510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Grafton DBM

K051195510(k) TraditionalCleared Dec 16, 2005220-day reviewClass II21 CFR 888.3045Recalls 30

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Osteotech, Inc. · MBP Filler, Bone Void, Osteoinduction (W/O Human Growth Factor)

Cleared on Dec 16, 2005 after a 220-day review. The FDA summary names no predicate by number. 140 later devices descend from it. It has been recalled 30 times.

Decided
Dec 16, 2005
Review
220 days
Class
Class II
Product code
MBP
Recalls
30
Predicates
0

140 later devices descend from it, 11 citing it directly.

Descendants, by year cleared 2006: 2 descendants cleared2011: 11 descendants cleared2007: 3 descendants cleared2018: 7 descendants cleared2009: 9 descendants cleared2013: 7 descendants cleared2008: 7 descendants cleared2017: 6 descendants cleared2010: 6 descendants cleared2015: 8 descendants cleared2024: 12 descendants cleared2012: 2 descendants cleared2021: 3 descendants cleared2016: 3 descendants cleared2014: 10 descendants cleared2020: 3 descendants cleared2023: 11 descendants cleared2025: 14 descendants cleared2022: 5 descendants cleared2019: 4 descendants cleared2026: 7 descendants cleared20052010201520202025
  1. This deviceK051195 Grafton DBMOsteotech, Inc. · 2005
  2. Descendant · cites it directly · 120 of its ownK060728 Novabone Putty - Bioactive Synthetic GraftNovabone Products, LLC · 2006
  3. Descendant · 2 generations on · 71 of its ownK080009 Novabone Putty- Bioactive Synthetic GraftNovabone Products, LLC · 2008
  4. Descendant · 2 generations on · 59 of its ownK082672 Novabone Putty-Bioactive Synthetic Bone GraftNovabone Products, LLC · 2008
  5. Descendant · 2 generations on · 51 of its ownK072184 Vitoss Bioactive Foam Bone Graft SubstituteOrthovita, Inc. · 2007
  6. Descendant · 3 generations on · 50 of its ownK081439 Vitoss Bioactive Foam Pack Bone Graft SubstituteOrthovita, Inc. · 2008

Showing 5 of 140 descendants. See all descendants

Descendants cleared per year

2006: 22007: 32008: 72009: 92010: 62011: 112012: 22013: 72014: 102015: 82016: 32017: 62018: 72019: 42020: 32021: 32022: 52023: 112024: 122025: 142026: 7
2006high 14 (2025) · low 2 (2006)2026

Most influential descendants

Showing 5 of 140 · See all
  • K060728Novabone Putty - Bioactive Synthetic GraftNovabone Products, LLC · 2006 · cites it directly120descendants
  • K080009Novabone Putty- Bioactive Synthetic GraftNovabone Products, LLC · 2008 · 2 generations on71descendants
  • K082672Novabone Putty-Bioactive Synthetic Bone GraftNovabone Products, LLC · 2008 · 2 generations on59descendants
  • K072184Vitoss Bioactive Foam Bone Graft SubstituteOrthovita, Inc. · 2007 · 2 generations on51descendants
  • K081439Vitoss Bioactive Foam Pack Bone Graft SubstituteOrthovita, Inc. · 2008 · 3 generations on50descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K051195Grafton DBMOsteotech, Inc.2005—
+1K060728Novabone Putty - Bioactive Synthetic GraftNovabone Products, LLC2006—
+2K080009Novabone Putty- Bioactive Synthetic GraftNovabone Products, LLC2008—
+2K082672Novabone Putty-Bioactive Synthetic Bone GraftNovabone Products, LLC2008—
+2K072184Vitoss Bioactive Foam Bone Graft SubstituteOrthovita, Inc.2007—
+3K081439Vitoss Bioactive Foam Pack Bone Graft SubstituteOrthovita, Inc.2008—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200620
200730
200870
200990
201060
2011110
201220
201370
2014100
201580
201630
201760
201870
201940
202030
202130
202250
2023110
2024120
2025140
202670
FIG. 1 — Lineage of K051195, 2005–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MBP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MBP device

  1. #1 K242799 PliaFX Floscore 9.0 · 3 generations · no Class I recall
Get a 510(k) Report

What was it cleared against?

The FDA summary names no predicate by number.

What descends from it? (140)

See all 140

11 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K060728Novabone Putty - Bioactive Synthetic GraftNovabone Products, LLC2006120
K123691Magnifuse Bone Graft (Cleared as Grafton II Edbm)Medtronic Sofamor Danek, Inc.20135
K091321Osteoselect Demineralized Bone Matrix PuttyBacterin International, Inc.20092
K091193Accell EVO3CIsotis Orthobiologics, Inc.20092
K063549Novabone Dental Putty-Bioactive Synthetic Bone GraftNovabone Products, LLC20072
K061668Optecure +CCCExactech, Inc.20062
K180121Rafugen DBMCellumed Co., Ltd.20180
K103742Accell EVO3 (Formerly Accell A2I)Integra Life Sciences20110
K082002Progenix PlusMedtronic Sofamor Danek20080
K081817Accell TBM-RIsotis Orthobiologics, Inc.20080

