510(k) · Cleared
Recalled · Class II · 2015Fujinon G5 Duodenscopes; G5 Duodenoscopes, ED-450XT5, ED-250XT5
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Fujinon, Inc. · FDT Duodenoscope And Accessories, Flexible/Rigid
Cleared on Nov 18, 2004 as substantially equivalent to K944620 (Marks & Murase, 1995) and 1 other predicate after a 108-day review. Its family tree reaches back 1 generation to 1995. 6 later devices descend from it. It has been recalled 4 times.
- Decided
- Nov 18, 2004
- Review
- 108 days
- Class
- Class II
- Product code
- FDT
- Recalls
- 4
- Predicates
- 2
Its longest line of descent runs back 1 generation to K944759 (1995). 2 ancestors in all. 6 later devices descend from it, 1 citing it directly.
- Root · 1 generation backK944759 200 Series Gastro-Intestinal Video SystemFujinon, Inc. · 1995 · Recalled
- This deviceK042076 Fujinon G5 Duodenscopes; G5 Duodenoscopes, ED-450XT5, ED-250XT5Fujinon, Inc. · 2004
- Descendant · cites it directly · 5 of its ownK152257 Fujifilm Duodenoscope Model ED-530XTFujifilm Medical Systems U.S.A, Inc. · 2017 · Recalled
- Descendant · 2 generations on · 3 of its ownK181745 FUJIFILM Duodenoscope ModelFujifilm Corporation · 2019
- Descendant · 3 generations on · 2 of its ownK191747 FUJIFILM Duodenoscope Model ED-580XTFujifilm Corporation · 2019
- Descendant · 4 generations on · 0 of its ownK251861 FUJIFILM Endoscope Model ED-S100TP; FUJIFILM Endoscope Model…Fujifilm Corporation · 2026
- Descendant · 4 generations on · 0 of its ownK203028 FUJIFILM Distal End CapFujifilm Corporation · 2020
Showing 5 of 6 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 6 · See all- K152257Fujifilm Duodenoscope Model ED-530XTFujifilm Medical Systems U.S.A, Inc. · 2017 · cites it directly5descendants
- K181745FUJIFILM Duodenoscope ModelFujifilm Corporation · 2019 · 2 generations on3descendants
- K191747FUJIFILM Duodenoscope Model ED-580XTFujifilm Corporation · 2019 · 3 generations on2descendants
- K251861FUJIFILM Endoscope Model ED-S100TP; FUJIFILM Endoscope Model ED-S100XP; FUJIFILM…Fujifilm Corporation · 2026 · 4 generations on0descendants
- K203028FUJIFILM Distal End CapFujifilm Corporation · 2020 · 4 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K944759 | 200 Series Gastro-Intestinal Video System | Fujinon, Inc. | 1995 | 5 |
| 0 (this device) | K042076 | Fujinon G5 Duodenscopes; G5 Duodenoscopes, ED-450XT5, ED-250XT5 | Fujinon, Inc. | 2004 | — |
| +1 | K152257 | Fujifilm Duodenoscope Model ED-530XT | Fujifilm Medical Systems U.S.A, Inc. | 2017 | 3 |
| +2 | K181745 | FUJIFILM Duodenoscope Model | Fujifilm Corporation | 2019 | — |
| +3 | K191747 | FUJIFILM Duodenoscope Model ED-580XT | Fujifilm Corporation | 2019 | — |
| +4 | K251861 | FUJIFILM Endoscope Model ED-S100TP; FUJIFILM Endoscope Model ED-S100XP… | Fujifilm Corporation | 2026 | — |
| +4 | K203028 | FUJIFILM Distal End Cap | Fujifilm Corporation | 2020 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1995 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2017 | 1 | 0 |
| 2018 | 1 | 0 |
| 2019 | 2 | 0 |
| 2020 | 1 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FDT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FDT device
- #1 K251861 FUJIFILM Endoscope Model ED-S100TP; FUJIFILM…score 13.7 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K944620 | Series 400 Gastroenterology SystemSummary page 1…G5 Duodenoscopes Common Name: Duodenoscopes Classification: Class 11, 21 CFR 876.1500, FAK Predicate Device 121 CFR 807.92(a)(3)J * Fujinon ED-400XL - K944620 * Fujinon ED-200XU - K944759 The G5 changes were described and cleared by FDA in the 510(k) submission for the Fuj… | 1995 |
| * | K944759 | 200 Series Gastro-Intestinal Video SystemSummary page 1…oscopes Classification: Class 11, 21 CFR 876.1500, FAK Predicate Device 121 CFR 807.92(a)(3)J * Fujinon ED-400XL - K944620 * Fujinon ED-200XU - K944759 The G5 changes were described and cleared by FDA in the 510(k) submission for the Fujinon Double Bal- loon Enteroscopy Sy… | 1995 |
What descends from it? (6)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K152257 | Fujifilm Duodenoscope Model ED-530XT | 2017 | 5 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Dec 23, 2015 | Class II | Z-1864-2016 | Other | This correction is in response to publicized reports of multi-drug resistant bacteria on endoscopes used for Endoscopic Retrograde Cholangiopancreatogram (ERCP) procedures. |
| Dec 23, 2015 | Class II | Z-1863-2016 | Other | This correction is in response to publicized reports of multi-drug resistant bacteria on endoscopes used for Endoscopic Retrograde Cholangiopancreatogram (ERCP) procedures. |
| Dec 23, 2015 | Class II | Z-1860-2016 | Other | This correction is in response to publicized reports of multi-drug resistant bacteria on endoscopes used for Endoscopic Retrograde Cholangiopancreatogram (ERCP) procedures. |
| Jun 30, 2015 | Class II | Z-2778-2015 | Device Design | The suction button on the ED-530XT endoscope may become lodged (stuck) in the active position following the use of x-ray contrast agents during ERCP procedures. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All FDT devicesFDT Duodenoscope And Accessories, Flexible/Rigid: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · FDT
2in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FDT
72in 2020
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FDT
7so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · FDT
3,946in 2025
MAUDE reports naming product code FDT, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 4 |
| 2008 | 1 | — | 0 | 11 |
| 2009 | 1 | — | 0 | 18 |
| 2010 | 0 | — | 0 | 12 |
| 2011 | 1 | — | 0 | 10 |
| 2012 | 0 | — | 0 | 15 |
| 2013 | 0 | — | 0 | 18 |
| 2014 | 0 | — | 0 | 36 |
| 2015 | 0 | — | 1 | 246 |
| 2016 | 1 | — | 1 | 247 |
| 2017 | 2 | — | 0 | 791 |
| 2018 | 3 | 63 | 1 | 481 |
| 2019 | 5 | 88 | 0 | 1,347 |
| 2020 | 4 | 72 | 0 | 822 |
| 2021 | 2 | — | 1 | 1,270 |
| 2022 | 1 | — | 0 | 2,619 |
| 2023 | 0 | — | 3 | 3,050 |
| 2024 | 2 | — | 0 | 3,019 |
| 2025 | 2 | — | 1 | 3,946 |
| 2026 | 2 | — | 7 | — |
What is it indicated for?
The device is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K042076 · printed Nov 4, 2026
Cite this page
510(k) Report. “K042076: Fujinon G5 Duodenscopes; G5 Duodenoscopes, ED-450XT5, ED-250XT5.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K042076
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026