510(k) · Cleared
Recalled · Class II · 2017Harvest Graft Delivery Syringe
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Harvest Technologies, Corp. · FMF Syringe, Piston
Cleared on Mar 11, 2005 as substantially equivalent to K003286 (Depuyacromed, 2001) and 3 other predicates after a 107-day review. Its family tree reaches back 3 generations to 2002. 51 later devices descend from it. It has been recalled 2 times.
- Decided
- Mar 11, 2005
- Review
- 107 days
- Class
- Class II
- Product code
- FMF
- Recalls
- 2
- Predicates
- 4
Its longest line of descent runs back 3 generations to K013966 (2002). 8 ancestors in all, the oldest K994337 (2000). 51 later devices descend from it, 4 citing it directly.
- Root · 3 generations backK013966 Apatight-TCP Bone Graft SubstituteCerabio, LLC · 2002
- 2 generations backK022629 WMT-TCP Bone Graft SubstituteWrightmedicaltechnologyinc · 2002
- 1 generation backK031817 Cellplex TCP Synthetic Cancellous BoneWrightmedicaltechnologyinc · 2003
- This deviceK043261 Harvest Graft Delivery SyringeHarvest Technologies, Corp. · 2005
- Descendant · cites it directly · 45 of its ownK100754 Graft Delivery System Model SA-6115Micromedics, Inc. · 2010
- Descendant · 2 generations on · 41 of its ownK111632 Infill Graft Delivery SystemPinnacle Spine Group, LLC · 2011
- Descendant · 3 generations on · 40 of its ownK121476 Infill Graft Delivery SystemPinnacle Spine Group, LLC · 2012
- Descendant · cites it directly · 37 of its ownK062986 Medtronic Graft Delivery SyringeMedtronic Sofamor Danek USA, Inc. · 2006
- Descendant · 2 generations on · 36 of its ownK121124 Arthrex Mixing and Delivery SystemArthrex, Inc. · 2012
Showing 5 of 51 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 51 · See all- K100754Graft Delivery System Model SA-6115Micromedics, Inc. · 2010 · cites it directly45descendants
- K111632Infill Graft Delivery SystemPinnacle Spine Group, LLC · 2011 · 2 generations on41descendants
- K121476Infill Graft Delivery SystemPinnacle Spine Group, LLC · 2012 · 3 generations on40descendants
- K062986Medtronic Graft Delivery SyringeMedtronic Sofamor Danek USA, Inc. · 2006 · cites it directly37descendants
- K121124Arthrex Mixing and Delivery SystemArthrex, Inc. · 2012 · 2 generations on36descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K013966 | Apatight-TCP Bone Graft Substitute | Cerabio, LLC | 2002 | — |
| −2 | K022629 | WMT-TCP Bone Graft Substitute | Wrightmedicaltechnologyinc | 2002 | — |
| −1 | K031817 | Cellplex TCP Synthetic Cancellous Bone | Wrightmedicaltechnologyinc | 2003 | — |
| 0 (this device) | K043261 | Harvest Graft Delivery Syringe | Harvest Technologies, Corp. | 2005 | — |
| +1 | K100754 | Graft Delivery System Model SA-6115 | Micromedics, Inc. | 2010 | — |
| +2 | K111632 | Infill Graft Delivery System | Pinnacle Spine Group, LLC | 2011 | — |
| +3 | K121476 | Infill Graft Delivery System | Pinnacle Spine Group, LLC | 2012 | — |
| +1 | K062986 | Medtronic Graft Delivery Syringe | Medtronic Sofamor Danek USA, Inc. | 2006 | — |
| +2 | K121124 | Arthrex Mixing and Delivery System | Arthrex, Inc. | 2012 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2000 | 1 | 0 |
| 2001 | 1 | 0 |
| 2002 | 1 | 0 |
| 2003 | 1 | 0 |
| 2004 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2006 | 2 | 0 |
| 2008 | 1 | 0 |
| 2010 | 1 | 0 |
| 2011 | 2 | 0 |
| 2012 | 2 | 0 |
| 2014 | 5 | 0 |
| 2015 | 3 | 0 |
| 2016 | 3 | 0 |
| 2017 | 2 | 0 |
| 2018 | 3 | 0 |
| 2019 | 2 | 0 |
| 2020 | 7 | 0 |
| 2021 | 2 | 0 |
| 2022 | 2 | 0 |
| 2023 | 6 | 0 |
| 2024 | 5 | 0 |
| 2025 | 2 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FMF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FMF device
