510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Philips Heartstart FR2+ Automated External Defibrillator, Models M3860A, M3861A, M3840A, M3841A

K051632510(k) TraditionalCleared Oct 5, 2005107-day reviewClass III21 CFR 870.5310Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Oct 5, 2005 after a 107-day review. The FDA summary names its predicate without a number. 14 later devices descend from it. It has been recalled 7 times.

Decided
Oct 5, 2005
Review
107 days
Class
Class III
Product code
MKJ
Recalls
7
Predicates
0

14 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2011: 1 descendant cleared2008: 1 descendant cleared2012: 4 descendants cleared2013: 1 descendant cleared2014: 4 descendants cleared2018: 1 descendant cleared2021: 1 descendant cleared2024: 1 descendant cleared20052010201520202025
  1. This deviceK051632 Philips Heartstart FR2+ Automated External Defibrillator, Models…Philips Medical Systems · 2005
  2. Descendant · cites it directly · 10 of its ownK072923 Zoll E Series DefibrillatorZoll Medical Corporation, World Wide Headquarters · 2008
  3. Descendant · 2 generations on · 8 of its ownK112432 Zoll X SeriesZoll Medical Corporation, World Wide Headquarters · 2012 · Recalled
  4. Descendant · 3 generations on · 4 of its ownK133269 Zoll X SeriesZoll Medical Corporation, World Wide Headquarters · 2014
  5. Descendant · 3 generations on · 3 of its ownK121367 Zoll Propaq XMZOLL Medical Corporation · 2012 · Recalled
  6. Descendant · 4 generations on · 2 of its ownK180482 ZOLL Propaq MZOLL Medical Corporation · 2018

Showing 5 of 14 descendants. See all descendants

Descendants cleared per year

2008: 12011: 12012: 42013: 12014: 42018: 12021: 12024: 1
2008high 4 (2012) · low 0 (2009)2026

Most influential descendants

Showing 5 of 14 · See all
  • K072923Zoll E Series DefibrillatorZoll Medical Corporation, World Wide Headquarters · 2008 · cites it directly10descendants
  • K112432Zoll X SeriesZoll Medical Corporation, World Wide Headquarters · 2012 · 2 generations on8descendants
  • K133269Zoll X SeriesZoll Medical Corporation, World Wide Headquarters · 2014 · 3 generations on4descendants
  • K121367Zoll Propaq XMZOLL Medical Corporation · 2012 · 3 generations on3descendants
  • K180482ZOLL Propaq MZOLL Medical Corporation · 2018 · 4 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K051632Philips Heartstart FR2+ Automated External Defibrillator, Models M3860A…Philips Medical Systems2005—
+1K072923Zoll E Series DefibrillatorZoll Medical Corporation, World Wide Headquarters2008—
+2K112432Zoll X SeriesZoll Medical Corporation, World Wide Headquarters20122
+3K133269Zoll X SeriesZoll Medical Corporation, World Wide Headquarters2014—
+3K121367Zoll Propaq XMZOLL Medical Corporation20121
+4K180482ZOLL Propaq MZOLL Medical Corporation2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
201110
201240
201310
201440
201810
202110
202410
FIG. 1 — Lineage of K051632, 2005–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MKJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Philips/Heartstream HeartStart FR2+ AED
Named without a submission number
—

What descends from it? (14)

See all 14

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K072923Zoll E Series DefibrillatorZoll Medical Corporation, World Wide Headquarters200810
K111693Philips Heartstart FR3 Automated External DefibrillatorPhilips Medical Systems20112
K120350Extended Life BatteryAed Battery Exchange, LLC20120

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 9, 2021Class IIZ-0078-2022Process controlCustomers were not notified of previous recalls associated with various defibrillator models.
May 28, 2010Not classifiedZ-0470-2011Nonconforming Material/ComponentFailure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable.
May 28, 2010Not classifiedZ-0472-2011Nonconforming Material/ComponentFailure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable.
May 28, 2010Not classifiedZ-0473-2011Nonconforming Material/ComponentFailure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable.
May 28, 2010Not classifiedZ-0471-2011Nonconforming Material/ComponentFailure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable.
Oct 1, 2009Not classifiedZ-0062-2010Nonconforming Material/ComponentA potential exists for a component failure, which, if it occurs, could result in the inability to deliver therapy.
Oct 1, 2009Not classifiedZ-0063-2010Nonconforming Material/ComponentA potential exists for a component failure, which, if it occurs, could result in the inability to deliver therapy.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K072923 (Zoll Medical Corporation, World Wide Headquarters, 2008) has gone furthest, with 10 descendants of its own.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

The HeartStart FR2+ is intended to be used with disposable defibrillation pads applied to a person who is experiencing the symptoms of sudden cardiac arrest (SCA): unresponsiveness and absence of breathing. If in doubt, apply the pads. When the patient is under 8 years or weighs less than 55 pounds (25 kg), apply FR2 infant/child reduced-energy defibrillator pads, if available. WARNING: Performance of the SMART CPR AUTOl and AUTO2 settings has not been…

Read the full statement (199 words)

The HeartStart FR2+ is intended to be used with disposable defibrillation pads applied to a person who is experiencing the symptoms of sudden cardiac arrest (SCA): unresponsiveness and absence of breathing.

If in doubt, apply the pads.

When the patient is under 8 years or weighs less than 55 pounds (25 kg), apply FR2 infant/child reduced-energy defibrillator pads, if available.

WARNING: Performance of the SMART CPR AUTOl and AUTO2 settings has not been established in patients under 8 years or 55 lb (25 kg). See Appendix E for more information.

The HeartStart FR2+ is intended for use by personnel who have been trained in its operation. The user should be qualified by training in basic life support (BLS), advanced life support (ALS), or other physician-authorized emergency medical response.

At the discretion of emergency care personnel, the M3860A FR2+ with ECG display enabled can also be used with the FR2+ ECG assessment module to display the rhythm of a responsive or breathing patient, regardless of age. The FR2+ and ECG assessment module system provides a non-diagnostic display for attended patient monitoring. While connected to the FR2+ ECG assessment module, the FR2+ evaluates the patient's ECG and disables its shock capability.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051632: Philips Heartstart FR2+ Automated External Defibrillator, Models M3860A, M3861A, M3840A, M3841A.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051632

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026