510(k) · Cleared
Recalled · 2012Millennium Multi-Leaf Collimator
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Varian Medical Systems · IYE Accelerator, Linear, Medical
Cleared on Mar 23, 2005 as substantially equivalent to K990085 (Varian Assoc., Inc., 1999) after a 29-day review. Its family tree reaches back 3 generations to 1995. 67 later devices descend from it. It has been recalled 2 times.
- Decided
- Mar 23, 2005
- Review
- 29 days
- Class
- Class II
- Product code
- IYE
- Recalls
- 2
- Predicates
- 1
Its longest line of descent runs back 3 generations to K926449 (1995). 5 ancestors in all. 67 later devices descend from it, 3 citing it directly.
- Root · 3 generations backK926449 Varian Multileaf Collimator, Release 4.0Varian Assoc., Inc. · 1995
- 2 generations backK973889 Varian Multileaf Collimator with Dynamic Arc Therapy FeatureVarian Medical Systems, Inc. · 1997
- 1 generation backK990085 Millennium MLC -120Varian Assoc., Inc. · 1999 · Recalled
- This deviceK050442 Millennium Multi-Leaf CollimatorVarian Medical Systems · 2005
- Descendant · cites it directly · 64 of its ownK071992 High-Definition 120 Multileaf Collimator (HD120 MLC)Varian Medical Systems · 2007 · Recalled
- Descendant · cites it directly · 61 of its ownK092871 Trilogy MXVarian Medical Systems, Inc. · 2009 · Recalled
- Descendant · 2 generations on · 43 of its ownK111106 TruebeamVarian Medical Systems, Inc. · 2011 · Recalled
- Descendant · 3 generations on · 37 of its ownK123291 TruebeamVarian Medical Systems, Inc. · 2012
- Descendant · 3 generations on · 36 of its ownK123808 AgilityElekta, Ltd. · 2013 · Recalled
Showing 5 of 67 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 67 · See all- K071992High-Definition 120 Multileaf Collimator (HD120 MLC)Varian Medical Systems · 2007 · cites it directly64descendants
- K092871Trilogy MXVarian Medical Systems, Inc. · 2009 · cites it directly61descendants
- K111106TruebeamVarian Medical Systems, Inc. · 2011 · 2 generations on43descendants
- K123291TruebeamVarian Medical Systems, Inc. · 2012 · 3 generations on37descendants
- K123808AgilityElekta, Ltd. · 2013 · 3 generations on36descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K926449 | Varian Multileaf Collimator, Release 4.0 | Varian Assoc., Inc. | 1995 | — |
| −2 | K973889 | Varian Multileaf Collimator with Dynamic Arc Therapy Feature | Varian Medical Systems, Inc. | 1997 | — |
| −1 | K990085 | Millennium MLC -120 | Varian Assoc., Inc. | 1999 | 2 |
| 0 (this device) | K050442 | Millennium Multi-Leaf Collimator | Varian Medical Systems | 2005 | — |
| +1 | K071992 | High-Definition 120 Multileaf Collimator (HD120 MLC) | Varian Medical Systems | 2007 | 4 |
| +1 | K092871 | Trilogy MX | Varian Medical Systems, Inc. | 2009 | 5 |
| +2 | K111106 | Truebeam | Varian Medical Systems, Inc. | 2011 | 5 |
| +3 | K123291 | Truebeam | Varian Medical Systems, Inc. | 2012 | — |
| +3 | K123808 | Agility | Elekta, Ltd. | 2013 | 2 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1996 | 1 | 0 |
| 1997 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2007 | 1 | 0 |
| 2009 | 2 | 0 |
| 2011 | 2 | 0 |
| 2012 | 2 | 0 |
| 2013 | 2 | 0 |
| 2014 | 3 | 0 |
| 2015 | 4 | 0 |
| 2016 | 2 | 0 |
| 2017 | 7 | 0 |
| 2018 | 6 | 0 |
| 2019 | 5 | 0 |
| 2020 | 2 | 0 |
| 2021 | 7 | 0 |
| 2022 | 4 | 0 |
| 2023 | 9 | 0 |
| 2024 | 2 | 0 |
| 2025 | 2 | 0 |
| 2026 | 5 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new IYE device
- #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K990085 | Millennium MLC -120Summary page 1…Product Code: 90 LYE 2. Predicate Devices to claim substantial equivalence: a. Varian Medical Systems' Millennium Multi-leaf Collimator (K990085) 3. Device Description: Millennium Multi-Leaf Collimator (MLC) Release v7.0 is a modification of… | 1999 |
What descends from it? (67)
See all 673 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jan 16, 2012 | Not classified | Z-1091-2012 | Device Design | When using the Standard and Millennium 52-leaf MLC multi-leaf collimators for beam shaping, under specific treatment conditions, dose may be delivered outside of the intended treatment field that exceed the expected transmission dose. |
| Jan 16, 2012 | Not classified | Z-1090-2012 | Device Design | When using the Standard and Millennium 52-leaf MLC multi-leaf collimators for beam shaping, under specific treatment conditions, dose may be delivered outside of the intended treatment field that exceed the expected transmission dose. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 10 years and passes through 3 companies; a new generation was cleared every 2.5 years on median.
- 3 of the 5 ancestors have a recall of any class.
- Of its descendants, K071992 (Varian Medical Systems, 2007) has gone furthest, with 64 descendants of its own.
- 1 other device also cites its predicate K990085.
How does its category compare?
All IYE devicesIYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · IYE
12in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · IYE
170in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · IYE
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · IYE
101in 2025
MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 23 | 44 | 78 | 155 |
| 2008 | 14 | 69 | 49 | 114 |
| 2009 | 20 | 63 | 22 | 174 |
| 2010 | 23 | 74 | 27 | 246 |
| 2011 | 11 | 94 | 73 | 280 |
| 2012 | 28 | 106 | 20 | 128 |
| 2013 | 18 | 88 | 11 | 96 |
| 2014 | 20 | 100 | 14 | 124 |
| 2015 | 20 | 108 | 8 | 100 |
| 2016 | 21 | 97 | 7 | 97 |
| 2017 | 23 | 94 | 7 | 125 |
| 2018 | 21 | 58 | 7 | 85 |
| 2019 | 19 | 83 | 2 | 91 |
| 2020 | 20 | 67 | 4 | 74 |
| 2021 | 19 | 86 | 5 | 68 |
| 2022 | 12 | 172 | 3 | 66 |
| 2023 | 18 | 129 | 7 | 128 |
| 2024 | 9 | 239 | 1 | 81 |
| 2025 | 12 | 170 | 1 | 101 |
| 2026 | 14 | 120 | 4 | — |
What is it indicated for?
The Varian Millennium 120 Multi-leaf Collimator (MLC) is provided to assist the radiation oncologist in the delivery of radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation. In static mode, the MLC performs the same function as the customized shadow blocks. In a dynamic mode, a series of MLC leaf positions can be indexed to either Clinac dose fraction or gantry angle to create a changing beam shape while the radiation beam is on to create a three dimensional dose distribution.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K050442 · printed Nov 3, 2026
Cite this page
510(k) Report. “K050442: Millennium Multi-Leaf Collimator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K050442
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026