Vericis Cardiovascular Image and Information System, Model 4.0

K051649510(k) TraditionalCleared Aug 2, 200542-day reviewClass II21 CFR 892.2050Recalls 17

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Camtronics Medical Systems · LLZ System, Image Processing, Radiological

Cleared on Aug 2, 2005 as substantially equivalent to K043146 (Mckesson Medical Imaging Company, 2005) after a 42-day review. Its family tree reaches back 3 generations to 1993. 10 later devices descend from it. It has been recalled 17 times.

Decided
Aug 2, 2005
Review
42 days
Class
Class II
Product code
LLZ
Recalls
17
Predicates
1

Its longest line of descent runs back 3 generations to K922164 (1993). 7 ancestors in all, the oldest K920319 (1992). 10 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor1993: 1 other ancestor1998: 1 other ancestor1992: 1 other ancestor2008: 1 descendant cleared2019: 2 descendants cleared2005: 1 descendant cleared2011: 1 descendant cleared2012: 1 descendant cleared2018: 1 descendant cleared2024: 1 descendant cleared2021: 1 descendant cleared2022: 1 descendant clearedReference device: K955519 · Analytical Review Station · Camtronics, Ltd. · 1996 · recalledReference device: K992259 · Echocardiography System Series 95000 · Camtronics, Ltd. · 1999K922164 · Raytel Digital Imaging Systems · Olicon Imaging Systems, Inc. · 1993K973959 · Olicon Imaging Systems, Inc. 02-Workstation and/or Pacsview Software · Olicon Imaging Systems, Inc. · 1997K043146 · Horizon Medical Imaging · Mckesson Medical Imaging Company · 2005 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors3 generations back: 2 other ancestors 1 generation forward: 3 descendants2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 2 descendants5 generations forward: 2 descendants6 generations forward: 1 descendant K043146 · Horizon Medical Imaging · Mckesson Medical Imaging Company · 2005 · recalledK973959 · Olicon Imaging Systems, Inc. 02-Workstation and/or Pacsview Software · Olicon Imaging Systems, Inc. · 1997K922164 · Raytel Digital Imaging Systems · Olicon Imaging Systems, Inc. · 19930+5
  1. Root · 3 generations backK922164 Raytel Digital Imaging SystemsOlicon Imaging Systems, Inc. · 1993
  2. 2 generations backK973959 Olicon Imaging Systems, Inc. 02-Workstation and/or Pacsview SoftwareOlicon Imaging Systems, Inc. · 1997
  3. 1 generation backK043146 Horizon Medical ImagingMckesson Medical Imaging Company · 2005 · Recalled
  4. This deviceK051649 Vericis Cardiovascular Image and Information System, Model 4.0Camtronics Medical Systems · 2005
  5. Descendant · cites it directly · 7 of its ownK053133 Syngo Dynamics, Version 5.0Siemens Medical Solutions USA, Inc. · 2005 · Recalled
  6. Descendant · 2 generations on · 6 of its ownK111319 Ibis Explorer and Markup SoftwareMedqia · 2011
  7. Descendant · 3 generations on · 5 of its ownK122205 EXINIExini Diagnostics AB · 2012
  8. Descendant · 4 generations on · 2 of its ownK191262 aBSIExini Diagnostics AB · 2019
  9. Descendant · 5 generations on · 1 of its ownK211655 aPROMISEExini Diagnostics AB · 2021

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2005: 12008: 12011: 12012: 12018: 12019: 22021: 12022: 12024: 1
2005high 2 (2019) · low 0 (2006)2026

Most influential descendants

Showing 5 of 10 · See all
  • K053133Syngo Dynamics, Version 5.0Siemens Medical Solutions USA, Inc. · 2005 · cites it directly7descendants
  • K111319Ibis Explorer and Markup SoftwareMedqia · 2011 · 2 generations on6descendants
  • K122205EXINIExini Diagnostics AB · 2012 · 3 generations on5descendants
  • K191262aBSIExini Diagnostics AB · 2019 · 4 generations on2descendants
  • K211655aPROMISEExini Diagnostics AB · 2021 · 5 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K922164Raytel Digital Imaging SystemsOlicon Imaging Systems, Inc.1993—
−2K973959Olicon Imaging Systems, Inc. 02-Workstation and/or Pacsview SoftwareOlicon Imaging Systems, Inc.1997—
−1K043146Horizon Medical ImagingMckesson Medical Imaging Company200511
0 (this device)K051649Vericis Cardiovascular Image and Information System, Model 4.0Camtronics Medical Systems2005—
+1K053133Syngo Dynamics, Version 5.0Siemens Medical Solutions USA, Inc.20052
+2K111319Ibis Explorer and Markup SoftwareMedqia2011—
+3K122205EXINIExini Diagnostics AB2012—
+4K191262aBSIExini Diagnostics AB2019—
+5K211655aPROMISEExini Diagnostics AB2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199210
199310
199810
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
200810
201110
201210
201810
201920
202110
202210
202410
FIG. 1 — Lineage of K051649, 1993–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K043146Horizon Medical Imaging
Summary page 1
…e: System, Image Processing, Radiological Regulation Class: II Product Code: LZZ C. Predicate Devices McKesson Horizon Medical Imaging (K043146) D. Device Description The VERICIS is an integrated cardiovascular information system classified as a picture…
Mckesson Medical Imaging Company200511
†K955519Analytical Review StationCamtronics, Ltd.19961
†K992259Echocardiography System Series 95000Camtronics, Ltd.1999—

