510(k) Report

510(k) · Cleared

Recalled · 2011

Arthoralix 8500, Models 110-0199G1, 110-0199G2, 110-0199G3, 110-0199G4

K050255510(k) TraditionalCleared Mar 17, 200542-day reviewClass II21 CFR 872.1800Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Gendex Dental Systems · EHD Unit, X-Ray, Extraoral With Timer

Cleared on Mar 17, 2005 after a 42-day review. The FDA summary names its predicate without a number. 1 later device descends from it. It has been recalled 2 times.

Decided
Mar 17, 2005
Review
42 days
Class
Class II
Product code
EHD
Recalls
2
Predicates
0

1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2009: 1 descendant cleared20052010201520202025
  1. This deviceK050255 Arthoralix 8500, Models 110-0199G1, 110-0199G2, 110-0199G3, 110-0199G4Gendex Dental Systems · 2005
  2. Descendant · cites it directly · 0 of its ownK091866 Progeny Vantage Panoramic X-Ray SystemMidmark · 2009

Most influential descendants

Showing 1 of 1 · See all
  • K091866Progeny Vantage Panoramic X-Ray SystemMidmark · 2009 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K050255Arthoralix 8500, Models 110-0199G1, 110-0199G2, 110-0199G3, 110-0199G4Gendex Dental Systems2005—
+1K091866Progeny Vantage Panoramic X-Ray SystemMidmark2009—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
FIG. 1 — Lineage of K050255, 2005–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in EHD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new EHD device

  1. #1 K260795 Hand-held Dental X-ray System (HyperLight-M)score 14.1 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Gendex Orthoralix 9200
Named without a submission number
—
*—Planmeca DIMAX2
Named without a submission number
—
*—Gendex Orthoralix 9200 DDE
Named without a submission number
—
*—Gendex GX-9000
Named without a submission number
—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K091866Progeny Vantage Panoramic X-Ray SystemMidmark20090

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 4, 2011Not classifiedZ-2040-2011OtherThere is an anomaly in the Orthoralix firmware which can cause a message corruption which will distort the patient image being transferred from the X-Ray system to the PC. This image distortion/corruption (distorted, fragmented, flipped…
Dec 21, 2010Not classifiedZ-1373-2011Labeling Change ControlThe Automatic Exposure Control (AEC) feature located on the control panel of the Orthoralix 8500 Panoramic Dental X-Ray System was functional, but was subsequently disabled in units manufactured starting 5/17/10. North American units…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K091866 (Midmark, 2009) has gone furthest, with 0 descendants of its own.

How does its category compare?

All EHD devices

EHD Unit, X-Ray, Extraoral With Timer: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · EHD

1in 2025

2007: 12008: 52009: 42010: 42011: 32012: 22013: 42014: 32015: 12016: 22017: 42018: 52019: 42020: 22021: 12022: 82023: 82024: 72025: 12026 (partial year): 6
2007high 8 (2022) · low 1 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · EHD

58in 2024

2008: 49 days median2009: 92 days median2010: 268 days median2011: 102 days median2013: 216 days median2014: 107 days median2017: 28 days median2018: 30 days median2019: 173 days median2022: 58 days median2023: 147 days median2024: 58 days median2026 (partial year): 128 days median
2007high 268 (2010) · low 28 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · EHD

1in 2017

2010: 32011: 32014: 12016: 12017: 1
2007high 3 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · EHD

6in 2024

2007: 52008: 62009: 382010: 472011: 262012: 222013: 212014: 202015: 1102016: 632017: 662018: 212019: 212020: 42021: 42022: 42023: 22024: 6
2007high 110 (2015) · low 0 (2025)2025

MAUDE reports naming product code EHD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—05
200854906
2009492038
20104268347
20113102326
20122—022
20134216021
20143107120
20151—0110
20162—163
2017428166
2018530021
20194173021
20202—04
20211—04
202285804
2023814702
202475806
20251—00
202661280—

What is it indicated for?

The Orthoralix 8500 Panoramic Dental X-Ray System is to be used as an extraoral source of x-rays for imaging of the dento-maxilofacial area.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K050255: Arthoralix 8500, Models 110-0199G1, 110-0199G2, 110-0199G3, 110-0199G4.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K050255

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026