510(k) · Cleared
Recalled · Class II · 2026FR.O.H. Calcaneus Repair System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
I.T.S. Implantat-Technologie-Systeme GmbH · NDF Plate, Fixation, Bone, Non-Spinal, Metallic
Cleared on Aug 22, 2005 as substantially equivalent to K991407 (Synthes (Usa), 1999) and 2 other predicates after a 63-day review. Its family tree reaches back 2 generations to 1998. 28 later devices descend from it. It has been recalled once.
- Decided
- Aug 22, 2005
- Review
- 63 days
- Class
- Class II
- Product code
- NDF
- Recalls
- 1
- Predicates
- 3
Its longest line of descent runs back 2 generations to K981372 (1998). 5 ancestors in all, the oldest K903810 (1990). 28 later devices descend from it, 4 citing it directly.
- Root · 2 generations backK981372 Synthes 3.0 MM PC-Fix SystemSynthes (Usa) · 1998
- 1 generation backK991407 Synthes Locking Calcaneal PlatesSynthes (Usa) · 1999
- This deviceK051642 FR.O.H. Calcaneus Repair SystemI.T.S. Implantat-Technologie-Systeme GmbH · 2005
- Descendant · cites it directly · 27 of its ownK060156 Straight Plate with Angular Stability & Screw SystemI.T.S. Implantat-Technologie-Systeme GmbH · 2006 · Recalled
- Descendant · 2 generations on · 9 of its ownK150693 PHIN-FX Cannulated Lag Screw SystemAscent Medical, LLC · 2015
- Descendant · cites it directly · 9 of its ownK063166 I.T.S. Pelvic Reconstruction SystemI.T.S. Implantat-Technologie-Systeme GmbH · 2006 · Recalled
- Descendant · 2 generations on · 8 of its ownK110895 Ortho Solutions Trauma Implants for OsteosynthesisOrtho Solutions Limited · 2011
- Descendant · 3 generations on · 4 of its ownK161904 GEO Bone Screw SystemGramercy Extremity Orthopeidics, LLC · 2017
Showing 5 of 28 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 28 · See all- K060156Straight Plate with Angular Stability & Screw SystemI.T.S. Implantat-Technologie-Systeme GmbH · 2006 · cites it directly27descendants
- K150693PHIN-FX Cannulated Lag Screw SystemAscent Medical, LLC · 2015 · 2 generations on9descendants
- K063166I.T.S. Pelvic Reconstruction SystemI.T.S. Implantat-Technologie-Systeme GmbH · 2006 · cites it directly9descendants
- K110895Ortho Solutions Trauma Implants for OsteosynthesisOrtho Solutions Limited · 2011 · 2 generations on8descendants
- K161904GEO Bone Screw SystemGramercy Extremity Orthopeidics, LLC · 2017 · 3 generations on4descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K981372 | Synthes 3.0 MM PC-Fix System | Synthes (Usa) | 1998 | — |
| −1 | K991407 | Synthes Locking Calcaneal Plates | Synthes (Usa) | 1999 | — |
| 0 (this device) | K051642 | FR.O.H. Calcaneus Repair System | I.T.S. Implantat-Technologie-Systeme GmbH | 2005 | — |
| +1 | K060156 | Straight Plate with Angular Stability & Screw System | I.T.S. Implantat-Technologie-Systeme GmbH | 2006 | 2 |
| +2 | K150693 | PHIN-FX Cannulated Lag Screw System | Ascent Medical, LLC | 2015 | — |
| +1 | K063166 | I.T.S. Pelvic Reconstruction System | I.T.S. Implantat-Technologie-Systeme GmbH | 2006 | 2 |
| +2 | K110895 | Ortho Solutions Trauma Implants for Osteosynthesis | Ortho Solutions Limited | 2011 | — |
| +3 | K161904 | GEO Bone Screw System | Gramercy Extremity Orthopeidics, LLC | 2017 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1990 | 1 | 0 |
| 1992 | 1 | 0 |
| 2000 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2006 | 2 | 0 |
| 2010 | 1 | 0 |
| 2011 | 1 | 0 |
| 2012 | 1 | 0 |
| 2013 | 1 | 0 |
| 2015 | 3 | 0 |
| 2016 | 1 | 0 |
| 2017 | 3 | 0 |
| 2020 | 2 | 0 |
| 2022 | 4 | 0 |
