510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

Model 6416 Temporary Pacing Lead System

K042190510(k) SpecialCleared Sep 7, 200426-day reviewClass II21 CFR 870.3680Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Vascular · LDF Electrode, Pacemaker, Temporary

Cleared on Sep 7, 2004 after a 26-day review. The FDA summary names its predicate without a number. 2 later devices descend from it. It has been recalled once.

Decided
Sep 7, 2004
Review
26 days
Class
Class II
Product code
LDF
Recalls
1
Predicates
0

2 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2016: 1 descendant cleared2026: 1 descendant cleared200020052010201520202025
  1. This deviceK042190 Model 6416 Temporary Pacing Lead SystemMedtronic Vascular · 2004
  2. Descendant · cites it directly · 1 of its ownK160260 Tempo Temporary Pacing LeadBiotrace Medical, Inc. · 2016
  3. Descendant · 2 generations on · 0 of its ownK252042 Sync AV II Temporary Pacing CatheterSwift Sync, Inc. · 2026

Most influential descendants

Showing 2 of 2 · See all
  • K160260Tempo Temporary Pacing LeadBiotrace Medical, Inc. · 2016 · cites it directly1descendant
  • K252042Sync AV II Temporary Pacing CatheterSwift Sync, Inc. · 2026 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K042190Model 6416 Temporary Pacing Lead SystemMedtronic Vascular2004—
+1K160260Tempo Temporary Pacing LeadBiotrace Medical, Inc.2016—
+2K252042Sync AV II Temporary Pacing CatheterSwift Sync, Inc.2026—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
202610
FIG. 1 — Lineage of K042190, 2004–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LDF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LDF device

  1. #1 K241334 Semi-Floating Temporary Pacing Electrode…score 12.1 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—current market released Model 6416 Bipolar Transvenous Temporary Pacing System
Named without a submission number
—

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K160260Tempo Temporary Pacing LeadBiotrace Medical, Inc.20161

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 16, 2016Class IIZ-2200-2016Nonconforming Material/ComponentThis Pacing Lead System does not comply with Section 8.5.2.3 of IEC 60601-1, and corresponding provisions of FDA 21 CFR 898, which relate to design standards to prevent connecting a patients lead to a possible hazardous voltage.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K160260 (Biotrace Medical, Inc., 2016) has gone furthest, with 1 descendant of its own.

How does its category compare?

All LDF devices

LDF Electrode, Pacemaker, Temporary: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LDF

5in 2025

2008: 12014: 12015: 22016: 32017: 12018: 12019: 12024: 12025: 52026 (partial year): 1
2007high 5 (2025) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LDF

266in 2025

2016: 138 days median2025: 266 days median
2007high 266 (2025) · low 138 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LDF

1in 2018

2008: 22009: 22011: 12012: 12018: 1, 1 Class I2026 (partial year): 1, 1 Class I
2007high 2 (2008) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · LDF

139in 2025

2007: 252008: 182009: 442010: 482011: 352012: 682013: 742014: 682015: 812016: 902017: 792018: 912019: 1302020: 832021: 1252022: 1172023: 2122024: 1902025: 139
2007high 212 (2023) · low 18 (2008)2025

MAUDE reports naming product code LDF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—025
20081—218
20090—244
20100—048
20110—135
20120—168
20130—074
20141—068
20152—081
20163138090
20171—079
20181—1 (1)91
20191—0130
20200—083
20210—0125
20220—0117
20230—0212
20241—0190
202552660139
20261—1 (1)—

What is it indicated for?

The Model 6416 Bipolar Transvenous Temporary Pacing System features an active fixation, bipolar lead and a soft-tipped lubricated guide catheter. The system is designed for temporary intracardiac pacing and/or EGM recording.

The system is disposable, for temporary single patient use with a contemplated implant duration of 7 days or less. The lead and accessories are supplied sterile.

The lead is introduced transvenously using the guide catheter. Once within the appropriate chamber, the helical tip electrode of the lead is actively fixed into the endocardium. After lead placement, the guide catheter is removed by sliding it over the lead's bifurcated connector.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K042190: Model 6416 Temporary Pacing Lead System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K042190

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026