510(k) Report

510(k) · Cleared

Recalled · 2006

Rexeed Series Dialyzers

K051187510(k) SpecialCleared Jun 8, 200530-day reviewClass II21 CFR 876.5860Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Asahi Kasei Medical Co., Ltd. · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Cleared on Jun 8, 2005 as substantially equivalent to K001250 (Asahi Medical Co., Ltd., 2000) and 3 other predicates after a 30-day review. Its family tree reaches back 2 generations to 2000. 20 later devices descend from it. It has been recalled once.

Decided
Jun 8, 2005
Review
30 days
Class
Class II
Product code
KDI
Recalls
1
Predicates
4

Its longest line of descent runs back 2 generations to K001250 (2000). 4 ancestors in all. 20 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 2 other ancestors2013: 2 descendants cleared2008: 2 descendants cleared2015: 2 descendants cleared2014: 2 descendants cleared2026: 3 descendants cleared2020: 3 descendants cleared2017: 1 descendant cleared2023: 1 descendant cleared2025: 3 descendants cleared2024: 1 descendant clearedK001250 · Aps Series Dialyzers (Wet Models), Aps Series Dialysis (Dry Models) · Asahi Medical Co., Ltd. · 2000 · recalledK041726 · High Flux Hemodialysis Membrane Dialyzer or High Flux Hollow Fiber Dialyzer, Models Aps-R, -M, -S, -E, and -EX · Asahi Medical Co., Ltd. · 2004 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 4 descendants2 generations forward: 2 descendants3 generations forward: 2 descendants4 generations forward: 4 descendants5 generations forward: 3 descendants6 generations forward: 3 descendants7 generations forward: 2 descendants K041726 · High Flux Hemodialysis Membrane Dialyzer or High Flux Hollow Fiber Dialyzer, Models Aps-R, -M, -S, -E, and -EX · Asahi Medical Co., Ltd. · 2004 · recalledK001250 · Aps Series Dialyzers (Wet Models), Aps Series Dialysis (Dry Models) · Asahi Medical Co., Ltd. · 2000 · recalled0+5
  1. Root · 2 generations backK001250 Aps Series Dialyzers (Wet Models), Aps Series Dialysis (Dry Models)Asahi Medical Co., Ltd. · 2000 · Recalled
  2. 1 generation backK041726 High Flux Hemodialysis Membrane Dialyzer or High Flux Hollow Fiber…Asahi Medical Co., Ltd. · 2004 · Recalled
  3. This deviceK051187 Rexeed Series DialyzersAsahi Kasei Medical Co., Ltd. · 2005
  4. Descendant · cites it directly · 16 of its ownK082414 Optiflux F250NR, Model 0500325EFresenius Medical Care North America · 2008 · Recalled
  5. Descendant · 2 generations on · 13 of its ownK123262 Optiflux Ultra DialyzerFresenius Medical Care North America · 2014
  6. Descendant · 3 generations on · 12 of its ownK152367 Optiflux dialyzerFresenius Medical Care North America · 2015 · Recalled
  7. Descendant · 4 generations on · 9 of its ownK162488 Optiflux dialyzerFresenius Medical Care North America · 2017 · Recalled
  8. Descendant · 5 generations on · 6 of its ownK220721 FX CorAL 60, FX CorAL 80, FX CorAL 100, FX CorAL 120, FX CorAL 600…Fresenius Medical Care Renal Therapies Group, LLC · 2023

Showing 5 of 20 descendants. See all descendants

Descendants cleared per year

2008: 22013: 22014: 22015: 22017: 12020: 32023: 12024: 12025: 32026: 3
2008high 3 (2020) · low 0 (2009)2026

