510(k) · Cleared
Recalled · 2011Multiple Analyte Urine Drugs of Abuse Calibrators and Controls
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Lin-Zhi International, Inc. · DKB Calibrators, Drug Mixture
Cleared on Jun 24, 2005 as substantially equivalent to K993755 (Syva Co., 1999) and 1 other predicate after a 57-day review. Its family tree reaches back 1 generation to 1998. 6 later devices descend from it. It has been recalled 3 times.
- Decided
- Jun 24, 2005
- Review
- 57 days
- Class
- Class II
- Product code
- DKB
- Recalls
- 3
- Predicates
- 2
Its longest line of descent runs back 1 generation to K983159 (1998). 2 ancestors in all. 6 later devices descend from it, 6 citing it directly.
- Root · 1 generation backK983159 Drugs of Abuse Urine Calibrators and ControlsDiagnostic Reagents, Inc. · 1998
- This deviceK051088 Multiple Analyte Urine Drugs of Abuse Calibrators and ControlsLin-Zhi International, Inc. · 2005
- Descendant · cites it directly · 0 of its ownK153693 Immunalysis Methamphetamine Urine Enzyme Immunoassay, Immunalysis…Immunalysis Corporation · 2016
- Descendant · cites it directly · 0 of its ownK151771 Immunalysis Benzodiazepines Urine Enzyme Immunoassay, Immunalysis…Immunalysis Corporation · 2016
- Descendant · cites it directly · 0 of its ownK152176 Immunalysis PCP Urine Enzyme Immunoassay, Immunalysis Multi-Drug…Immunalysis Corporation · 2015
- Descendant · cites it directly · 0 of its ownK143502 Immunalysis Opiates Urine Enzyme Immunoassay, Immunalysis Opiates…Immunalysis Corporation · 2015
- Descendant · cites it directly · 0 of its ownK133710 Lzi Multiple Analyte Set a;B;C Drugs of Abuse ControlsLin-Zhi International, Inc. · 2014
Showing 5 of 6 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 6 · See all- K153693Immunalysis Methamphetamine Urine Enzyme Immunoassay, Immunalysis Multi-Drug CalibratorsImmunalysis Corporation · 2016 · cites it directly0descendants
- K151771Immunalysis Benzodiazepines Urine Enzyme Immunoassay, Immunalysis Multi-Drug CalibratorsImmunalysis Corporation · 2016 · cites it directly0descendants
- K152176Immunalysis PCP Urine Enzyme Immunoassay, Immunalysis Multi-Drug CalibratorsImmunalysis Corporation · 2015 · cites it directly0descendants
- K143502Immunalysis Opiates Urine Enzyme Immunoassay, Immunalysis Opiates Urine Calibrators 300…Immunalysis Corporation · 2015 · cites it directly0descendants
- K133710Lzi Multiple Analyte Set a;B;C Drugs of Abuse ControlsLin-Zhi International, Inc. · 2014 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K983159 | Drugs of Abuse Urine Calibrators and Controls | Diagnostic Reagents, Inc. | 1998 | — |
| 0 (this device) | K051088 | Multiple Analyte Urine Drugs of Abuse Calibrators and Controls | Lin-Zhi International, Inc. | 2005 | — |
| +1 | K153693 | Immunalysis Methamphetamine Urine Enzyme Immunoassay, Immunalysis Multi-Drug… | Immunalysis Corporation | 2016 | — |
| +1 | K151771 | Immunalysis Benzodiazepines Urine Enzyme Immunoassay, Immunalysis Multi-Drug… | Immunalysis Corporation | 2016 | — |
| +1 | K152176 | Immunalysis PCP Urine Enzyme Immunoassay, Immunalysis Multi-Drug Calibrators | Immunalysis Corporation | 2015 | — |
| +1 | K143502 | Immunalysis Opiates Urine Enzyme Immunoassay, Immunalysis Opiates Urine… | Immunalysis Corporation | 2015 | — |
| +1 | K133710 | Lzi Multiple Analyte Set a;B;C Drugs of Abuse Controls | Lin-Zhi International, Inc. | 2014 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1999 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2013 | 1 | 0 |
| 2014 | 1 | 0 |
| 2015 | 2 | 0 |
| 2016 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DKB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K993755 | Emit Calibrater/Control Level 0, Level 1, Level 2, Level 3, Level 4, Level 5Summary page 1…brators and Controls are substantially equivalent to the Drugs of Abuse Urine Calibrators and Controls (Dade Behring, Microgenics), cleared under prernarket notification K993755 (Dade Behring), K983 159 (DRI, now Microgenics). 10 Device D… | 1999 |
| * | K983159 | Drugs of Abuse Urine Calibrators and ControlsSummary page 1…substantially equivalent to the Drugs of Abuse Urine Calibrators and Controls (Dade Behring, Microgenics), cleared under prernarket notification K993755 (Dade Behring), K983 159 (DRI, now Microgenics). 10 Device Description All of the DA… | 1998 |
