510(k) Report

510(k) · Cleared

Recalled · 2007

Ceegraph/Sleepscan Netlink Traveler

K052066510(k) SpecialCleared Aug 26, 200525-day reviewClass II21 CFR 882.1400Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bio-Logic Systems Corp. · OLV Standard Polysomnograph With Electroencephalograph

Cleared on Aug 26, 2005 as substantially equivalent to K002570 (Bio-Logic Systems Corp., 2000) and 3 other predicates after a 25-day review. Its family tree reaches back 2 generations to 1994. 5 later devices descend from it. It has been recalled 2 times.

Decided
Aug 26, 2005
Review
25 days
Class
Class II
Product code
OLV
Recalls
2
Predicates
4

Its longest line of descent runs back 2 generations to K930790 (1994). 5 ancestors in all. 5 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 2 other ancestors1996: 1 other ancestor2010: 1 descendant cleared2014: 1 descendant cleared2018: 1 descendant cleared2020: 1 descendant cleared2023: 1 descendant clearedK930790 · Addition of Built-in Oxymeter to Sleepscan Prod. · Bio-Logic Systems Corp. · 1994K962103 · Bio-Logic Sleepscan Product with Built-in Oximeter · Bio-Logic Systems Corp. · 19961990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 1 descendant5 generations forward: 1 descendant K962103 · Bio-Logic Sleepscan Product with Built-in Oximeter · Bio-Logic Systems Corp. · 1996K930790 · Addition of Built-in Oxymeter to Sleepscan Prod. · Bio-Logic Systems Corp. · 19940+5
  1. Root · 2 generations backK930790 Addition of Built-in Oxymeter to Sleepscan Prod.Bio-Logic Systems Corp. · 1994
  2. 1 generation backK962103 Bio-Logic Sleepscan Product with Built-in OximeterBio-Logic Systems Corp. · 1996
  3. This deviceK052066 Ceegraph/Sleepscan Netlink TravelerBio-Logic Systems Corp. · 2005
  4. Descendant · cites it directly · 4 of its ownK093728 BE Micro, TreaEb Neuro S.P.A. · 2010
  5. Descendant · 2 generations on · 3 of its ownK133355 Cadwell Flex Eeg/PSGCadwell Industries, Inc. · 2014
  6. Descendant · 3 generations on · 2 of its ownK180269 Cadwell Apollo SystemCadwell Industries, Inc. · 2018
  7. Descendant · 4 generations on · 1 of its ownK201819 Cadwell Apollo SystemCadwell Industries, Inc. · 2020
  8. Descendant · 5 generations on · 0 of its ownK221959 Q21Neurofield, Inc. · 2023

Most influential descendants

Showing 5 of 5 · See all
  • K093728BE Micro, TreaEb Neuro S.P.A. · 2010 · cites it directly4descendants
  • K133355Cadwell Flex Eeg/PSGCadwell Industries, Inc. · 2014 · 2 generations on3descendants
  • K180269Cadwell Apollo SystemCadwell Industries, Inc. · 2018 · 3 generations on2descendants
  • K201819Cadwell Apollo SystemCadwell Industries, Inc. · 2020 · 4 generations on1descendant
  • K221959Q21Neurofield, Inc. · 2023 · 5 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K930790Addition of Built-in Oxymeter to Sleepscan Prod.Bio-Logic Systems Corp.1994—
−1K962103Bio-Logic Sleepscan Product with Built-in OximeterBio-Logic Systems Corp.1996—
0 (this device)K052066Ceegraph/Sleepscan Netlink TravelerBio-Logic Systems Corp.2005—
+1K093728BE Micro, TreaEb Neuro S.P.A.2010—
+2K133355Cadwell Flex Eeg/PSGCadwell Industries, Inc.2014—
+3K180269Cadwell Apollo SystemCadwell Industries, Inc.2018—
+4K201819Cadwell Apollo SystemCadwell Industries, Inc.2020—
+5K221959Q21Neurofield, Inc.2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
200020

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201410
201810
202010
202310
FIG. 1 — Lineage of K052066, 1994–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OLV? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OLV device

