# K040696: Modification to Nxstage System One

- Pathway: 510(k) (Special)
- Applicant: Nxstage Medical, Inc.
- Decided: May 26, 2004 after 70 days
- Product code: KDI (Dialyzer, High Permeability With Or Without Sealed Dialysate System)
- Device class: 2
- Regulation: 21 CFR 876.5860
- Recalls of this device: 3 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
No later device cites it as a predicate yet.
0 of 1 device in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. The FDA summary names no predicate by number.

## Recalls of this device
- Apr 14, 2004 · Not classified · Z-1071-04: Pinhole leaks on the replacement (RF) side FMP bag in certain lots of its disposable cartridge balance chambers resulting in loss of dialysate or ultrafiltrate
- Apr 14, 2004 · Not classified · Z-1069-04: Pinhole leaks on the replacement (RF) side FMP bag in certain lots of its disposable cartridge balance chambers resulting in loss of dialysate or ultrafiltrate
- Apr 14, 2004 · Not classified · Z-1070-04: Pinhole leaks on the replacement (RF) side FMP bag in certain lots of its disposable cartridge balance chambers resulting in loss of dialysate or ultrafiltrate

## Indications for use
The NxStage System One is indicated for treatment of renal failure or fluid overload using hemofiltration, hemodialysis and/or ultrafiltration. All treatments must be administered by a health care provider, under physician prescription.