| 2023 | 2 | 0 |
| 2024 | 1 | 0 |
| 2025 | 1 | 0 |
| 2026 | 5 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in NDF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K991407 | Synthes Locking Calcaneal PlatesSummary page 1…fixation fastener (see 21 CFR, Sec. 888.3040) DEVICE PRODUCT CODE: HRS & HWC SUBSTANTIALLY Synthes Locking Calcaneal Plates (1K991407) EQUIVALENT DEVICES DePuy/Ace Calcaneal Peri-articular Plate (K(993465) Ace/DePuy Cannulated Cance… | 1999 |
| * | K993465 | Depuy Ace Calcaneal Peri-Articular PlateSummary page 1…E PRODUCT CODE: HRS & HWC SUBSTANTIALLY Synthes Locking Calcaneal Plates (1K991407) EQUIVALENT DEVICES DePuy/Ace Calcaneal Peri-articular Plate (K(993465) Ace/DePuy Cannulated Cancellous Bone Screw (K(903810) DEVICE DESCRIPTION: The 1.T.S. FR.O.H. Cal… | 2000 |
| * | K903810 | Ace Cannulated Self Tapping Cancellous Bone ScrewSummary page 1…eal Plates (1K991407) EQUIVALENT DEVICES DePuy/Ace Calcaneal Peri-articular Plate (K(993465) Ace/DePuy Cannulated Cancellous Bone Screw (K(903810) DEVICE DESCRIPTION: The 1.T.S. FR.O.H. Calcaneus Repair System is a combination of fracture redu… | 1990 |
What descends from it? (28)
See all 284 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jan 23, 2026 | Class II | Z-1595-2026 | No Marketing Application | Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 7 years and passes through 2 companies; a new generation was cleared every 6.2 years on median.
- 1 of the 5 ancestors has a recall of any class.
- Of its descendants, K060156 (I.T.S. Implantat-Technologie-Systeme GmbH, 2006) has gone furthest, with 27 descendants of its own.
- 15 other devices also cite its predicate K991407.
How does its category compare?
All NDF devicesNDF Plate, Fixation, Bone, Non-Spinal, Metallic: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · NDF
FDA has decided no NDF device since 2007.
Median 510(k) review, days · NDF
Too few NDF 510(k)s in any year since 2007 to give a median.
Recalls per year · NDF
1in 2013
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · NDF
1in 2022
MAUDE reports naming product code NDF, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 1 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 0 | 1 |
| 2010 | 0 | — | 0 | 1 |
| 2011 | 0 | — | 0 | 0 |
| 2012 | 0 | — | 0 | 0 |
| 2013 | 0 | — | 1 | 1 |
| 2014 | 0 | — | 0 | 0 |
| 2015 | 0 | — | 0 | 0 |
| 2016 | 0 | — | 0 | 0 |
| 2017 | 0 | — | 0 | 1 |
| 2018 | 0 | — | 0 | 0 |
| 2019 | 0 | — | 0 | 0 |
| 2020 | 0 | — | 0 | 0 |
| 2021 | 0 | — | 0 | 0 |
| 2022 | 0 | — | 0 | 1 |
| 2023 | 0 | — | 0 | 0 |
| 2024 | 0 | — | 0 | 0 |
| 2025 | 0 | — | 0 | 0 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The I.T.S. FR.O.M. Calcaneus Repair System is a titanium implant fracture fixation system for repairing fractures located in the calcaneus heel bone of the foot.
Indications for Use include: Intra and extra-articular fracture(s) of the calcaneus, corrective osteotomy, joint depression, non-displaced and tongue type, severely comminuted fractures, multifragmentary fractures, revision procedures, joint fusion, stabilization and fixation of fresh fractures, reconstruction of the calcaneus bones, and open and closed fractures of the calcaneus. The system can be used in both adult and pediatric patients.
This system is not intended for spinal use.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K051642 · printed Nov 6, 2026
Cite this page
510(k) Report. “K051642: FR.O.H. Calcaneus Repair System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051642