510(k) Report

510(k) · Cleared

Cleared

MPS (or MPS-T)

K041838510(k) TraditionalCleared Mar 1, 2005237-day reviewClass II21 CFR 882.1620

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Innerspace, Inc. · GWM Device, Monitoring, Intracranial Pressure

Cleared on Mar 1, 2005 as substantially equivalent to K013005 (Innerspace, Inc., 2002) and 2 other predicates after a 237-day review. Its family tree reaches back 2 generations to 2000. 3 later devices descend from it.

Decided
Mar 1, 2005
Review
237 days
Class
Class II
Product code
GWM
Recalls
None
Predicates
3

Its longest line of descent runs back 2 generations to K002765 (2000). 4 ancestors in all. 3 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor2001: 1 other ancestor2008: 1 descendant cleared2009: 1 descendant cleared2016: 1 descendant clearedK002765 · Licox CMP Monitor-AC3.1, Licox Imc Oxygen Catheters-CC1,CC1.SB Luicox Imc Introducer Kits-IM,IM2,IM3,II1 Licoximc Tempe · Integra Neurosciences · 2000 · recalledK013005 · Act 11 MP Ventricular Catheter · Innerspace, Inc. · 2002200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant K013005 · Act 11 MP Ventricular Catheter · Innerspace, Inc. · 2002K002765 · Licox CMP Monitor-AC3.1, Licox Imc Oxygen Catheters-CC1,CC1.SB Luicox Imc Introducer Kits-IM,IM2,IM3,II1 Licoximc Tempe · Integra Neurosciences · 2000 · recalled0
  1. Root · 2 generations backK002765 Licox CMP Monitor-AC3.1, Licox Imc Oxygen Catheters-CC1,CC1.SB…Integra Neurosciences · 2000 · Recalled
  2. 1 generation backK013005 Act 11 MP Ventricular CatheterInnerspace, Inc. · 2002
  3. This deviceK041838 MPS (or MPS-T)Innerspace, Inc. · 2005
  4. Descendant · cites it directly · 2 of its ownK072379 MPS Oxiport and MPS Oxiport PlusInnerspace, Inc. · 2008
  5. Descendant · 2 generations on · 1 of its ownK083378 SynergyInnerspace, Inc. · 2009
  6. Descendant · 3 generations on · 0 of its ownK161010 HICP200 Patient/Monitor Interconnect CableInnerspace Neuro Solutions, Inc. · 2016

Most influential descendants

Showing 3 of 3 · See all
  • K072379MPS Oxiport and MPS Oxiport PlusInnerspace, Inc. · 2008 · cites it directly2descendants
  • K083378SynergyInnerspace, Inc. · 2009 · 2 generations on1descendant
  • K161010HICP200 Patient/Monitor Interconnect CableInnerspace Neuro Solutions, Inc. · 2016 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K002765Licox CMP Monitor-AC3.1, Licox Imc Oxygen Catheters-CC1,CC1.SB Luicox Imc…Integra Neurosciences20003
−1K013005Act 11 MP Ventricular CatheterInnerspace, Inc.2002—
0 (this device)K041838MPS (or MPS-T)Innerspace, Inc.2005—
+1K072379MPS Oxiport and MPS Oxiport PlusInnerspace, Inc.2008—
+2K083378SynergyInnerspace, Inc.2009—
+3K161010HICP200 Patient/Monitor Interconnect CableInnerspace Neuro Solutions, Inc.2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
200910
201610
FIG. 1 — Lineage of K041838, 2000–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GWM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GWM device

  1. #1 K252241 Pressio 3 Multi-parameter Neuromonitoring Systemscore 12.9 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K013005Act 11 MP Ventricular Catheter
Summary page 1
…CFR 882.4100 (84HCA 2.4 Equivalent device The bolt system is substantially equivalent to ACT IIMP plus coveted in 510-K 013005. The 005 device is a bolt that can receive the ventricular catheter of ACT II (705) and hold a second parameter p…
Innerspace, Inc.2002—
*K013705Act Iii Ventricular Catheter
Summary page 1
…r catheter of ACT II (705) and hold a second parameter probe. The pressure sensing system used is the same as that used in ACT IJR. It is covered in 510-K 013705 (705). The drainage catheter system is similar to that used in 705 except the catheter size is smaller. The MIP…
Innerspace, Inc.2002—
*K002765Licox CMP Monitor-AC3.1, Licox Imc Oxygen Catheters-CC1,CC1.SB Luicox Imc Introducer…
Summary page 1
…e MIPS- OXIPORT bolt guide tubes are similar to that described in ACT II MN(005). The guide tubes are substantially equivalent to guide tubes described in Licox (002765) as regards having the tubes extend beyond the dura - 2.5 Description of the Catheter's ICP Mon…
Integra Neurosciences20003

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K072379MPS Oxiport and MPS Oxiport PlusInnerspace, Inc.20082

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 5 years and passes through 2 companies; a new generation was cleared every 3.0 years on median.
  • 1 of the 4 ancestors has a recall of any class.
  • Of its descendants, K072379 (Innerspace, Inc., 2008) has gone furthest, with 2 descendants of its own.

How does its category compare?

All GWM devices

GWM Device, Monitoring, Intracranial Pressure: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GWM

1in 2025

2007: 22008: 22009: 12010: 22011: 32012: 42013: 22014: 22016: 32017: 32018: 12019: 22020: 32021: 22024: 22025: 12026 (partial year): 1
2007high 4 (2012) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GWM

30in 2020

2011: 123 days median2012: 100 days median2016: 192 days median2017: 37 days median2020: 30 days median
2007high 192 (2016) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GWM

3in 2025

2008: 12010: 92011: 62012: 12013: 122014: 22019: 12023: 1, 1 Class I2024: 4, 2 Class I2025: 3
2007high 12 (2013) · low 0 (2007)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · GWM

441in 2025

2007: 1262008: 1052009: 1172010: 1042011: 1842012: 1512013: 1202014: 3602015: 4862016: 4192017: 4452018: 3022019: 2602020: 2682021: 2112022: 4242023: 7442024: 6412025: 441
2007high 744 (2023) · low 104 (2010)2025

MAUDE reports naming product code GWM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—0126
20082—1105
20091—0117
20102—9104
201131236184
201241001151
20132—12120
20142—2360
20150—0486
201631920419
20173370445
20181—0302
20192—1260
20203300268
20212—0211
20220—0424
20230—1 (1)744
20242—4 (2)641
20251—3441
20261—0—

What is it indicated for?

The use of MPS (or MPS-T) by a qualified neurosurgeon is indicated when direct measurement of intracranial pressure is clinically important, when the patient may require CSF drainage in the course of their care and when data from a second or third parameter may be deemed useful in optimum patient management.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K041838: MPS (or MPS-T).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K041838

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026