510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Hemoglide Series and Hemoglide Star Series XK Long-Term Hemodialysis Catheters

K051748510(k) SpecialCleared Aug 12, 200544-day reviewClass II21 CFR 876.5540Recalls 13

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

C.R. Bard, Inc. · MSD Catheter, Hemodialysis, Implanted

Cleared on Aug 12, 2005 as substantially equivalent to K030020 (C.R. Bard, Inc., 2003) and 1 other predicate after a 44-day review. Its family tree reaches back 5 generations to 1998. 8 later devices descend from it. It has been recalled 13 times.

Decided
Aug 12, 2005
Review
44 days
Class
Class II
Product code
MSD
Recalls
13
Predicates
2

Its longest line of descent runs back 5 generations to K981994 (1998). 7 ancestors in all, the oldest K972207 (1998). 8 later devices descend from it, 2 citing it directly.

Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 4 descendants3 generations forward: 1 descendant4 generations forward: 1 descendant K033294 · Hemosplit XK, Hemosplit XK Catheter, 16FR X 42 CM, Kit;16FR X 19 CM, Kit with Microintroducer · C.R. Bard, Inc. · 2003K030020 · Hemosplit Long-Term Hemodialysis Catheter · C.R. Bard, Inc. · 2003K020465 · Medcomp Ash Split-Cath II · Medcomp · 2002K012562 · Medcomp 14.5F Double Lumen Hemodialysis Catheter · Medical Components, Inc. · 2002K981994 · Vascath Opti-Flow Long Term Dual Lumen Hemodialysis Catheter · C.R. Bard, Inc. · 1998−50
  1. Root · 5 generations backK981994 Vascath Opti-Flow Long Term Dual Lumen Hemodialysis CatheterC.R. Bard, Inc. · 1998
  2. 4 generations backK012562 Medcomp 14.5F Double Lumen Hemodialysis CatheterMedical Components, Inc. · 2002
  3. 3 generations backK020465 Medcomp Ash Split-Cath IIMedcomp · 2002
  4. 2 generations backK030020 Hemosplit Long-Term Hemodialysis CatheterC.R. Bard, Inc. · 2003
  5. 1 generation backK033294 Hemosplit XK, Hemosplit XK Catheter, 16FR X 42 CM, Kit;16FR X 19 CM…C.R. Bard, Inc. · 2003
  6. This deviceK051748 Hemoglide Series and Hemoglide Star Series XK Long-Term Hemodialysis…C.R. Bard, Inc. · 2005
  7. Descendant · cites it directly · 5 of its ownK190527 GlidePath Long-Term Hemodialysis Catheters, HemoStar Long-Term…C. R. Bard · 2020
  8. Descendant · 2 generations on · 2 of its ownK200627 GlidePath 10F Long-Term Dialysis CatheterC.R. Bard, Inc. · 2020
  9. Descendant · 3 generations on · 1 of its ownK202150 GlidePath 7.5F Long-Term Dialysis CatheterC.R. Bard, Inc. · 2020
  10. Descendant · cites it directly · 1 of its ownK053589 Long-Term Hemodialysis Catheters with Biobloc CoatingBard Access Systems, Inc. · 2006
  11. Descendant · 4 generations on · 0 of its ownK241581 Hemodialysis CatheterHaolang Medical USA Corporation · 2025

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2006: 12007: 12020: 32021: 22025: 1
2006high 3 (2020) · low 0 (2008)2026

Most influential descendants

Showing 5 of 8 · See all
  • K190527GlidePath Long-Term Hemodialysis Catheters, HemoStar Long-Term Hemodialysis Catheters…C. R. Bard · 2020 · cites it directly5descendants
  • K200627GlidePath 10F Long-Term Dialysis CatheterC.R. Bard, Inc. · 2020 · 2 generations on2descendants
  • K202150GlidePath 7.5F Long-Term Dialysis CatheterC.R. Bard, Inc. · 2020 · 3 generations on1descendant
  • K053589Long-Term Hemodialysis Catheters with Biobloc CoatingBard Access Systems, Inc. · 2006 · cites it directly1descendant
  • K241581Hemodialysis CatheterHaolang Medical USA Corporation · 2025 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K981994Vascath Opti-Flow Long Term Dual Lumen Hemodialysis CatheterC.R. Bard, Inc.1998—
−4K012562Medcomp 14.5F Double Lumen Hemodialysis CatheterMedical Components, Inc.2002—
−3K020465Medcomp Ash Split-Cath IIMedcomp2002—
−2K030020Hemosplit Long-Term Hemodialysis CatheterC.R. Bard, Inc.2003—
−1K033294Hemosplit XK, Hemosplit XK Catheter, 16FR X 42 CM, Kit;16FR X 19 CM, Kit with…C.R. Bard, Inc.2003—
0 (this device)K051748Hemoglide Series and Hemoglide Star Series XK Long-Term Hemodialysis CathetersC.R. Bard, Inc.2005—
+1K190527GlidePath Long-Term Hemodialysis Catheters, HemoStar Long-Term Hemodialysis…C. R. Bard2020—
+2K200627GlidePath 10F Long-Term Dialysis CatheterC.R. Bard, Inc.2020—
+3K202150GlidePath 7.5F Long-Term Dialysis CatheterC.R. Bard, Inc.2020—
+1K053589Long-Term Hemodialysis Catheters with Biobloc CoatingBard Access Systems, Inc.2006—
+4K241581Hemodialysis CatheterHaolang Medical USA Corporation2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200710
202030
202120
202510
FIG. 1 — Lineage of K051748, 1998–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MSD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MSD device

