Smart Bitebloc

K042665510(k) TraditionalCleared Nov 23, 200455-day reviewClass II21 CFR 868.1400Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Oridion Medical 1987, Ltd. · CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase

Cleared on Nov 23, 2004 as substantially equivalent to K954352 (United States Endoscopy Group, Inc., 1996) and 5 other predicates after a 55-day review. Its family tree reaches back 1 generation to 1990. 3 later devices descend from it. It has been recalled 2 times.

Decided
Nov 23, 2004
Review
55 days
Class
Class II
Product code
CCK
Recalls
2
Predicates
6

Its longest line of descent runs back 1 generation to K896691 (1990). 6 ancestors in all. 3 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor1993: 2 other ancestors1992: 2 other ancestors2008: 1 descendant cleared2014: 1 descendant cleared2024: 1 descendant clearedK896691 · Bite Block · Endoscopix · 19901990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 5 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant K896691 · Bite Block · Endoscopix · 19900
  1. Root · 1 generation backK896691 Bite BlockEndoscopix · 1990
  2. This deviceK042665 Smart BiteblocOridion Medical 1987, Ltd. · 2004
  3. Descendant · cites it directly · 2 of its ownK080527 Twinguard with Capnograph AccessoryTrawax Pty Limited · 2008
  4. Descendant · 2 generations on · 1 of its ownK140473 DualguardFlexicare Medical, Ltd. · 2014
  5. Descendant · 3 generations on · 0 of its ownK240937 AIM (N/A)Inventeur, LLC · 2024

Most influential descendants

Showing 3 of 3 · See all
  • K080527Twinguard with Capnograph AccessoryTrawax Pty Limited · 2008 · cites it directly2descendants
  • K140473DualguardFlexicare Medical, Ltd. · 2014 · 2 generations on1descendant
  • K240937AIM (N/A)Inventeur, LLC · 2024 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K896691Bite BlockEndoscopix1990—
0 (this device)K042665Smart BiteblocOridion Medical 1987, Ltd.2004—
+1K080527Twinguard with Capnograph AccessoryTrawax Pty Limited2008—
+2K140473DualguardFlexicare Medical, Ltd.2014—
+3K240937AIM (N/A)Inventeur, LLC2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199220
199320
199610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
201410
202410
FIG. 1 — Lineage of K042665, 1990–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CCK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CCK device

  1. #1 K253030 Capnostream™35 Portable Respiratory Monitor…score 15.8 · 16 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (6)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K954352Bite Block
Summary page 2
…s. Oridion considers this device to be a capnography accessory. SUBSTANTIAL EQUIVALENCE INFORMATION Predicate Devices United States Endoscopy K954352 United States Endoscopy K924304 Endoscopix Bite Block K896691 GI Supply Bite Bloc…
United States Endoscopy Group, Inc.1996—
*K924304Bite Block
Summary page 2
…aphy accessory. SUBSTANTIAL EQUIVALENCE INFORMATION Predicate Devices United States Endoscopy K954352 United States Endoscopy K924304 Endoscopix Bite Block K896691 GI Supply Bite Block K915816 GI Supply Oxy-Block K9…
United States Endoscopy Group, Inc.1993—
*K896691Bite Block
Summary page 2
…RMATION Predicate Devices United States Endoscopy K954352 United States Endoscopy K924304 Endoscopix Bite Block K896691 GI Supply Bite Block K915816 GI Supply Oxy-Block K931044 Stantex Pty Ltd Oxiguard…
Endoscopix1990—
*K915816Bite Block/Mouthpiece for Endoscopy
Summary page 2
…United States Endoscopy K954352 United States Endoscopy K924304 Endoscopix Bite Block K896691 GI Supply Bite Block K915816 GI Supply Oxy-Block K931044 Stantex Pty Ltd Oxiguard K914978 The Smart BiteBlocTM has the same…
Gi Supply1992—
*K931044Oxy-Bloc
Summary page 2
…United States Endoscopy K924304 Endoscopix Bite Block K896691 GI Supply Bite Block K915816 GI Supply Oxy-Block K931044 Stantex Pty Ltd Oxiguard K914978 The Smart BiteBlocTM has the same intended use as the predicate devices identi…
Gi Supply1993—
*K914978Oxyguard
Summary page 2
…Endoscopix Bite Block K896691 GI Supply Bite Block K915816 GI Supply Oxy-Block K931044 Stantex Pty Ltd Oxiguard K914978 The Smart BiteBlocTM has the same intended use as the predicate devices identified in this section. Any technological differenc…
Stantex Pty, Ltd.1992—

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K080527Twinguard with Capnograph AccessoryTrawax Pty Limited20082

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 16, 2012Not classifiedZ-1333-2012Packaging process controlCO2 sampling lines may have paper and plastic particulate present in the packaged units at levels that may exceed visual specification.
Feb 16, 2012Not classifiedZ-1334-2012Packaging process controlCO2 sampling lines may have paper and plastic particulate present in the packaged units at levels that may exceed visual specification.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 6 ancestors has a recall.
  • Of its descendants, K080527 (Trawax Pty Limited, 2008) has gone furthest, with 2 descendants of its own.

How does its category compare?

All CCK devices

CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CCK

3in 2025

2007: 52008: 32009: 32010: 32011: 22012: 72013: 42014: 92015: 52016: 22017: 52018: 62019: 72020: 42021: 42022: 32023: 42024: 32025: 32026 (partial year): 3
2007high 9 (2014) · low 2 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CCK

169in 2025

2007: 56 days median2008: 248 days median2009: 101 days median2010: 125 days median2012: 150 days median2013: 132 days median2014: 166 days median2015: 105 days median2017: 134 days median2018: 296 days median2019: 261 days median2020: 188 days median2021: 227 days median2022: 226 days median2023: 555 days median2024: 592 days median2025: 169 days median2026 (partial year): 224 days median
2007high 592 (2024) · low 56 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CCK

1in 2022

2008: 12009: 192010: 42011: 142012: 32013: 2, 1 Class I2015: 12018: 52019: 7, 7 Class I2020: 12022: 1
2007high 19 (2009) · low 0 (2007)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · CCK

1,259in 2025

2007: 72008: 202009: 212010: 122011: 182012: 132013: 282014: 512015: 1142016: 882017: 582018: 632019: 1412020: 2,3022021: 2,2722022: 2,4752023: 1,6372024: 1,2162025: 1,259
2007high 2,475 (2022) · low 7 (2007)2025

MAUDE reports naming product code CCK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200755607
20083248120
200931011921
20103125412
20112—1418
20127150313
201341322 (1)28
20149166051
201551051114
20162—088
20175134058
20186296563
201972617 (7)141
2020418812,302
2021422702,272
2022322612,475
2023455501,637
2024359201,216
2025316901,259
202632240—

What is it indicated for?

The proposed device, Smart BiteBloc TM, will be used during endoscopy procedures that require a bite block that enables and supports the convenient use of, and will not interfere with, an oral nasal cannula (KO1 1536) for collecting samples of the patient's breathing to measure CO 2 with a capnograph while simultaneously administering supplemental oxygen near the nose for inhalation. It can be used for non intubated patients who weigh more than 55 lbs (25 kg).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K042665: Smart Bitebloc.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K042665

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026