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 12, 2026Class IIZ-0138-2027Process change controlMedtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product…
Sep 26, 2023Class IIZ-0161-2024Employee errorSpecific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with…
Sep 26, 2023Class IIZ-0163-2024Employee errorSpecific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with…
Sep 26, 2023Class IIZ-0164-2024Employee errorSpecific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with…
Sep 26, 2023Class IIZ-0165-2024Employee errorSpecific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with…
Sep 26, 2023Class IIZ-0166-2024Employee errorSpecific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with…
Sep 26, 2023Class IIZ-0167-2024Employee errorSpecific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with…
Sep 26, 2023Class IIZ-0162-2024Employee errorSpecific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have been released with…
Feb 8, 2023Class IIZ-1324-2023Process controlPotential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
Feb 8, 2023Class IIZ-1325-2023Process controlPotential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
Feb 8, 2023Class IIZ-1327-2023Process controlPotential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
Feb 8, 2023Class IIZ-1333-2023Process controlPotential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
Feb 8, 2023Class IIZ-1326-2023Process controlPotential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
Feb 8, 2023Class IIZ-1328-2023Process controlPotential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
Feb 8, 2023Class IIZ-1323-2023Process controlPotential for packaging non-conformances directly related to the dual-barrier, sterile pouch packaging system. Non-conforming packaging may lead to a breach in the sterile barrier.
Jan 6, 2014Class IIZ-0977-2014Packaging process controlProduct was labeled as being measured as 2.5 cm x 10 cm, but contained product that measured as 2.5 cm x 5 cm.
Apr 20, 2012Class IIZ-0084-2013Device DesignMultiple lots of GRAFTON and XPANSE , which are used for bone grafting procedures are being recalled because there is a possibility that sterility of the outer surface of the inner pouch may become compromised.
Apr 20, 2012Class IIZ-0087-2013Device DesignMultiple lots of GRAFTON and XPANSE , which are used for bone grafting procedures are being recalled because there is a possibility that sterility of the outer surface of the inner pouch may become compromised.
Apr 20, 2012Class IIZ-0088-2013Device DesignMultiple lots of GRAFTON and XPANSE , which are used for bone grafting procedures are being recalled because there is a possibility that sterility of the outer surface of the inner pouch may become compromised.
Apr 20, 2012Class IIZ-0091-2013Device DesignMultiple lots of GRAFTON and XPANSE , which are used for bone grafting procedures are being recalled because there is a possibility that sterility of the outer surface of the inner pouch may become compromised.
Apr 20, 2012Class IIZ-0092-2013Device DesignMultiple lots of GRAFTON and XPANSE , which are used for bone grafting procedures are being recalled because there is a possibility that sterility of the outer surface of the inner pouch may become compromised.
Apr 20, 2012Class IIZ-0093-2013Device DesignMultiple lots of GRAFTON and XPANSE , which are used for bone grafting procedures are being recalled because there is a possibility that sterility of the outer surface of the inner pouch may become compromised.
Apr 20, 2012Class IIZ-0083-2013Device DesignMultiple lots of GRAFTON and XPANSE , which are used for bone grafting procedures are being recalled because there is a possibility that sterility of the outer surface of the inner pouch may become compromised.
Apr 20, 2012Class IIZ-0089-2013Device DesignMultiple lots of GRAFTON and XPANSE , which are used for bone grafting procedures are being recalled because there is a possibility that sterility of the outer surface of the inner pouch may become compromised.
Apr 20, 2012Class IIZ-0090-2013Device DesignMultiple lots of GRAFTON and XPANSE , which are used for bone grafting procedures are being recalled because there is a possibility that sterility of the outer surface of the inner pouch may become compromised.
Apr 20, 2012Class IIZ-0085-2013Device DesignMultiple lots of GRAFTON and XPANSE , which are used for bone grafting procedures are being recalled because there is a possibility that sterility of the outer surface of the inner pouch may become compromised.
Apr 20, 2012Class IIZ-0086-2013Device DesignMultiple lots of GRAFTON and XPANSE , which are used for bone grafting procedures are being recalled because there is a possibility that sterility of the outer surface of the inner pouch may become compromised.
Apr 20, 2012Class IIZ-0094-2013Device DesignMultiple lots of GRAFTON and XPANSE , which are used for bone grafting procedures are being recalled because there is a possibility that sterility of the outer surface of the inner pouch may become compromised.
Aug 8, 2011Not classifiedZ-0416-2012Material/Component ContaminationAseptic conditions during the production of these products may have been compromised.
Aug 8, 2011Not classifiedZ-0415-2012Material/Component ContaminationAseptic conditions during the production of these products may have been compromised.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K060728 (Novabone Products, LLC, 2006) has gone furthest, with 120 descendants of its own.

How does its category compare?

All MBP devices

MBP Filler, Bone Void, Osteoinduction (W/O Human Growth Factor): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MBP

1in 2024

2007: 52008: 52009: 12011: 12012: 12013: 32015: 12024: 1
2007high 5 (2007) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MBP

181in 2013

2007: 188 days median2008: 93 days median2013: 181 days median
2007high 188 (2007) · low 93 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MBP

7in 2023

2011: 22012: 132014: 12016: 12023: 72026 (partial year): 2
2007high 13 (2012) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MBP

2in 2025

2007: 72008: 152009: 92010: 142011: 182012: 242013: 162014: 152015: 112016: 62017: 52018: 92019: 32020: 62021: 22022: 42023: 62024: 12025: 2
2007high 24 (2012) · low 1 (2024)2025

MAUDE reports naming product code MBP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007518807
2008593015
20091—09
20100—014
20111—218
20121—1324
20133181016
20140—115
20151—011
20160—16
20170—05
20180—09
20190—03
20200—06
20210—02
20220—04
20230—76
20241—01
20250—02
20260—2—

What is it indicated for?

GRAFTON® DBM is intended for use as a bone graft extender, bone graft substitute, and bone void filler in bony voids or gaps of the skeletal system (i.e., spine, pelvis and extremities) not intrinsic to the stability of the bony structure. The voids or gaps may be surgically created defects or defects created by traumatic injury to the bone. GRAFTON® DBM is resorbed/remodeled and is replaced by host bone during the healing process.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051195: Grafton DBM.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051195

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026