  1. #1 K241581 Hemodialysis Catheterscore 12.4 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K030020Hemosplit Long-Term Hemodialysis Catheter
Summary page 1
…Gastroenterology and Renal I-. -~ tditevOevicbEhIfit <410r~- dH~~tt HemoSplit Long-Term Heniodialysis Catheter K030020 6/16/2003 HemoSplit XK Long-Terin Heniodialywsi Catheter K033294 11/26/2003…
C.R. Bard, Inc.2003—
*K033294Hemosplit XK, Hemosplit XK Catheter, 16FR X 42 CM, Kit;16FR X 19 CM, Kit with…
Summary page 1
…10r~- dH~~tt HemoSplit Long-Term Heniodialysis Catheter K030020 6/16/2003 HemoSplit XK Long-Terin Heniodialywsi Catheter K033294 11/26/2003 0o0038 ===…
C.R. Bard, Inc.2003—
†K010567Optiflow Catheter #5641190 Catheter Only #5643150 Catheter Kit #5631000 Catheterless…C.R. Bard, Inc.2001—

Also cleared against K030020: 3 other devices →

What descends from it? (8)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K190527GlidePath Long-Term Hemodialysis Catheters, HemoStar Long-Term Hemodialysis Catheters…C. R. Bard20205
K053589Long-Term Hemodialysis Catheters with Biobloc CoatingBard Access Systems, Inc.20061

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 31, 2019Class IIZ-2388-2019Nonconforming Material/ComponentThe action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
May 31, 2019Class IIZ-2390-2019Nonconforming Material/ComponentThe action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
May 31, 2019Class IIZ-2386-2019Nonconforming Material/ComponentThe action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
May 31, 2019Class IIZ-2387-2019Nonconforming Material/ComponentThe action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
May 31, 2019Class IIZ-2385-2019Nonconforming Material/ComponentThe action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
May 31, 2019Class IIZ-2392-2019Nonconforming Material/ComponentThe action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
May 31, 2019Class IIZ-2393-2019Nonconforming Material/ComponentThe action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
May 31, 2019Class IIZ-2384-2019Nonconforming Material/ComponentThe action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
May 31, 2019Class IIZ-2395-2019Nonconforming Material/ComponentThe action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
May 31, 2019Class IIZ-2391-2019Nonconforming Material/ComponentThe action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
May 31, 2019Class IIZ-2394-2019Nonconforming Material/ComponentThe action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
May 31, 2019Class IIZ-2383-2019Nonconforming Material/ComponentThe action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.
May 31, 2019Class IIZ-2389-2019Nonconforming Material/ComponentThe action is being initiated due to complaints received concerning the molded tip of the tunneler breaking or is damaged during the placement procedure of the catheter.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 7 years and passes through 3 companies; a new generation was cleared every 1.1 years on median.
  • None of the 7 ancestors has a recall.
  • Of its descendants, K190527 (C. R. Bard, 2020) has gone furthest, with 5 descendants of its own.
  • 3 other devices also cite its predicate K030020.

How does its category compare?

All MSD devices

MSD Catheter, Hemodialysis, Implanted: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MSD

1in 2025

2007: 72008: 32009: 62010: 42011: 72012: 22013: 52014: 82015: 12016: 22017: 22018: 22019: 12020: 42021: 42025: 1
2007high 8 (2014) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MSD

153in 2021

2007: 92 days median2008: 89 days median2009: 81 days median2010: 116 days median2011: 70 days median2013: 105 days median2014: 246 days median2020: 106 days median2021: 153 days median
2007high 246 (2014) · low 70 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MSD

1in 2013

2008: 472009: 492010: 112011: 242013: 12026 (partial year): 4, 4 Class I
2007high 49 (2009) · low 0 (2007)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · MSD

1,054in 2025

2007: 1182008: 1742009: 1732010: 3062011: 2682012: 3672013: 3632014: 5062015: 5502016: 5322017: 4842018: 4632019: 6452020: 4012021: 5232022: 6102023: 5422024: 8412025: 1,054
2007high 1,054 (2025) · low 118 (2007)2025

MAUDE reports naming product code MSD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077920118
200838947174
200968149173
2010411611306
201177024268
20122—0367
201351051363
201482460506
20151—0550
20162—0532
20172—0484
20182—0463
20191—0645
202041060401
202141530523
20220—0610
20230—0542
20240—0841
20251—01,054
20260—4 (4)—

What is it indicated for?

The HemoGlide® Star series and HemoGlide® Star series XK long-term hemodialysis catheters are indicated for use in attaining short-term or long-term vascular access for hemodialysis, hemoperfusion or apheresis therapy. Access is attained via the internal jugular vein, external jugular vein, subclavian vein, or femoral vein.

Catheters greater than 40 cm are intended for femoral vein insertion.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051748: Hemoglide Series and Hemoglide Star Series XK Long-Term Hemodialysis Catheters.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K051748

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026