Also cleared against K043146: 4 other devices →

What descends from it? (10)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K053133Syngo Dynamics, Version 5.0Siemens Medical Solutions USA, Inc.20057
K192276Merge CardioMerge Healthcare Incorporated20190
K082706Infinitt Cardiology PacsInfinitt Co., Ltd.20080

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 19, 2025Class IIZ-1401-2025Software designWhen following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent…
Feb 19, 2025Class IIZ-1398-2025Software designWhen following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent…
Feb 19, 2025Class IIZ-1399-2025Software designWhen following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent…
Feb 19, 2025Class IIZ-1400-2025Software designWhen following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent…
May 8, 2020Class IIZ-2403-2020Software Design ChangeAn error in the calculation of the Michigan Reference Ranges can generate an inaccurate Z-Score calculation. The error may reduce the sensitivity of the Z-Score in detecting an abnormality resulting in a false negative.
Apr 13, 2016Class IIZ-2123-2017Software designFetal patient report was automatically pulling prior measurement data for a prior fetus since the fetal study is based on the mother's MRN, resulting in the wrong fetal measurements getting referenced.
Apr 4, 2016Class IIZ-1486-2017Software designCardio study list does not show STAT studies without refreshing.
Apr 4, 2016Class IIZ-1415-2017Software designA situation can occur allowing two physicians to access the same study report in EchoIMS when launched from the Cardio Study List without receiving the read-only notification prompt.
Apr 4, 2016Class IIZ-1403-2017Software designWhen taking measurements from images on the Cardio workstation or from the US cart, numbers are not crossing to the report in the correct unit of measure.
Mar 28, 2016Class IIZ-0729-2017Software designIf images are sent without an order in the system, they will be matched with the latest order on the current patient/modality matching potentially resulting in the matching of the report to the incorrect accession number (but still…
Mar 28, 2016Class IIZ-1046-2017Software designIf a reader selects Multi-study review prior to the original study completing its loading process and then immediately returns to image review of the original study, it is possible that not all current images will be present for review…
Feb 9, 2016Class IIZ-2709-2016Software designA report can be confirmed with the incorrect patient demographics resulting in it being saved under the incorrect patient record in the Electronic Medical Record (EMR).
Jan 30, 2016Class IIZ-2865-2016Software designPatient information in the header is only found on the first page of the report and not on the subsequent pages.
Oct 26, 2015Class IIZ-0555-2016Process controlThe firm, Merge Healthcare, sent out Merge HEMO V10.0 & Merge CARDIO V10.1 software to 3 customers before the product was completely validated in house.
Oct 1, 2015Class IIZ-0730-2017Software designReporting feature times out after inactivity for more than an hour sending the user back to the study list, which causes all reporting data being entered to be lost.
Sep 22, 2015Class IIZ-1496-2017Process designFor sites using the Issuer of Patient ID (IPID), the system will display the study list and images from different patients with different IPIDs as though they are for the same patient if they all have the first name, last name, and…
Sep 22, 2015Class IIZ-1517-2017Software designUsers can merge a device import file with an image study that already has a confirmed report, which may result in including information not present at the time of physician interpretation.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 3 companies; a new generation was cleared every 4.9 years on median.
  • 1 of the 7 ancestors has a recall of any class.
  • Of its descendants, K053133 (Siemens Medical Solutions USA, Inc., 2005) has gone furthest, with 7 descendants of its own.
  • 4 other devices also cite its predicate K043146.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

VERICIS is a system intended to be used to acquire, store, print, transfer, and archive clinical information from Camtronics and other vendors systems including images, Hemodynamic studies and reports, measurements (via import from DICOM Structured Reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data. VERICIS is intended to allow users to review diagnostic and non-diagnostic…

Read the full statement (215 words)

VERICIS is a system intended to be used to acquire, store, print, transfer, and archive clinical information from Camtronics and other vendors systems including images, Hemodynamic studies and reports, measurements (via import from DICOM Structured Reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data. VERICIS is intended to allow users to review diagnostic and non-diagnostic quality images, annotate studies, perform digital subtraction on images, to perform quantitative measurements on images (including but not limited to quantitative coronary analysis, left ventricular analysis, time, area, length, velocity, angle, volume, and velocity-time integrals), to generate physician-generated clinical reports (via structure reporting and template based tools), and to store this information in a database.

*VERICIS is software comprised of modules that perform under standard "off-the-shelf' personal computers and servers running the Microsoft Windows 2000/2003/XP operating systems. eVERICIS is image data storage and display software that accepts DICOM (Digital Imaging and Communications in Medicine) image data files from multiple modalities. It accepts text data using other standards-based formats including but not limited to HL7 and XML. eVERICIS is an Intemet/Intranet network system that is designed for small and large, multi-user environments. The VERICIS network structure (including server and workstations) provides for the system's database management, storage, printing, and all DICOM/HL-7 interface services.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051649: Vericis Cardiovascular Image and Information System, Model 4.0.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051649

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026