K051918510(k) SpecialCleared Aug 17, 200533-day reviewClass II21 CFR 882.5960Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Jerome Medical · HAX Tong, Skull For Traction

Cleared on Aug 17, 2005 as substantially equivalent to K023959 (The Jerome Group, Inc., 2003) after a 33-day review. Its family tree reaches back 2 generations to 1982. It has been recalled 7 times.

Decided
Aug 17, 2005
Review
33 days
Class
Class II
Product code
HAX
Recalls
7
Predicates
1

Its longest line of descent runs back 2 generations to K822780 (1982). No later device cites it as a predicate yet.

K822780 · Cervical Brace · Jerome Medical · 1982 · recalledK023959 · Jerome Glass Ring · The Jerome Group, Inc. · 2003 · recalled198019851990199520002005
  1. Root · 2 generations backK822780 Cervical BraceJerome Medical · 1982 · Recalled
  2. 1 generation backK023959 Jerome Glass RingThe Jerome Group, Inc. · 2003 · Recalled
  3. This deviceK051918 Resolve Open Back Halo RingsJerome Medical · 2005
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K822780Cervical BraceJerome Medical19821
−1K023959Jerome Glass RingThe Jerome Group, Inc.20036
0 (this device)K051918Resolve Open Back Halo RingsJerome Medical2005—
FIG. 1 — Lineage of K051918, 1982–2005Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HAX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K023959Jerome Glass Ring
Summary page 1
…ication Name: Tong, Skull for Traction, Sec. 882.5960, Neurological Devices, Class II, HAX Substantially Equ~ivalent To: ReSOLVE Closed Back Halo Ring (K023959) manufactured by Jerome Medical. Also similar to the V1 Halo Ring (K9301 53) manufactured by Jerome Medical.…
The Jerome Group, Inc.20036

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 21, 2018Class IIZ-0999-2019Labeling designHalo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line…
Jun 21, 2018Class IIZ-1002-2019Labeling designHalo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line…
Jun 21, 2018Class IIZ-1000-2019Labeling designHalo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line…
Jun 21, 2018Class IIZ-1001-2019Labeling designHalo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line…
Nov 9, 2008Not classifiedZ-1509-2009Process change controlOssur initiated the recall after their investigation of some complaints found that in some instances the ceramic pins are difficult to screw through the halo ring.
Nov 9, 2008Not classifiedZ-1507-2009Process change controlOssur initiated the recall after their investigation of some complaints found that in some instances the ceramic pins are difficult to screw through the halo ring.
Nov 9, 2008Not classifiedZ-1508-2009Process change controlOssur initiated the recall after their investigation of some complaints found that in some instances the ceramic pins are difficult to screw through the halo ring.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 2 companies; a new generation was cleared every 20.4 years on median.
  • 2 of the 2 ancestors have a recall of any class.

How does its category compare?

All HAX devices

HAX Tong, Skull For Traction: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · HAX

1in 2009

2007: 12009: 1
2007high 1 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HAX

Too few HAX 510(k)s in any year since 2007 to give a median.

Recalls per year · HAX

4in 2018

2008: 32014: 62018: 4
2007high 6 (2014) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HAX

1in 2020

2008: 12010: 12011: 22015: 12016: 12017: 12020: 1
2007high 2 (2011) · low 0 (2007)2025

MAUDE reports naming product code HAX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20080—31
20091—00
20100—01
20110—02
20120—00
20130—00
20140—60
20150—01
20160—01
20170—01
20180—40
20190—00
20200—01
20210—00
20220—00
20230—00
20240—00
20250—00
20260—0—

What is it indicated for?

The ReSOLVE Open Back Halo Ring is intended for use with a Jerome Halo Vest to immobilize

and/or provide traction for a cervical spine injury.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051918: Resolve Open Back Halo Rings.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K051918

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026