# K042651: Free T3 Calibrators on the Access Immunoassay Systems

- Pathway: 510(k) (Traditional)
- Applicant: Beckman Coulter, Inc.
- Decided: Oct 5, 2004 after 7 days
- Product code: JIT (Calibrator, Secondary)
- Device class: 2
- Regulation: 21 CFR 862.1150
- Recalls of this device: 1 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 1 generation back to K003098 (2000). 1 ancestor in all.
No later device cites it as a predicate yet.
0 of 2 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K003098: Access Ultrasensitive HGH Calibrators, Beckman Coulter, Inc. (2000) Summary page 1.

## Recalls of this device
- Dec 17, 2009 · Not classified · Z-2178-2011: Beckman Coulter has confirmed reports that the Access Free T3 calibrator lot numbers 989902, 989903, 989904, 989905, 989906 and 989907 will show a decline in performance before reaching their stated e

## Indications for use
The Access Free T3 Calibrator set is a device intended for medical purposes for use in the Access Immunoassay Systems to establish points of reference that are used in the determination of values in the measurement of free T3 levels in human serum and plasma.