Anspach Minimal Access Spinal Attachment (Masa) System

K042783510(k) TraditionalCleared Dec 17, 200472-day reviewClass II21 CFR 882.4310Recalls 31

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

The Anspach Effort, Inc. · HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)

Cleared on Dec 17, 2004 after a 72-day review. The FDA summary names its predicate without a number. 14 later devices descend from it. It has been recalled 31 times.

Decided
Dec 17, 2004
Review
72 days
Class
Class II
Product code
HBE
Recalls
31
Predicates
0

14 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2011: 1 descendant cleared2018: 2 descendants cleared2019: 3 descendants cleared2014: 1 descendant cleared2023: 1 descendant cleared2024: 2 descendants cleared2022: 1 descendant cleared2025: 2 descendants cleared2026: 1 descendant cleared200020052010201520202025
  1. This deviceK042783 Anspach Minimal Access Spinal Attachment (Masa) SystemThe Anspach Effort, Inc. · 2004
  2. Descendant · cites it directly · 13 of its ownK113476 Anspach Dissection ToolsThe Anspach Effort, Inc. · 2011 · Recalled
  3. Descendant · 2 generations on · 4 of its ownK130755 Hicut Highspeed InstrumentAdeor Medical AG · 2014
  4. Descendant · 3 generations on · 3 of its ownK232684 Surgify HaloSurgify Medical OY · 2023
  5. Descendant · 2 generations on · 3 of its ownK173066 OSSEODUO Shaver and Drill SystemBien-Air Surgery SA · 2018
  6. Descendant · 4 generations on · 2 of its ownK250380 Surgify Halo (54.085.SHD.H1); Surgify Halo (54.140.SHD.H1); Surgify…Surgify Medical OY · 2025 · Recalled

Showing 5 of 14 descendants. See all descendants

Descendants cleared per year

2011: 12014: 12018: 22019: 32022: 12023: 12024: 22025: 22026: 1
2011high 3 (2019) · low 0 (2012)2026

Most influential descendants

Showing 5 of 14 · See all
  • K113476Anspach Dissection ToolsThe Anspach Effort, Inc. · 2011 · cites it directly13descendants
  • K130755Hicut Highspeed InstrumentAdeor Medical AG · 2014 · 2 generations on4descendants
  • K232684Surgify HaloSurgify Medical OY · 2023 · 3 generations on3descendants
  • K173066OSSEODUO Shaver and Drill SystemBien-Air Surgery SA · 2018 · 2 generations on3descendants
  • K250380Surgify Halo (54.085.SHD.H1); Surgify Halo (54.140.SHD.H1); Surgify Halo…Surgify Medical OY · 2025 · 4 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K042783Anspach Minimal Access Spinal Attachment (Masa) SystemThe Anspach Effort, Inc.2004—
+1K113476Anspach Dissection ToolsThe Anspach Effort, Inc.201139
+2K130755Hicut Highspeed InstrumentAdeor Medical AG2014—
+3K232684Surgify HaloSurgify Medical OY2023—
+2K173066OSSEODUO Shaver and Drill SystemBien-Air Surgery SA2018—
+4K250380Surgify Halo (54.085.SHD.H1); Surgify Halo (54.140.SHD.H1); Surgify Halo…Surgify Medical OY20255

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201410
201820
201930
202210
202310
202420
202520
202610
FIG. 1 — Lineage of K042783, 2004–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HBE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HBE device

  1. #1 K251433 Surgify Halo (54.085.SHD.H1); Surgify Halo…score 13.2 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Anspach Angled Micro-nose Attachment
Named without a submission number
—
*—Medtronic (Midas Rex) "Micro-Telescoping System"
Named without a submission number
—

What descends from it? (14)

See all 14

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K113476Anspach Dissection ToolsThe Anspach Effort, Inc.201113