Most influential descendants

Showing 5 of 20 · See all
  • K082414Optiflux F250NR, Model 0500325EFresenius Medical Care North America · 2008 · cites it directly16descendants
  • K123262Optiflux Ultra DialyzerFresenius Medical Care North America · 2014 · 2 generations on13descendants
  • K152367Optiflux dialyzerFresenius Medical Care North America · 2015 · 3 generations on12descendants
  • K162488Optiflux dialyzerFresenius Medical Care North America · 2017 · 4 generations on9descendants
  • K220721FX CorAL 60, FX CorAL 80, FX CorAL 100, FX CorAL 120, FX CorAL 600, FX CorAL 800, FX…Fresenius Medical Care Renal Therapies Group, LLC · 2023 · 5 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K001250Aps Series Dialyzers (Wet Models), Aps Series Dialysis (Dry Models)Asahi Medical Co., Ltd.20002
−1K041726High Flux Hemodialysis Membrane Dialyzer or High Flux Hollow Fiber Dialyzer…Asahi Medical Co., Ltd.20041
0 (this device)K051187Rexeed Series DialyzersAsahi Kasei Medical Co., Ltd.2005—
+1K082414Optiflux F250NR, Model 0500325EFresenius Medical Care North America20083
+2K123262Optiflux Ultra DialyzerFresenius Medical Care North America2014—
+3K152367Optiflux dialyzerFresenius Medical Care North America20151
+4K162488Optiflux dialyzerFresenius Medical Care North America20174
+5K220721FX CorAL 60, FX CorAL 80, FX CorAL 100, FX CorAL 120, FX CorAL 600, FX CorAL…Fresenius Medical Care Renal Therapies Group, LLC2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200120

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200820
201320
201420
201520
201710
202030
202310
202410
202530
202630
FIG. 1 — Lineage of K051187, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KDI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KDI device

  1. #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K001250Aps Series Dialyzers (Wet Models), Aps Series Dialysis (Dry Models)
Summary page 1
…fibers to the crimped was implemented. The REXEED dialyzers with crimped hollow fibers is added to the existing APS Series Dialyzers (hereafter called APS) cleared under K001250 and; (2) the REXEED dialyzers membrane surface area ranges from 1.5m 2 to 2.5m 2, while that of APS ranges from 0.8 m 2 to 2.1 m…
Asahi Medical Co., Ltd.20002
*K041726High Flux Hemodialysis Membrane Dialyzer or High Flux Hollow Fiber Dialyzer, Models…
Summary page 1
…e or chronic renal failure. REXEED-U/S are intended for single use and REXEED-UR/R for reuse. This is the same intended use as the existing APS cleared under K001250 and K041726, with the excwtion of using mixture of hydrogen peroxide and peroxyacetic acid (e.g., Renalin ) as reprocessing agent. REXEED…
Asahi Medical Co., Ltd.20041
*K003498Fresenius Optiflux 200A
Summary page 2
…nciple of operation, technology, materials, specifications, and performance to the existing APS cleared under K001250 and K041726, Fresenius Optiflux F200A cleared under K003498, and Gambro Polyflux 24S and Polyflux 24R cleared under K010667, with the exception of using mixture of hydrogen peroxide and per…
Fresenius Medical Care North America20012
*K010667Gambro Polyflux 24S, Gambro Polyflux 24R
Summary page 2
…performance to the existing APS cleared under K001250 and K041726, Fresenius Optiflux F200A cleared under K003498, and Gambro Polyflux 24S and Polyflux 24R cleared under K010667, with the exception of using mixture of hydrogen peroxide and peroxyacetic acid (e.g., Renalin®). REXEED Series Dialyzers…
Gambro Renal Products2001—

Also cleared against K001250: 7 other devices →

What descends from it? (20)

See all 20

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K082414Optiflux F250NR, Model 0500325EFresenius Medical Care North America200816
K082515Asahi Rexeed-SX/LX Dialyzers, Models: 13SX, 15SX, 18SX, 21SX, 25SX, 13LX, 15LX, 18LX, 21LXAsahi Kasei Medical Co., Ltd.20081
K153344Asahi REXEED-S Series DialyzerAsahi Kasei Medical Co., Ltd.20150
K121409Asahi Rexeed-SX/LX Series DialyzerAsahi Kasei Medical Co., Ltd.20130

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 12, 2006Not classifiedZ-0172-2007OtherThere have been reports of difficulty separating the Gambro Cartridge Blood Set from Asahi Reuse/Multiple Use Dialyzers at the end of treatment, possibly damaging the dialyzer nozzle.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 5 years and passes through 2 companies; a new generation was cleared every 3.9 years on median.
  • 3 of the 4 ancestors have a recall of any class.
  • Of its descendants, K082414 (Fresenius Medical Care North America, 2008) has gone furthest, with 16 descendants of its own.
  • 7 other devices also cite its predicate K001250.

