510(k) Report

510(k) · Cleared

Recalled · 2010

Zimmer Trabecular Metal Humeral Stem, Sizes 6MM and 8MM

K050761510(k) SpecialCleared Apr 11, 200518-day reviewClass II21 CFR 888.3650Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zimmer, Inc. · KWT Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented

Cleared on Apr 11, 2005 as substantially equivalent to K041549 (Zimmer, Inc., 2004) after an 18-day review. Its family tree reaches back 3 generations to 1979. It has been recalled 6 times.

Decided
Apr 11, 2005
Review
18 days
Class
Class II
Product code
KWT
Recalls
6
Predicates
1

Its longest line of descent runs back 3 generations to K790987 (1979). 14 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 1999: 1 other ancestor1996: 4 other ancestors1995: 1 other ancestor1993: 1 other ancestor1992: 1 other ancestor1989: 1 other ancestor1987: 1 other ancestor1986: 1 other ancestorK790987 · Zimmer R Total Shoulder II · Zimmer, Inc. · 1979K982981 · New Zimmer Shoulder System · Zimmer, Inc. · 1998 · recalledK041549 · Zimmer Trabecular Metal Humeral Stem · Zimmer, Inc. · 2004 · recalled1975198019851990199520002005
Other ancestors, by generation 2 generations back: 1 other ancestor3 generations back: 10 other ancestors K041549 · Zimmer Trabecular Metal Humeral Stem · Zimmer, Inc. · 2004 · recalledK982981 · New Zimmer Shoulder System · Zimmer, Inc. · 1998 · recalledK790987 · Zimmer R Total Shoulder II · Zimmer, Inc. · 19790
  1. Root · 3 generations backK790987 Zimmer R Total Shoulder IIZimmer, Inc. · 1979
  2. 2 generations backK982981 New Zimmer Shoulder SystemZimmer, Inc. · 1998 · Recalled
  3. 1 generation backK041549 Zimmer Trabecular Metal Humeral StemZimmer, Inc. · 2004 · Recalled
  4. This deviceK050761 Zimmer Trabecular Metal Humeral Stem, Sizes 6MM and 8MMZimmer, Inc. · 2005
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K790987Zimmer R Total Shoulder IIZimmer, Inc.1979—
−2K982981New Zimmer Shoulder SystemZimmer, Inc.19981
−1K041549Zimmer Trabecular Metal Humeral StemZimmer, Inc.20048
0 (this device)K050761Zimmer Trabecular Metal Humeral Stem, Sizes 6MM and 8MMZimmer, Inc.2005—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198710
198910
199210
199310
199510
199640
199910
FIG. 1 — Lineage of K050761, 1979–2005Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KWT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KWT device

  1. #1 K231657 AGILON® XO Shoulder Replacement Systemscore 17.7 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041549Zimmer Trabecular Metal Humeral Stem
Our reading of the summary gave K041459, FDA’s number for Lifepak 12 Defibrillator/Monitor (Medtronic Emergency Response Systems, Inc.). Corrected to the number with two digits swapped whose FDA name carries the summary’s words for the device, “Zimmer TrabecularMetal Humeral Stem”.
Zimmer, Inc.20048

Also cleared against K041549: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 10, 2010Not classifiedZ-0570-2011Device DesignThe results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating of the implant…
Sep 10, 2010Not classifiedZ-0574-2011Device DesignThe results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating of the implant…
Sep 10, 2010Not classifiedZ-0575-2011Device DesignThe results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating of the implant…
Sep 10, 2010Not classifiedZ-0571-2011Device DesignThe results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating of the implant…
Sep 10, 2010Not classifiedZ-0573-2011Device DesignThe results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating of the implant…
Sep 10, 2010Not classifiedZ-0576-2011Device DesignThe results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating of the implant…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 1 company; a new generation was cleared every 5.7 years on median.
  • 5 of the 14 ancestors have a recall of any class.
  • 1 other device also cites its predicate K041549.

How does its category compare?

All KWT devices

KWT Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KWT

4in 2025

2007: 12010: 12011: 52012: 12013: 12014: 12015: 22016: 22017: 12018: 22019: 42020: 12021: 12022: 12023: 12024: 12025: 42026 (partial year): 1
2007high 5 (2011) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KWT

242in 2025

2011: 165 days median2019: 91 days median2025: 242 days median
2007high 242 (2025) · low 91 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KWT

1in 2023

2008: 12010: 72012: 22013: 12016: 12017: 12018: 12019: 32023: 1
2007high 7 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KWT

749in 2025

2007: 322008: 482009: 582010: 712011: 1122012: 832013: 1132014: 1682015: 1762016: 4452017: 5602018: 4802019: 4192020: 2152021: 2282022: 2732023: 3612024: 6012025: 749
2007high 749 (2025) · low 32 (2007)2025

MAUDE reports naming product code KWT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—032
20080—148
20090—058
20101—771
201151650112
20121—283
20131—1113
20141—0168
20152—0176
20162—1445
20171—1560
20182—1480
20194913419
20201—0215
20211—0228
20221—0273
20231—1361
20241—0601
202542420749
20261—0—

What is it indicated for?

Prosthetic replacement with this device may be indicated for the treatment of severe pain or significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint; ununited humeral head fractures of long duration; irreducible 3-and 4-part proximal humeral fractures; avascular necrosis of the humeral head; or other difficult clinical management problems where arthrodesis or resectional arthoplasty is not acceptable. The…

Read the full statement (109 words)

Prosthetic replacement with this device may be indicated for the treatment of severe pain or significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint; ununited humeral head fractures of long duration; irreducible 3-and 4-part proximal humeral fractures; avascular necrosis of the humeral head; or other difficult clinical management problems where arthrodesis or resectional arthoplasty is not acceptable. The assembled humeral component may be used alone for hemiarthroplasty or combined with the glenoid component for total shoulder arthroplasty. Humeral heads with heights of 27mm or greater may be used for difficult clinical management problems involving rotator cuff deficiency where arthrodesis or conventional nonconstrained arthroplasty is not acceptable.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K050761: Zimmer Trabecular Metal Humeral Stem, Sizes 6MM and 8MM.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K050761

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026