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 6, 2016Class IIZ-2003-2017Under Investigation by firmSupplied Directions for Use (DFU) did not contain a recommended service interval.
Oct 6, 2016Class IIZ-2009-2017Under Investigation by firmSupplied Directions for Use (DFU) did not contain a recommended service interval.
Oct 6, 2016Class IIZ-2004-2017Under Investigation by firmSupplied Directions for Use (DFU) did not contain a recommended service interval.
Oct 6, 2016Class IIZ-2007-2017Under Investigation by firmSupplied Directions for Use (DFU) did not contain a recommended service interval.
Oct 6, 2016Class IIZ-2011-2017Under Investigation by firmSupplied Directions for Use (DFU) did not contain a recommended service interval.
Oct 6, 2016Class IIZ-2005-2017Under Investigation by firmSupplied Directions for Use (DFU) did not contain a recommended service interval.
Oct 6, 2016Class IIZ-2006-2017Under Investigation by firmSupplied Directions for Use (DFU) did not contain a recommended service interval.
Oct 26, 2012Class IIZ-0708-2014Labeling False and MisleadingDuring a recent review of documentation Anspach determined that the current directions for use for the MCA cutting burrs and MCA Micro Curved Attachment do not provide clear details regarding the intended use specifically for the MCA…
Mar 29, 2012Class IIZ-0715-2014Error in labelingCurrent directions for use for the MCA cutting burs (item 18-0001) do not provide enough details regarding the intended use specifically for the MCA burrs.
Mar 29, 2012Class IIZ-0753-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0762-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0763-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0764-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0758-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0759-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0761-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0768-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0750-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0754-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0749-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0751-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0752-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0755-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0756-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0757-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0760-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0765-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0766-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0767-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Mar 29, 2012Class IIZ-0769-2014Labeling Change ControlThe Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA…
Oct 13, 2010Not classifiedZ-2903-2011Nonconforming Material/ComponentOn July 30th, 2010 the Anspach Effort, Inc. initiated a voluntary recall on their Anspach Single Use, Sterile Bone Cutting Burrs. A complaint was recieved stating that the product had failed during a procedure.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K113476 (The Anspach Effort, Inc., 2011) has gone furthest, with 13 descendants of its own.

How does its category compare?

All HBE devices

HBE Drills, Burrs, Trephines & Accessories (Simple, Powered): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · HBE

6in 2025

2007: 22008: 22011: 22012: 12013: 12014: 22015: 32016: 22017: 22018: 22019: 52021: 62022: 22023: 22025: 62026 (partial year): 1
2007high 6 (2021) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HBE

30in 2025

2015: 155 days median2019: 133 days median2021: 193 days median2025: 30 days median
2007high 193 (2021) · low 30 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HBE

1in 2024

2008: 472009: 72010: 232011: 12012: 262013: 42016: 72017: 32021: 12023: 32024: 12026 (partial year): 19
2007high 47 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HBE

1,924in 2025

2007: 142008: 142009: 2422010: 4922011: 1,0132012: 1,1462013: 1,5832014: 5592015: 4832016: 8482017: 8482018: 7872019: 8362020: 1,0252021: 1,0472022: 9002023: 9482024: 1,1532025: 1,924
2007high 1,924 (2025) · low 14 (2007)2025

MAUDE reports naming product code HBE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—014
20082—4714
20090—7242
20100—23492
20112—11,013
20121—261,146
20131—41,583
20142—0559
201531550483
20162—7848
20172—3848
20182—0787
201951330836
20200—01,025
2021619311,047
20222—0900
20232—3948
20240—11,153
202563001,924
20261—19—

What is it indicated for?

The Anspach Minimal Access Spinal Attachment system is indicated for use as a nosepiece attachment with Anspach pneumatic and electric surgical drill systems for surgical cutting, shaping and removal of bones, including bones of the spine and cranium.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K042783: Anspach Minimal Access Spinal Attachment (Masa) System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K042783

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026