510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Fassier-Duval Telscopic IM System

K041393510(k) TraditionalCleared Aug 17, 200485-day reviewClass II21 CFR 888.3020Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pega Medical, Inc. · HSB Rod, Fixation, Intramedullary And Accessories

Cleared on Aug 17, 2004 as substantially equivalent to K020885 (Pega Medical, Inc., 2003) after an 85-day review. Its family tree reaches back 3 generations to 1996. 14 later devices descend from it. It has been recalled 5 times.

Decided
Aug 17, 2004
Review
85 days
Class
Class II
Product code
HSB
Recalls
5
Predicates
1

Its longest line of descent runs back 3 generations to K960642 (1996). 5 ancestors in all, the oldest K813371 (1982). 14 later devices descend from it, 5 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1998: 1 other ancestor1982: 1 other ancestor2015: 1 descendant cleared2025: 4 descendants cleared2016: 1 descendant cleared2021: 1 descendant cleared2012: 1 descendant cleared2020: 1 descendant cleared2023: 1 descendant cleared2024: 1 descendant cleared2026: 3 descendants clearedK960642 · Nancy Nail · Landos, Inc. · 1996K971783 · Synthes Elastic Intramedullary Nail (Ein) System · Synthes (Usa) · 1997K020885 · Fassire-Duval Telescopic IM System · Pega Medical, Inc. · 2003 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors 1 generation forward: 5 descendants2 generations forward: 4 descendants3 generations forward: 3 descendants4 generations forward: 2 descendants K020885 · Fassire-Duval Telescopic IM System · Pega Medical, Inc. · 2003 · recalledK971783 · Synthes Elastic Intramedullary Nail (Ein) System · Synthes (Usa) · 1997K960642 · Nancy Nail · Landos, Inc. · 19960
  1. Root · 3 generations backK960642 Nancy NailLandos, Inc. · 1996
  2. 2 generations backK971783 Synthes Elastic Intramedullary Nail (Ein) SystemSynthes (Usa) · 1997
  3. 1 generation backK020885 Fassire-Duval Telescopic IM SystemPega Medical, Inc. · 2003 · Recalled
  4. This deviceK041393 Fassier-Duval Telscopic IM SystemPega Medical, Inc. · 2004
  5. Descendant · cites it directly · 7 of its ownK111232 Gap Endo-Exo Medullary SystemPega Medical, Inc. · 2012 · Recalled
  6. Descendant · cites it directly · 6 of its ownK160545 Gap Endo-Exo Medullary SystemPega Medical, Inc. · 2016
  7. Descendant · cites it directly · 6 of its ownK143355 The Simple Locking IntraMedullary (SLIM) SystemPega Medical, Inc. · 2015 · Recalled
  8. Descendant · 2 generations on · 5 of its ownK192710 The Simple Locking Intramedullary (Slim) SystemPega Medical, Inc. · 2020
  9. Descendant · 2 generations on · 2 of its ownK250163 T2 Alpha Femur Retrograde Nailing SystemStryker GmbH · 2025

Showing 5 of 14 descendants. See all descendants

Descendants cleared per year

2012: 12015: 12016: 12020: 12021: 12023: 12024: 12025: 42026: 3
2012high 4 (2025) · low 0 (2013)2026

Most influential descendants

Showing 5 of 14 · See all
  • K111232Gap Endo-Exo Medullary SystemPega Medical, Inc. · 2012 · cites it directly7descendants
  • K160545Gap Endo-Exo Medullary SystemPega Medical, Inc. · 2016 · cites it directly6descendants
  • K143355The Simple Locking IntraMedullary (SLIM) SystemPega Medical, Inc. · 2015 · cites it directly6descendants
  • K192710The Simple Locking Intramedullary (Slim) SystemPega Medical, Inc. · 2020 · 2 generations on5descendants
  • K250163T2 Alpha Femur Retrograde Nailing SystemStryker GmbH · 2025 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K960642Nancy NailLandos, Inc.1996—
−2K971783Synthes Elastic Intramedullary Nail (Ein) SystemSynthes (Usa)1997—
−1K020885Fassire-Duval Telescopic IM SystemPega Medical, Inc.20034
0 (this device)K041393Fassier-Duval Telscopic IM SystemPega Medical, Inc.2004—
+1K111232Gap Endo-Exo Medullary SystemPega Medical, Inc.20121
+1K160545Gap Endo-Exo Medullary SystemPega Medical, Inc.2016—
+1K143355The Simple Locking IntraMedullary (SLIM) SystemPega Medical, Inc.20151
+2K192710The Simple Locking Intramedullary (Slim) SystemPega Medical, Inc.2020—
+2K250163T2 Alpha Femur Retrograde Nailing SystemStryker GmbH2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198210
199810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201510
201610
202010
202110
202310
202410
202540
202630
FIG. 1 — Lineage of K041393, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HSB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HSB device