- #1 K241821 Luer Lock Syringe with Safety Needle; Luer…score 16.1 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K003286 | Symphony Graft Delivery SystemSummary page 1…other bone graft delivery systems; e.g., Symphony Graft Delivery System marketed by Depuy Acrorned (K003286), Wright Medical Technology Bone Graft Syringe (K023088) and INFLITRATE Marrow… | 2001 |
| * | K023088 | Bone Graft Syringe, Model 8600-00X0Summary page 1…Delivery System marketed by Depuy Acrorned (K003286), Wright Medical Technology Bone Graft Syringe (K023088) and INFLITRATE Marrow Infusion Chamber (component of K031817), and the… | 2002 |
| * | K031817 | Cellplex TCP Synthetic Cancellous BoneSummary page 1…edical Technology Bone Graft Syringe (K023088) and INFLITRATE Marrow Infusion Chamber (component of K031817), and the IsoTis Ortholliologics Aspirex-IBone Marrow Aspirate… | 2003 |
| * | K041991 | Aspirex - Bone Marrow Aspirate KitSummary page 1…n Chamber (component of K031817), and the IsoTis Ortholliologics Aspirex-IBone Marrow Aspirate Kit (K041991). Device Description: The Harvest Graft Delivery Systerm consists of a Graft S… | 2004 |
What descends from it? (51)
See all 514 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 20, 2017 | Class II | Z-0075-2018 | Packaging | Certain lots of the Harvest Graft Delivery System GDP-10 Procedure Packs may have a lack of a seal on the inner pouch. |
| Sep 11, 2013 | Class II | Z-0022-2014 | Device Design | Potential for leur connector to leak or have cracks. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 3 years and passes through 3 companies; a new generation was cleared every 8 months on median.
- None of the 8 ancestors has a recall.
- Of its descendants, K100754 (Micromedics, Inc., 2010) has gone furthest, with 45 descendants of its own.
- 6 other devices also cite its predicate K003286.
How does its category compare?
All FMF devicesFMF Syringe, Piston: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · FMF
13in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FMF
114in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FMF
19in 2023
Class I (19 since 2007) Class II/III not classified
Adverse event reports per year · FMF
4,181in 2025
MAUDE reports naming product code FMF, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 23 | 90 | 6 | 674 |
| 2008 | 9 | 74 | 11 | 691 |
| 2009 | 14 | 94 | 15 | 913 |
| 2010 | 11 | 130 | 12 | 679 |
| 2011 | 16 | 109 | 3 | 186 |
| 2012 | 9 | 86 | 6 | 307 |
| 2013 | 12 | 154 | 1 | 326 |
| 2014 | 13 | 84 | 1 | 244 |
| 2015 | 11 | 87 | 2 | 273 |
| 2016 | 12 | 146 | 0 | 684 |
| 2017 | 13 | 135 | 3 | 1,858 |
| 2018 | 8 | 87 | 2 | 3,952 |
| 2019 | 11 | 146 | 1 | 5,863 |
| 2020 | 15 | 107 | 2 | 5,334 |
| 2021 | 20 | 175 | 1 | 5,874 |
| 2022 | 24 | 211 | 2 | 5,233 |
| 2023 | 22 | 173 | 19 (18) | 5,214 |
| 2024 | 14 | 130 | 0 | 5,122 |
| 2025 | 13 | 114 | 0 | 4,181 |
| 2026 | 9 | 239 | 2 (1) | — |
What is it indicated for?
The Harvest Graft Delivery System is intended for the aspiration of bone marrow, autologous blood, plasma, or other body fluids. The system is indicated for the delivery of allograft, autograft, or synthetic bone graft materials to an orthopedic surgical site. In addition, it is designed to facilitate pre-mixing of bone graft materials with the aspirate(s), IV. fluids, blood, plasma, platelet rich plasma, bone marrow or other specific blood component(s) as deemed necessary by the clinical use requirements.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K043261 · printed Nov 6, 2026
Cite this page
510(k) Report. “K043261: Harvest Graft Delivery Syringe.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K043261
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026