510(k) Report

510(k) · Cleared

Recalled · Class I · 2014

Reduced Energy Pediatric Electrode Pads for the Samaritan Aed and Pad, Models Sde 251, SPP 351

K052465510(k) TraditionalCleared Dec 12, 2006460-day reviewClass III21 CFR 870.5310Recalls 2 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Heartsine Technologies, Inc. · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Dec 12, 2006 as substantially equivalent to K052057 (Medtronic Emergency Response Systems, Inc., 2006) after a 460-day review. Its family tree reaches back 7 generations to 1993. 1 of the 27 devices in its family, counting itself, has a Class I recall. 2 later devices descend from it. It has been recalled 2 times, including a Class I recall.

Decided
Dec 12, 2006
Review
460 days
Class
Class III
Product code
MKJ
Recalls
2 (1 Class I)
Predicates
1

Its longest line of descent runs back 7 generations to K921669 (1993). 24 ancestors in all, the oldest K802288 (1980). 2 later devices descend from it, 1 citing it directly. 1 of the 27 devices in this family, counting this one, has a Class I recall.

  1. Root · 7 generations backK921669 Marquette SL Series Transport Remote AcquisitionMarquette Electronics, Inc. · 1993
  2. 6 generations backK964750 Eagle 4000 Patient MonitorMarquette Electronics, Inc. · 1997
  3. 5 generations backK973486 Life Pak 12Physio-Control Corp. · 1998 · Recalled
  4. 4 generations backK991910 Lifepak 12 BiphasicPhysio-Control Corp. · 1999 · Recalled
  5. 3 generations backK012274 Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/PacemakerMedtronic Physio-Control Corp. · 2002 · Recalled
  6. 2 generations backK033275 Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp. · 2003 · Recalled
  7. 1 generation backK052057 Lifepak 500 Automated External DefibrillatorMedtronic Emergency Response Systems, Inc. · 2006 · Recalled
  8. This deviceK052465 Reduced Energy Pediatric Electrode Pads for the Samaritan Aed and…Heartsine Technologies, Inc. · 2006
  9. Descendant · cites it directly · 1 of its ownK123881 Heartsine Samaritan Pad 350PHeartsine Technologies, Inc. · 2013 · Recalled
  10. Descendant · 2 generations on · 0 of its ownK142709 samaritan PAD 450PHeartsine Technologies, Inc. · 2015

Most influential descendants

Showing 2 of 2 · See all
  • K123881Heartsine Samaritan Pad 350PHeartsine Technologies, Inc. · 2013 · cites it directly1descendant
  • K142709samaritan PAD 450PHeartsine Technologies, Inc. · 2015 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K921669Marquette SL Series Transport Remote AcquisitionMarquette Electronics, Inc.1993—
−6K964750Eagle 4000 Patient MonitorMarquette Electronics, Inc.1997—
−5K973486Life Pak 12Physio-Control Corp.19983
−4K991910Lifepak 12 BiphasicPhysio-Control Corp.19993
−3K012274Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/PacemakerMedtronic Physio-Control Corp.200217
−2K033275Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp.20031
−1K052057Lifepak 500 Automated External DefibrillatorMedtronic Emergency Response Systems, Inc.20063
0 (this device)K052465Reduced Energy Pediatric Electrode Pads for the Samaritan Aed and Pad, Models…Heartsine Technologies, Inc.2006—
+1K123881Heartsine Samaritan Pad 350PHeartsine Technologies, Inc.20131
+2K142709samaritan PAD 450PHeartsine Technologies, Inc.2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198410
199010
199110
199210
199410
199540
199620
199910
200130
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201510
FIG. 1 — Lineage of K052465, 1993–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052057Lifepak 500 Automated External Defibrillator
Cleared while this submission was under review
Summary page 1
…llators, automatic, external 74 MKJ Common/Usual Name: Automated external defibrillator and electrodes. 2. Equivalent legally marketed device: Medtronic LIFEPAK 500, K052057 3. Indications for Use: The samaritan Pediatric-Pak (Model SPP 351) is indicated for use to treat patients in cardiopulmon…
Medtronic Emergency Response Systems, Inc.20063
†K042088Samaritan Pad Pak Defibrillation Electrode, Model SPP-301Heartsine Technologies, Inc.2004—

Also cleared against K052057: 1 other device →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K123881Heartsine Samaritan Pad 350PHeartsine Technologies, Inc.20131

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 12, 2014Class IIZ-2115-2014Packaging process controlA small number of sealed foil pouches containing the electrodes were found to be difficult to open, resulting in potential delay in therapy.
Sep 13, 2012Class IZ-0124-2013Device DesignCertain Samaritan 300/300P PAD devices may experience one or both of the following conditions that could affect the ability to deliver therapy to a patient in an SCA event if needed: Issue 1 (On/Off Issue): the device may turn itself on…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 5 companies; a new generation was cleared every 1.7 years on median.
  • 12 of the 24 ancestors have a recall of any class.
  • Of its descendants, K123881 (Heartsine Technologies, Inc., 2013) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K052057.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

The samaritan® Pediatric-Pak (Model SPP 351) is indicated for use to treat patients in cardiopulmonary arrest who are unconscious, without a pulse and not breathing spontaneously. They should only be used by personnel who have been trained in its operation.

The samaritan® Pediatric-Pak is specially designed for use only with the samaritan® PAD (Model SAM 300P). This product allows the samaritan PAD to deliver lower-energy therapy to children from 1 year of age to 8 years or up to 55lbs (25kg).

DO NOT DELAY THERAPY IF YOU ARE NOT SURE OF EXACT AGE OR WEIGHT.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K052465: Reduced Energy Pediatric Electrode Pads for the Samaritan Aed and Pad, Models Sde 251, SPP 351.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K052465

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.