  1. #1 K200869 Gamma3 Systemscore 19.8 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K020885Fassire-Duval Telescopic IM System
Summary page 2
…substantially equivalent in design and function to itself as predicate device: The intended uses remained unchanged from the original Fassier-Duval Pediatric IM System (K020885) to the modified device, subject of this application. Design changes have been validated via in-vitro biomechanical testing wher…
Pega Medical, Inc.20034

Also cleared against K020885: 4 other devices →

What descends from it? (14)

See all 14

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K111232Gap Endo-Exo Medullary SystemPega Medical, Inc.20127
K160545Gap Endo-Exo Medullary SystemPega Medical, Inc.20166
K143355The Simple Locking IntraMedullary (SLIM) SystemPega Medical, Inc.20156
K211292Fassier-Duval Telescopic IM SystemPega Medical, Inc.20212
K241983Fassier-Duval Telescoping IM SystemOrthopediatrics Canada Ulc Dba Pega Medical20250

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 4, 2024Class IIZ-2941-2024Process controlDuring the assembly of the FRT250 cartridge assembly, an incorrect component was used. Specifically, the FRT240-HEX (2) was used instead of the FRT250-HEX.
Apr 10, 2015Class IIZ-1546-2015Nonconforming Material/ComponentThe hexagonal tip of the Female Driver instrument for the Fassier-Duval IM Telescopic System is 0.06 mm under the specifications. This may result in a reduction of the total strength of the instrument, possibly leading to deformation…
Mar 5, 2014Class IIZ-1472-2014Error in labelingThe outer package label contained the incorrect length of the device. There is a label on the outer package and an additional label on the product contained in the package.
Feb 7, 2014Class IIZ-1448-2014Nonconforming Material/ComponentThe Male Components in this lot are made of material with lower strength than manufacturing specifications.
Jan 29, 2010Not classifiedZ-1494-2011Process controlPossible breakage. Structural weakness has been detected on the Male Driver 3.2mm and 4.0mm sizes. In certain cases, this weakness could lead to a premature failure of the instrument if it is submitted to large torsional loads higher than…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 8 years and passes through 3 companies; a new generation was cleared every 1.5 years on median.
  • 1 of the 5 ancestors has a recall of any class.
  • Of its descendants, K111232 (Pega Medical, Inc., 2012) has gone furthest, with 7 descendants of its own.
  • 4 other devices also cite its predicate K020885.

How does its category compare?

All HSB devices

HSB Rod, Fixation, Intramedullary And Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · HSB

15in 2025

2007: 82008: 112009: 182010: 142011: 162012: 172013: 182014: 192015: 222016: 82017: 232018: 172019: 132020: 172021: 102022: 122023: 112024: 132025: 152026 (partial year): 18
2007high 23 (2017) · low 8 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HSB

106in 2025

2007: 68 days median2008: 57 days median2009: 86 days median2010: 117 days median2011: 78 days median2012: 112 days median2013: 105 days median2014: 147 days median2015: 111 days median2016: 124 days median2017: 119 days median2018: 134 days median2019: 132 days median2020: 112 days median2021: 94 days median2022: 172 days median2023: 162 days median2024: 195 days median2025: 106 days median2026 (partial year): 151 days median
2007high 195 (2024) · low 57 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HSB

18in 2025

2008: 92009: 812010: 182011: 22012: 72013: 142014: 32015: 32016: 62017: 662018: 562019: 62020: 62021: 22022: 12023: 12024: 12025: 182026 (partial year): 1
2007high 81 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HSB

2,911in 2025

2007: 1622008: 1842009: 3342010: 4542011: 6622012: 1,0792013: 1,7732014: 1,4742015: 1,6712016: 1,9202017: 1,4472018: 3,5262019: 2,5172020: 3,2992021: 4,1112022: 2,7332023: 2,2192024: 2,8692025: 2,911
2007high 4,111 (2021) · low 162 (2007)2025

MAUDE reports naming product code HSB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078680162
200811579184
2009188681334
20101411718454
201116782662
20121711271,079
201318105141,773
20141914731,474
20152211131,671
2016812461,920
201723119661,447
201817134563,526
20191313262,517
20201711263,299
2021109424,111
20221217212,733
20231116212,219
20241319512,869
202515106182,911
2026181511—

What is it indicated for?

This implant is indicated as a temporary implant to aid in the healing of long diaphysis fractures, osteotomies, malunions and nonunions and to prevent further fractures in femur, tibia and humerus in pediatric patients suffering from Osteogenesis Imperfecta without disrupting the bone growth plate. It can be used in procedures such as bone lengthening/shortening concomitantly with external fixators in pediatric or small stature patients with limb length discrepancy.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K041393: Fassier-Duval Telscopic IM System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K041393

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026