Vitros Chemistry Products DLDL Reagent, Calibrator Kit 19, FS Calibrator 1 and Performance Verifiers I and II

K041720510(k) TraditionalCleared Aug 25, 200462-day reviewClass I21 CFR 862.1475Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ortho-Clinical Diagnostics, Inc. · MRR System, Test, Low Density, Lipoprotein

Cleared on Aug 25, 2004 as substantially equivalent to K020852 (Polymedco, Inc., 2002) and 1 other predicate after a 62-day review. Its family tree reaches back 1 generation to 1990. 7 later devices descend from it. It has been recalled once.

Decided
Aug 25, 2004
Review
62 days
Class
Class I
Product code
MRR
Recalls
1
Predicates
2

Its longest line of descent runs back 1 generation to K904768 (1990). 2 ancestors in all. 7 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor2004: 3 descendants cleared2005: 1 descendant cleared2006: 2 descendants cleared2013: 1 descendant clearedK904768 · Kodatrol I and II Control and Diluent Set · Eastman Kodak Company · 1990 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 1 descendant3 generations forward: 2 descendants4 generations forward: 1 descendant K904768 · Kodatrol I and II Control and Diluent Set · Eastman Kodak Company · 1990 · recalled0
  1. Root · 1 generation backK904768 Kodatrol I and II Control and Diluent SetEastman Kodak Company · 1990 · Recalled
  2. This deviceK041720 Vitros Chemistry Products DLDL Reagent, Calibrator Kit 19, FS…Ortho-Clinical Diagnostics, Inc. · 2004
  3. Descendant · cites it directly · 4 of its ownK042477 Vitros Chemistry Products Trfrn Reahent; Vitros Chemistry Products…Ortho-Clinical Diagnostics, Inc. · 2004
  4. Descendant · 2 generations on · 3 of its ownK052819 Vitros Chemistry Products Aat Reagent, Calibrator Kit 99 and Aat…Ortho-Clinical Diagnostics, Inc. · 2005
  5. Descendant · 3 generations on · 1 of its ownK061588 Vitros Chemistry Products Hcy Reagent, Vitros Chemistry Products…Ortho-Clinical Diagnostics, Inc. · 2006 · Recalled
  6. Descendant · 4 generations on · 0 of its ownK123930 Vitros Chemistry Products Hcy 2 Reagent, Calibrator Kit 27, Hcy 2…Ortho-Clinical Diagnostics, Inc. · 2013
  7. Descendant · 3 generations on · 0 of its ownK060940 Vitros Chemistry Products Calibrator Kit 99, Vitros Chemistry…Ortho-Clinical Diagnostics, Inc. · 2006

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2004: 32005: 12006: 22013: 1
2004high 3 (2004) · low 0 (2007)2026

Most influential descendants

Showing 5 of 7 · See all
  • K042477Vitros Chemistry Products Trfrn Reahent; Vitros Chemistry Products Calibrator Kit 20…Ortho-Clinical Diagnostics, Inc. · 2004 · cites it directly4descendants
  • K052819Vitros Chemistry Products Aat Reagent, Calibrator Kit 99 and Aat Performance Verifiers…Ortho-Clinical Diagnostics, Inc. · 2005 · 2 generations on3descendants
  • K061588Vitros Chemistry Products Hcy Reagent, Vitros Chemistry Products Calibrator Kit 27, Hcy…Ortho-Clinical Diagnostics, Inc. · 2006 · 3 generations on1descendant
  • K123930Vitros Chemistry Products Hcy 2 Reagent, Calibrator Kit 27, Hcy 2 Performance Verifiers…Ortho-Clinical Diagnostics, Inc. · 2013 · 4 generations on0descendants
  • K060940Vitros Chemistry Products Calibrator Kit 99, Vitros Chemistry Products Aat/HPT…Ortho-Clinical Diagnostics, Inc. · 2006 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K904768Kodatrol I and II Control and Diluent SetEastman Kodak Company19901
0 (this device)K041720Vitros Chemistry Products DLDL Reagent, Calibrator Kit 19, FS Calibrator 1 and…Ortho-Clinical Diagnostics, Inc.2004—
+1K042477Vitros Chemistry Products Trfrn Reahent; Vitros Chemistry Products Calibrator…Ortho-Clinical Diagnostics, Inc.2004—
+2K052819Vitros Chemistry Products Aat Reagent, Calibrator Kit 99 and Aat Performance…Ortho-Clinical Diagnostics, Inc.2005—
+3K061588Vitros Chemistry Products Hcy Reagent, Vitros Chemistry Products Calibrator…Ortho-Clinical Diagnostics, Inc.20061
+4K123930Vitros Chemistry Products Hcy 2 Reagent, Calibrator Kit 27, Hcy 2 Performance…Ortho-Clinical Diagnostics, Inc.2013—
+3K060940Vitros Chemistry Products Calibrator Kit 99, Vitros Chemistry Products Aat/HPT…Ortho-Clinical Diagnostics, Inc.2006—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200430
200510
200620
201310
FIG. 1 — Lineage of K041720, 1990–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MRR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MRR device

