510(k) · Cleared
Recalled · Class II · 2014Datex-Ohmeda S/5 Neuromuscular Transmission Module, E-NMT and Accessories
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
GE Healthcare · KOI Stimulator, Nerve, Peripheral, Electric
Cleared on Jul 15, 2005 as substantially equivalent to K955026 (Datex Medical Instrumentation, Inc., 1997) after a 25-day review. Its family tree reaches back 1 generation to 1997. 7 later devices descend from it. It has been recalled once.
- Decided
- Jul 15, 2005
- Review
- 25 days
- Class
- Class II
- Product code
- KOI
- Recalls
- 1
- Predicates
- 1
Its longest line of descent runs back 1 generation to K955026 (1997). 7 later devices descend from it, 1 citing it directly.
- Root · 1 generation backK955026 AS/3 NMT ModuleDatex Medical Instrumentation, Inc. · 1997 · Recalled
- This deviceK051635 Datex-Ohmeda S/5 Neuromuscular Transmission Module, E-NMT and…GE Healthcare · 2005
- Descendant · cites it directly · 6 of its ownK190795 Tetragraph Neuromuscular Transmission MonitorSenzime AB · 2019
- Descendant · 2 generations on · 3 of its ownK201949 Smart Cable NMT Module and Accessories, Life Scope BSM 3000 Series…Nihon Kohden Corporation · 2021
- Descendant · 2 generations on · 1 of its ownK220530 Tetragraph Neuromuscular Transmission MonitorSenzime AB · 2022
- Descendant · 3 generations on · 1 of its ownK213316 Life Scope PT BSM-1700 Series, AY Series, Data Acquisition Unit…Nihon Kohden Corporation · 2021
- Descendant · 3 generations on · 0 of its ownK261098 TetraGraph Neuromuscular Transmission Monitor (SEN…Senzime AB · 2026
Showing 5 of 7 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 7 · See all- K190795Tetragraph Neuromuscular Transmission MonitorSenzime AB · 2019 · cites it directly6descendants
- K201949Smart Cable NMT Module and Accessories, Life Scope BSM 3000 Series and Life Scope BSM…Nihon Kohden Corporation · 2021 · 2 generations on3descendants
- K220530Tetragraph Neuromuscular Transmission MonitorSenzime AB · 2022 · 2 generations on1descendant
- K213316Life Scope PT BSM-1700 Series, AY Series, Data Acquisition Unit, LIFE SCOPE BSM 6000…Nihon Kohden Corporation · 2021 · 3 generations on1descendant
- K261098TetraGraph Neuromuscular Transmission Monitor (SEN 2015)Accessories:TetraSens (SEN…Senzime AB · 2026 · 3 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K955026 | AS/3 NMT Module | Datex Medical Instrumentation, Inc. | 1997 | 1 |
| 0 (this device) | K051635 | Datex-Ohmeda S/5 Neuromuscular Transmission Module, E-NMT and Accessories | GE Healthcare | 2005 | — |
| +1 | K190795 | Tetragraph Neuromuscular Transmission Monitor | Senzime AB | 2019 | — |
| +2 | K201949 | Smart Cable NMT Module and Accessories, Life Scope BSM 3000 Series and Life… | Nihon Kohden Corporation | 2021 | — |
| +2 | K220530 | Tetragraph Neuromuscular Transmission Monitor | Senzime AB | 2022 | — |
| +3 | K213316 | Life Scope PT BSM-1700 Series, AY Series, Data Acquisition Unit, LIFE SCOPE… | Nihon Kohden Corporation | 2021 | — |
| +3 | K261098 | TetraGraph Neuromuscular Transmission Monitor (SEN 2015)Accessories:TetraSens… | Senzime AB | 2026 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2019 | 1 | 0 |
| 2021 | 3 | 0 |
| 2022 | 2 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in KOI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KOI device
- #1 K261098 TetraGraph Neuromuscular Transmission Monitor…score 21.1 · 39 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K955026 | AS/3 NMT ModuleSummary page 1…uscular Transmission Module, E-NMT is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda M NMT Module (K955026). Page 2 of 4 DE… | 1997 |
What descends from it? (7)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K190795 | Tetragraph Neuromuscular Transmission Monitor | 2019 | 6 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Apr 9, 2014 | Class II | Z-1640-2014 | Software design | When the E-NMT-01 module is used in conjunction with the ElectroSensor, the Neuromuscular Transmission (NMT) values may indicate a deeper level of muscle relaxation than the actual level of muscle relaxation. In the clinical situation… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All KOI devicesKOI Stimulator, Nerve, Peripheral, Electric: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · KOI
2in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KOI
174in 2022
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KOI
No recall names KOI since 2007.
Adverse event reports per year · KOI
5in 2025
MAUDE reports naming product code KOI, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 0 | — | 0 | 1 |
| 2012 | 0 | — | 0 | 3 |
| 2013 | 0 | — | 0 | 2 |
| 2014 | 0 | — | 0 | 9 |
| 2015 | 0 | — | 0 | 1 |
| 2016 | 1 | — | 0 | 0 |
| 2017 | 0 | — | 0 | 0 |
| 2018 | 2 | — | 0 | 1 |
| 2019 | 1 | — | 0 | 4 |
| 2020 | 1 | — | 0 | 4 |
| 2021 | 2 | — | 0 | 3 |
| 2022 | 3 | 174 | 0 | 4 |
| 2023 | 0 | — | 0 | 1 |
| 2024 | 1 | — | 0 | 2 |
| 2025 | 2 | — | 0 | 5 |
| 2026 | 1 | — | 0 | — |
What is it indicated for?
The Datex-Ohmeda S/5TM E-NMT module is indicated for monitoring the relaxation of the patient and regional block stimulation for nerve location.
The device is indicated for use by qualified medical personnel only.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K051635 · printed Nov 6, 2026
Cite this page
510(k) Report. “K051635: Datex-Ohmeda S/5 Neuromuscular Transmission Module, E-NMT and Accessories.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051635
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026