# K050055: GE Datex-Ohmeda Aestiva/5 Mri Anesthesia System

- Pathway: 510(k) (Special)
- Applicant: Datex-Ohmeda, Inc.
- Decided: Feb 4, 2005 after 24 days
- Product code: BSZ (Gas-Machine, Anesthesia)
- Device class: 2
- Regulation: 21 CFR 868.5160
- Recalls of this device: 2 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
64 later devices descend from it; 4 cite it directly.
26 of 65 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. The FDA summary names its predicate without a number.
- (no number) Datex-Ohmeda Aestiva SmartVent MRI Anesthesia System: named in the summary without a submission number

## Most influential descendants
- K063404: Servo-I Ventilator System, Models 64, 87 and 800 (2007), 61 descendants of its own
- K081842: Engstrom Carestation, Engstrom Pro (2008), 37 descendants of its own
- K073179: Servo-I Ventilator System Model 64 87 800 E4073, Heliox Option, Model 6675585 (2008), 28 descendants of its own
- K093886: Engstrom Carestation, Engstrom Pro (2010), 21 descendants of its own
- K103803: Hamilton-G5 (2011), 14 descendants of its own

## Recalls of this device
- Jul 19, 2019 · Class II · Z-0114-2020: Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opp
- Dec 15, 2015 · Class II · Z-0679-2016: The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unr

## Indications for use
The Aestiva/5 MRI Anesthesia System provides the functional feature set offered by the conventional Aestiva/5 to the clinician with the added ability to be used in the MR environment. Among those standard Aestiva/5 features is the Datex-Ohmeda user interface, all the ventilation parameters of the SmartVent along with the Aestiva breathing circuit. The Aestiva/5 MRI Anesthesia System performed to specifications when tested directly next to 1.5 and 3.0 Tesla shielded MRI devices in a field strength that did not exceed 300 gauss.