What descends from it? (6)
6 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K153693 | Immunalysis Methamphetamine Urine Enzyme Immunoassay, Immunalysis Multi-Drug Calibrators | 2016 | 0 |
| K151771 | Immunalysis Benzodiazepines Urine Enzyme Immunoassay, Immunalysis Multi-Drug Calibrators | 2016 | 0 |
| K152176 | Immunalysis PCP Urine Enzyme Immunoassay, Immunalysis Multi-Drug Calibrators | 2015 | 0 |
| K143502 | Immunalysis Opiates Urine Enzyme Immunoassay, Immunalysis Opiates Urine Calibrators 300… | 2015 | 0 |
| K133710 | Lzi Multiple Analyte Set a;B;C Drugs of Abuse Controls | 2014 | 0 |
| K132090 | Benzodiazepine Calibrator Set, Benzodiazepine Control Level 1, and Benzodiazepine… | 2013 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 11, 2011 | Not classified | Z-0847-2012 | Labeling mix-ups | Catalog # 0804 Multi-Analyte Urine DAU Intermediate Calibrators were incorrectly labeled and mistakenly filled with Catalog # 0805 DAU High Calibrators. |
| Aug 23, 2011 | Not classified | Z-0868-2012 | Nonconforming Material/Component | Multi Analyte Urine DAU Cutoff Calibrator Gives a decreased absorbance reading due to a loss of the BZO (oxazepam) analyte from the solution and may cause false positive readings. |
| Aug 5, 2011 | Not classified | Z-0838-2012 | Error in labeling | Customer complaint investigation found that the Multi-Analyte Urine DAU Level 2 control was incorrectly labeled and actually contained Multi-Analyte DAU Low Level Calibrators. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All DKB devicesDKB Calibrators, Drug Mixture: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DKB
2in 2014
2026 to Oct 25, outlined.
Median 510(k) review, days · DKB
Too few DKB 510(k)s in any year since 2007 to give a median.
Recalls per year · DKB
2in 2022
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · DKB
1in 2017
MAUDE reports naming product code DKB, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 2 | — | 3 | 0 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 2 | — | 2 | 1 |
| 2012 | 1 | — | 1 | 4 |
| 2013 | 0 | — | 0 | 0 |
| 2014 | 2 | — | 0 | 0 |
| 2015 | 0 | — | 0 | 0 |
| 2016 | 0 | — | 0 | 0 |
| 2017 | 0 | — | 0 | 1 |
| 2018 | 0 | — | 0 | 0 |
| 2019 | 0 | — | 0 | 0 |
| 2020 | 0 | — | 0 | 0 |
| 2021 | 0 | — | 0 | 0 |
| 2022 | 0 | — | 2 | 0 |
| 2023 | 0 | — | 0 | 0 |
| 2024 | 0 | — | 0 | 0 |
| 2025 | 0 | — | 0 | 0 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The Multiple Analyte (Benzoylecgonine, Methamphetamine, Methadone, Morphine, Oxazepam, Secobarbital, Phencyclidine, and Propoxyphene) Urine Drugs of Abuse Calibrators are intended for in vitro diagnostic use for the calibration of their respective enzyme immunoassays to detect d-methainphetamine, benzoylecgonine, opiate, benzodiazepines, barbiturates, methadone, phencyclidine or propoxyphene in human urine. The Multiple Analyte (Benzoylecgonine…
Read the full statement (92 words)
The Multiple Analyte (Benzoylecgonine, Methamphetamine, Methadone, Morphine,
Oxazepam, Secobarbital, Phencyclidine, and Propoxyphene) Urine Drugs of Abuse
Calibrators are intended for in vitro diagnostic use for the calibration of their respective
enzyme immunoassays to detect d-methainphetamine, benzoylecgonine, opiate,
benzodiazepines, barbiturates, methadone, phencyclidine or propoxyphene in human urine.
The Multiple Analyte (Benzoylecgonine, Methamphetamine, Methadone, Morphine,
Oxazepam, Secobarbital, Phencyclidine, and Propoxyphene) Urine Drugs of Abuse Controls
are intended for in vitro diagnostic use for the validation of their respective enzyme
immunoassays to detect d-methamphetamine, benzoylecgonine, opiate, benzodiazepines.
barbiturates, methadone, phencyclidine or propoxyphene in human urine.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K051088 · printed Nov 6, 2026
Cite this page
510(k) Report. “K051088: Multiple Analyte Urine Drugs of Abuse Calibrators and Controls.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051088
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026