510(k) Report

510(k) · Cleared

Recalled · 2006

Stratus CS Acute Care NT-Probnp Testpak, PBNP Calpak Calibrator, PBNP Dilpak Diluent

K043476510(k) TraditionalCleared Feb 15, 200561-day reviewClass II21 CFR 862.1117Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Dade Behring, Inc. · NBC Test, Natriuretic Peptide

Cleared on Feb 15, 2005 as substantially equivalent to K032646 (Roche Diagnostics Corp., 2003) and 1 other predicate after a 61-day review. Its family tree reaches back 3 generations to 2000. 2 later devices descend from it. It has been recalled 2 times.

Decided
Feb 15, 2005
Review
61 days
Class
Class II
Product code
NBC
Recalls
2
Predicates
2

Its longest line of descent runs back 3 generations to DEN000010 (2000). 2 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2006: 1 descendant cleared2007: 1 descendant clearedDEN000010 · Triage B-Type Natriuretic Peptide (BNP) Test · Biosite Incorporated · 2000K022516 · Elecsys Probnp Immunoassay · Roche Diagnostics Corp. · 2002K032646 · Elecsys Probnp Immunoassay · Roche Diagnostics Corp. · 2003200020052010201520202025
  1. Root · 3 generations backDEN000010 Triage B-Type Natriuretic Peptide (BNP) TestBiosite Incorporated · 2000
  2. 2 generations backK022516 Elecsys Probnp ImmunoassayRoche Diagnostics Corp. · 2002
  3. 1 generation backK032646 Elecsys Probnp ImmunoassayRoche Diagnostics Corp. · 2003
  4. This deviceK043476 Stratus CS Acute Care NT-Probnp Testpak, PBNP Calpak Calibrator…Dade Behring, Inc. · 2005
  5. Descendant · cites it directly · 1 of its ownK060548 Stratus CS Acute Care Betahcg and NT-Probnp Testpak and Testpak…Dade Behring, Inc. · 2006 · Recalled
  6. Descendant · cites it directly · 0 of its ownK071834 Stratus CS Acute Care NT-Probnp Testpak, Model CpbnpmDade Behring, Inc. · 2007 · Recalled

Most influential descendants

Showing 2 of 2 · See all
  • K060548Stratus CS Acute Care Betahcg and NT-Probnp Testpak and Testpak Assay, Calpak and DilpakDade Behring, Inc. · 2006 · cites it directly1descendant
  • K071834Stratus CS Acute Care NT-Probnp Testpak, Model CpbnpmDade Behring, Inc. · 2007 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3DEN000010Triage B-Type Natriuretic Peptide (BNP) TestBiosite Incorporated2000—
−2K022516Elecsys Probnp ImmunoassayRoche Diagnostics Corp.2002—
−1K032646Elecsys Probnp ImmunoassayRoche Diagnostics Corp.2003—
0 (this device)K043476Stratus CS Acute Care NT-Probnp Testpak, PBNP Calpak Calibrator, PBNP Dilpak…Dade Behring, Inc.2005—
+1K060548Stratus CS Acute Care Betahcg and NT-Probnp Testpak and Testpak Assay, Calpak…Dade Behring, Inc.20062
+1K071834Stratus CS Acute Care NT-Probnp Testpak, Model CpbnpmDade Behring, Inc.20072

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200710
FIG. 1 — Lineage of K043476, 2000–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NBC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NBC device

  1. #1 K262109 Access NT-proBNPscore 13.9 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K032646Elecsys Probnp Immunoassay
Summary page 1
…A Reviewers' - 11/30/2000 Predicate Device(s): Roche Diagnostics Elecsys®)proBNP immunoassay and CalSet calibrator (K032646/K0225 16) Device Description(s): Method The Stratus® CS Acute Care"m NT-proBNP method is a two-site sandwich assay based upo…
Roche Diagnostics Corp.2003—
*K022516Elecsys Probnp Immunoassay
Summary page 1
…ers' - 11/30/2000 Predicate Device(s): Roche Diagnostics Elecsys®)proBNP immunoassay and CalSet calibrator (K032646/K0225 16) Device Description(s): Method The Stratus® CS Acute Care"m NT-proBNP method is a two-site sandwich assay based upon sol…
Roche Diagnostics Corp.2002—

Also cleared against K032646: 7 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K060548Stratus CS Acute Care Betahcg and NT-Probnp Testpak and Testpak Assay, Calpak and DilpakDade Behring, Inc.20061
K071834Stratus CS Acute Care NT-Probnp Testpak, Model CpbnpmDade Behring, Inc.20070

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 3, 2006Not classifiedZ-0879-06OtherFoil sealing of the TestPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the…
Apr 3, 2006Not classifiedZ-0885-06OtherFoil sealing of the DilPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 5 years and passes through 3 companies; a new generation was cleared every 1.3 years on median.
  • None of the 3 ancestors has a recall.
  • Of its descendants, K060548 (Dade Behring, Inc., 2006) has gone furthest, with 1 descendant of its own.
  • 7 other devices also cite its predicate K032646.

How does its category compare?

All NBC devices

NBC Test, Natriuretic Peptide: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · NBC

1in 2025

2007: 22008: 72010: 12020: 12021: 22022: 12023: 22024: 12025: 12026 (partial year): 4
2007high 7 (2008) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NBC

126in 2008

2008: 126 days median2026 (partial year): 94 days median
2007high 126 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NBC

1in 2020

2010: 12012: 22014: 12015: 12018: 12020: 1
2007high 2 (2012) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NBC

17in 2025

2007: 172008: 152009: 352010: 612011: 532012: 232013: 132014: 702015: 432016: 312017: 92018: 72019: 92020: 62022: 42023: 262024: 142025: 17
2007high 70 (2014) · low 0 (2021)2025

MAUDE reports naming product code NBC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—017
20087126015
20090—035
20101—161
20110—053
20120—223
20130—013
20140—170
20150—143
20160—031
20170—09
20180—17
20190—09
20201—16
20212—00
20221—04
20232—026
20241—014
20251—017
20264940—

What is it indicated for?

The Stratus® CS Acute CareTM NT-proBNP method (pBNP) is an in vitro diagnostic assay for the quantitative determination of N-terminal pro-brain natriuretic peptide (NT-proBNP) in human plasma. In individuals suspected of having congestive heart failure (CHF), measurements of NT-proBNP are used as an aid in the diagnosis and assessment of severity. The test is further indicated for the risk stratification of patients with acute coronary syndrome and heart…

Read the full statement (121 words)

The Stratus® CS Acute CareTM NT-proBNP method (pBNP) is an in vitro diagnostic assay for the quantitative determination of N-terminal pro-brain natriuretic peptide (NT-proBNP) in human plasma. In individuals suspected of having congestive heart failure (CHF), measurements of NT-proBNP are used as an aid in the diagnosis and assessment of severity. The test is further indicated for the risk stratification of patients with acute coronary syndrome and heart failure.

The NT-proBNP Calibrator (pBNP CalPak) is an in vitro diagnostic product intended to be used for calibration of the NT-proBNP method on the Stratus® CS analyzer.

The NT-proBNP Dilution Pak (DilPak) is an in vitro diagnostic product intended to be used for dilution of the NT-proBNP Method on the Stratus® CS analyzer.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K043476: Stratus CS Acute Care NT-Probnp Testpak, PBNP Calpak Calibrator, PBNP Dilpak Diluent.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K043476

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026