  1. #1 K241288 Noxturnal Webscore 17.3 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K002570Bio-Logic Ceegraph Netlink
Summary page 1
…Digital EEBG/Sleep Recorder. CLASSIFICATION NAME: Electroencephalograph (per CFR 882.1400). PREDICATE DEVICE: Bio-logic Ceegraph Netlink (K002570), Sleepscan Netlink (K003681I), Ceegraph Traveler (K954954) and Sleepscan Traveler (K962 103).…
Bio-Logic Systems Corp.2000—
*K003681Bio-Logic Sleepscan Netlink
Summary page 1
…p Recorder. CLASSIFICATION NAME: Electroencephalograph (per CFR 882.1400). PREDICATE DEVICE: Bio-logic Ceegraph Netlink (K002570), Sleepscan Netlink (K003681I), Ceegraph Traveler (K954954) and Sleepscan Traveler (K962 103).…
Bio-Logic Systems Corp.2000—
*K954954Ceegraph Traveler Ambulatory Eeg Recorder
Summary page 1
…AME: Electroencephalograph (per CFR 882.1400). PREDICATE DEVICE: Bio-logic Ceegraph Netlink (K002570), Sleepscan Netlink (K003681I), Ceegraph Traveler (K954954) and Sleepscan Traveler (K962 103).…
Bio-Logic Systems Corp.1996—
*K962103Bio-Logic Sleepscan Product with Built-in Oximeter
Summary page 1
…halograph (per CFR 882.1400). PREDICATE DEVICE: Bio-logic Ceegraph Netlink (K002570), Sleepscan Netlink (K003681I), Ceegraph Traveler (K954954) and Sleepscan Traveler (K962 103). Page 8…
Bio-Logic Systems Corp.1996—

What descends from it? (5)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K093728BE Micro, TreaEb Neuro S.P.A.20104

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 26, 2007Not classifiedZ-0001-2008Device DesignBio-logic System Corp Camera Pole may weaken and detach at the base of the telescoping camera pole during transport causing pole to break at the welded base and fall.
Feb 26, 2007Not classifiedZ-0761-2007OtherBattery and Battery Compartment; incorrect placement in the Netlink Traveler unit which caused a short in the wiring of the battery pack and overheating.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 1 company; a new generation was cleared every 9.0 years on median.
  • None of the 5 ancestors has a recall.
  • Of its descendants, K093728 (Eb Neuro S.P.A., 2010) has gone furthest, with 4 descendants of its own.

How does its category compare?

All OLV devices

OLV Standard Polysomnograph With Electroencephalograph: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · OLV

1in 2025

2008: 22010: 12011: 22013: 12015: 12017: 32018: 32020: 22022: 12024: 22025: 12026 (partial year): 2
2007high 3 (2017) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OLV

141in 2018

2017: 83 days median2018: 141 days median
2007high 141 (2018) · low 83 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OLV

No recall names OLV since 2007.

Adverse event reports per year · OLV

1in 2025

2013: 12016: 42018: 32019: 12021: 32022: 32023: 22024: 22025: 1
2007high 4 (2016) · low 0 (2007)2025

MAUDE reports naming product code OLV, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20082—00
20090—00
20101—00
20112—00
20120—00
20131—01
20140—00
20151—00
20160—04
201738300
2018314103
20190—01
20202—00
20210—03
20221—03
20230—02
20242—02
20251—01
20262—0—

What is it indicated for?

The Bio-logic Ceegraph/Sleepscan Netlink Traveler is indicated for use in the recording and analysis of EEG tests. Typical routine EEG tests are 20-30 minutes in duration, but the Ceegraph/Sleepscan Netlink Traveler can also be used for longer tests, including continuous long-term EEG monitoring with patient video. Similarly, the Bio-logic Ceegraph/Sleepscan Netlink Traveler is indicated for use in the recording and analysis of human physiological data…

Read the full statement (179 words)

The Bio-logic Ceegraph/Sleepscan Netlink Traveler is indicated for use in the recording and analysis of EEG tests. Typical routine EEG tests are 20-30 minutes in duration, but the Ceegraph/Sleepscan Netlink Traveler can also be used for longer tests, including continuous long-term EEG monitoring with patient video. Similarly, the Bio-logic Ceegraph/Sleepscan Netlink Traveler is indicated for use in the recording and analysis of human physiological data necessary for the diagnosis of Sleep-related disorders. It is intended to record and present this data in a form that can improve the speed of diagnosis and assist in potential treatment decisions. In general, EEG and Sleep testing is indicated for use whenever it is necessary to measure and record a patient's electrophysiological activity, including the electrical activity of the brain, by attaching multiple electrodes at various locations on the body.

The Netlink Traveler can be used for patients of all ages, from children to adults, including geriatric patients.

The use of the Ceegraph/Sleepscan Netlink Traveler is to be performed under the prescription and supervision of a physician or other trained health care professional.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K052066: Ceegraph/Sleepscan Netlink Traveler.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K052066

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026