 ancestors in all.
2 later devices descend from it; 1 cites it directly.
0 of 5 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K031004: Syva Emit II Plus Amphetamines Assay, Dade Behring, Inc. (2003) Summary page 2.

## Most influential descendants
- K062128: The Dimension Vista Flex Reagent Cartridges / Kit Assays (2006), 1 descendant of its own
- K102045: Dimension Visa HBA1C, Model K3105A (2011), 0 descendants of its own

## Indications for use
The AMPH Flexc reagent cartridge used on the Dimension 0 clinical chemistry system provides reagents for an in viiro diagnostic test intended for the qualitative and semi-quantitative determination of amphetamines in human urine using a cutoff of either 300, 500 or 10OOng/mL. Measurements obtained with the AMPH method are used in the diagnosis and treatment of amphetamines use or overdose. The AMPIH method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.