510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Roche Diagnostics Tina-Quant Iga Gen.2

K040435510(k) SpecialCleared Mar 10, 200420-day reviewClass II21 CFR 866.5510Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Diagnostics Corp. · CZP Iga, Antigen, Antiserum, Control

Cleared on Mar 10, 2004 as substantially equivalent to K955907 (Boehringer Mannheim Corp., 1996) after a 20-day review. Its family tree reaches back 1 generation to 1996. 4 later devices descend from it. It has been recalled once.

Decided
Mar 10, 2004
Review
20 days
Class
Class II
Product code
CZP
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to K955907 (1996). 4 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2012: 1 descendant cleared2019: 2 descendants cleared2025: 1 descendant clearedK955907 · Boehringer Mannheim Iga Assay · Boehringer Mannheim Corp. · 19961995200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 2 descendants3 generations forward: 1 descendant K955907 · Boehringer Mannheim Iga Assay · Boehringer Mannheim Corp. · 19960
  1. Root · 1 generation backK955907 Boehringer Mannheim Iga AssayBoehringer Mannheim Corp. · 1996
  2. This deviceK040435 Roche Diagnostics Tina-Quant Iga Gen.2Roche Diagnostics Corp. · 2004
  3. Descendant · cites it directly · 3 of its ownK103824 Human Iga Liquid Reagent Kit for Use on Spa PlusThe Binding Site · 2012
  4. Descendant · 2 generations on · 1 of its ownK191985 Optilite IgA KitThe Binding Site Group, Ltd. · 2019
  5. Descendant · 3 generations on · 0 of its ownK250159 Immunoglobulin Isotypes (GAM) for the EXENT Analyser; EXENT AnalyserThe Binding Site Group, Ltd. · 2025 · Recalled
  6. Descendant · 2 generations on · 0 of its ownK192116 Human IgA liquid reagent kit for Use on SPAPlusThe Binding Site Group, Ltd. · 2019

Most influential descendants

Showing 4 of 4 · See all
  • K103824Human Iga Liquid Reagent Kit for Use on Spa PlusThe Binding Site · 2012 · cites it directly3descendants
  • K191985Optilite IgA KitThe Binding Site Group, Ltd. · 2019 · 2 generations on1descendant
  • K250159Immunoglobulin Isotypes (GAM) for the EXENT Analyser; EXENT AnalyserThe Binding Site Group, Ltd. · 2025 · 3 generations on0descendants
  • K192116Human IgA liquid reagent kit for Use on SPAPlusThe Binding Site Group, Ltd. · 2019 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K955907Boehringer Mannheim Iga AssayBoehringer Mannheim Corp.1996—
0 (this device)K040435Roche Diagnostics Tina-Quant Iga Gen.2Roche Diagnostics Corp.2004—
+1K103824Human Iga Liquid Reagent Kit for Use on Spa PlusThe Binding Site2012—
+2K191985Optilite IgA KitThe Binding Site Group, Ltd.2019—
+3K250159Immunoglobulin Isotypes (GAM) for the EXENT Analyser; EXENT AnalyserThe Binding Site Group, Ltd.20251
+2K192116Human IgA liquid reagent kit for Use on SPAPlusThe Binding Site Group, Ltd.2019—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201920
202510
FIG. 1 — Lineage of K040435, 1996–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CZP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CZP device

  1. #1 K193525 Yumizen C1200 Immunoglobulin A, Yumizen C1200…score 0.2 · 3 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K955907Boehringer Mannheim Iga Assay
Summary page 1
…stry analyzers. Predicate We claim subs~tantial equivalence to the currently marketed Roche Device Diagnostics Tina-quant IgA assay. (K95 5907). 22 510(k) Summary -…
Boehringer Mannheim Corp.1996—

Also cleared against K955907: 1 other device →

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K103824Human Iga Liquid Reagent Kit for Use on Spa PlusThe Binding Site20123

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 3, 2019Class IIZ-1979-2019Material/Component ContaminationCalibration failures with Std.E, Sens.E, or Dup.E error due to atypical reaction kinetics and quality control failures have occurred due to contamination of R1 reagent with R3 reagent.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K103824 (The Binding Site, 2012) has gone furthest, with 3 descendants of its own.
  • 1 other device also cites its predicate K955907.

How does its category compare?

All CZP devices

CZP Iga, Antigen, Antiserum, Control: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CZP

1in 2020

2020: 1
2007high 1 (2020) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CZP

Too few CZP 510(k)s in any year since 2007 to give a median.

Recalls per year · CZP

1in 2022

2022: 1
2007high 1 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CZP

1in 2025

2011: 12015: 12016: 22017: 12021: 12022: 12024: 22025: 1
2007high 2 (2016) · low 0 (2007)2025

MAUDE reports naming product code CZP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—01
20120—00
20130—00
20140—00
20150—01
20160—02
20170—01
20180—00
20190—00
20201—00
20210—01
20220—11
20230—00
20240—02
20250—01
20260—0—

What is it indicated for?

Immanoturbidimetric assay for the quantitative in vitro determination of IgA in human serum and plasma on Roche automated clinical chemistry analyzers.

Measurement aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K040435: Roche Diagnostics Tina-Quant Iga Gen.2.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K040435

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026