How does its category compare?

All KDI devices

KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KDI

9in 2025

2007: 112008: 62009: 42010: 92011: 52012: 62013: 112014: 82015: 52016: 52017: 82018: 22019: 42020: 92021: 22022: 72023: 72024: 72025: 92026 (partial year): 6
2007high 11 (2007) · low 2 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KDI

199in 2025

2007: 91 days median2008: 148 days median2009: 236 days median2010: 186 days median2011: 85 days median2012: 153 days median2013: 247 days median2014: 164 days median2015: 69 days median2016: 230 days median2017: 188 days median2019: 263 days median2020: 266 days median2022: 259 days median2023: 157 days median2024: 235 days median2025: 199 days median2026 (partial year): 154 days median
2007high 266 (2020) · low 69 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KDI

2in 2025

2007: 42009: 32010: 12011: 42012: 12013: 22014: 32015: 52016: 42017: 12018: 12019: 62021: 22022: 172023: 8, 8 Class I2024: 4, 2 Class I2025: 22026 (partial year): 2
2007high 17 (2022) · low 0 (2008)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · KDI

2,786in 2025

2007: 9162008: 1,0422009: 5222010: 8232011: 2,2702012: 1,7652013: 2,3602014: 6,1542015: 4,1822016: 7662017: 9632018: 1,7162019: 1,5872020: 1,6332021: 1,9442022: 2,1802023: 2,6922024: 2,4412025: 2,786
2007high 6,154 (2014) · low 522 (2009)2025

MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711914916
2008614801,042
200942363522
201091861823
201158542,270
2012615311,765
20131124722,360
2014816436,154
201556954,182
201652304766
201781881963
20182—11,716
2019426361,587
2020926601,633
20212—21,944
20227259172,180
202371578 (8)2,692
202472354 (2)2,441
2025919922,786
202661542—

What is it indicated for?

A. REXEED Series Dialyzers is intended for use for hemodialysis treatment of patients who have chronic or acute renal failure. B. REXEED Series Dialyzers must be used in accordance with the instructions for a physician familiar with hemodialysis and familiar with the conditions of the patient. C. REXEED Series Dialyzers have been tested in-vitro and under single or initial use conditions and REXEED-UR/R Series Dialyzers under reprocessing and reuse…

Read the full statement (298 words)

A. REXEED Series Dialyzers is intended for use for hemodialysis treatment of patients who have chronic or acute renal failure.

B. REXEED Series Dialyzers must be used in accordance with the instructions for a physician familiar with hemodialysis and familiar with the conditions of the patient.

C. REXEED Series Dialyzers have been tested in-vitro and under single or initial use conditions and REXEED-UR/R Series Dialyzers under reprocessing and reuse conditions for up to 15t use cycles. Based on the results from these evaluations, REXEED-UR/R Series Dialyzers may be reprocessed for reuse on the same patient. If reprocessing and reuse is practiced, it is recommended that the reuse be done under as conditions as existed in the in vitro and confirmatory clinical studies recommended immediately below. It is noted that REXEED-UR/R Series Dialyzers have not been tested for reuse when reprocessed with agents and/or processes other than these, and the performance of the dialyzers under other conditions are not known and cannot be recommended. Accordingly:

(1) The reprocessed dialyzer may be used only if the residual Total Cell Volume (TCV) is at least 80% of the original TCV and if such dialyzer otherwise meets the acceptance criteria of these instruction for use and the instruction of the reprocessing system utilized. Furthermore, the policies, instructions, and criteria of the institution for reuse (e.g., concerning dialyzer performance, residual blood, and/or dialyzer leakage or damage) should be followed.

(2) The reprocessing agent may mixture of hydrogen peroxide and peroxyacetic acid (e.g., Renalin®) in conjunction with the Renatron® Dialyzer Reprocessing System (RS 8300); manufactured by Renal Systems, Inc.

(3) The instructions provided by the manufacturer of the chosen reprocessing agent must be followed in reprocessing the dialyzer.

(4) The reprocessed dialyzer may be used only on dialysis systems equipped with volumetric ultrafiltration controllers.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051187: Rexeed Series Dialyzers.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051187

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026