  1. #1 K241800 Atellica® CH HDL Cholesterol (HDLC)& Atellica®…score 8.0 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K020852Poly-Chem Glucose
Summary page 3
…alibrator 1 are substantially equivalent to the Polymedco Lipi+Plus Direct LDL assay (predicate device) which was cleared by the FDA (K020852) for IVD use. The relationship between the VITROS dLDL and the predicate device,…
Polymedco, Inc.2002—
*K904768Kodatrol I and II Control and Diluent Set
Summary page 4
…tantially equivalent to the VITROS Chemistry Products Performance Verifiers (predicate device) which were cleared by the FDA (K904768) for IVD use. Table 2 Table 2 lists the similarities and differences of the device characteristics between…
Eastman Kodak Company19901

What descends from it? (7)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K042477Vitros Chemistry Products Trfrn Reahent; Vitros Chemistry Products Calibrator Kit 20…Ortho-Clinical Diagnostics, Inc.20044
K042643Vitros Chemistry Product Malb Reagent, Calibrator Kit 24, Malb Performance Verifiers I…Ortho-Clinical Diagnostics, Inc.20040
K042494Vitros Chemistry Products C3 and CR Reagent, Calibrator Kit 20, Protein Performance…Ortho-Clinical Diagnostics20040

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 8, 2018Class IIZ-0476-2019Under Investigation by firmVITROS Chemistry Products dLDL Reagent GENs 26 and 27, negatively biased results

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K042477 (Ortho-Clinical Diagnostics, Inc., 2004) has gone furthest, with 4 descendants of its own.

How does its category compare?

All MRR devices

MRR System, Test, Low Density, Lipoprotein: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MRR

1in 2024

2008: 12009: 22017: 12023: 12024: 1
2007high 2 (2009) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MRR

Too few MRR 510(k)s in any year since 2007 to give a median.

Recalls per year · MRR

1in 2025

2008: 22018: 12025: 1
2007high 2 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MRR

1in 2024

2011: 32017: 32018: 22020: 12021: 12022: 12023: 12024: 1
2007high 3 (2011) · low 0 (2007)2025

MAUDE reports naming product code MRR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20081—20
20092—00
20100—00
20110—03
20120—00
20130—00
20140—00
20150—00
20160—00
20171—03
20180—12
20190—00
20200—01
20210—01
20220—01
20231—01
20241—01
20250—10
20260—0—

What is it indicated for?

For in vitro diagnostic use only. VITROS Chemistry Products dLDL Reagent is used to quantitatively measure LDL Cholesterol (LDLC) concentration in serum and plasma. Low Density Lipoprotein (LDL) cholesterol is used to evaluate the risk of developing coronary heart disease (CHD). The risk of CHD increases with higher LDL cholesterol concentrations. For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 19 is used in conjunction with…

Read the full statement (119 words)

For in vitro diagnostic use only. VITROS Chemistry Products dLDL Reagent is used to quantitatively measure LDL Cholesterol (LDLC) concentration in serum and plasma. Low Density Lipoprotein (LDL) cholesterol is used to evaluate the risk of developing coronary heart disease (CHD). The risk of CHD increases with higher LDL cholesterol concentrations.

For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 19 is used in conjunction with VITROS FS Calibrator 1 to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of HDL and LDL cholesterol using VITROS dHDL and dLDL Reagents.

For in vitro diagnostic use only. VITROS Chemistry Products Performance Verifiers I and II are assayed controls used to monitor performance on VITROS Chemistry Systems.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K041720: Vitros Chemistry Products DLDL Reagent, Calibrator Kit 19, FS Calibrator 1 and Performance Verifiers I and II.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K041720

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026