510(k) · Cleared
Recalled · 2006Smartsuction Harmony Device
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Haemonetics Corp. · BTA Pump, Portable, Aspiration (Manual Or Powered)
Cleared on Oct 5, 2005 as substantially equivalent to K021368 (Medela AG, 2002) after a 12-day review. Its family tree reaches back 1 generation to 2002. 1 later device descends from it. It has been recalled once.
- Decided
- Oct 5, 2005
- Review
- 12 days
- Class
- Class II
- Product code
- BTA
- Recalls
- 1
- Predicates
- 1
Its longest line of descent runs back 1 generation to K021368 (2002). 1 later device descends from it, 1 citing it directly.
Most influential descendants
Showing 1 of 1 · See all- K062801Smartsuction Solo DeviceHaemonetics Corp. · 2006 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K021368 | Medela Basic 30 and Dominant 50 Suction Pumps, Models 037, 057 | Medela AG | 2002 | — |
| 0 (this device) | K052626 | Smartsuction Harmony Device | Haemonetics Corp. | 2005 | — |
| +1 | K062801 | Smartsuction Solo Device | Haemonetics Corp. | 2006 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2006 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in BTA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new BTA device
- #1 K222547 Electric nasal aspiratorscore 20.7 · 6 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K021368 | Medela Basic 30 and Dominant 50 Suction Pumps, Models 037, 057Summary page 1…cation Name: AC Powered Suction Pump Predicate Device The predicate device is the Medela® Basic 30 Fluid Management System. The Medela device was cleared under K021368 on 5/15/02. Description Powered suction pumps are described in FDA regulations, 21 CFR 878.4780, as: "A powered suction pu… | 2002 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K062801 | Smartsuction Solo Device | 2006 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jul 28, 2006 | Not classified | Z-0035-2007 | Other | Faulty circuit board may short circuit and cause electric shock to operator |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All BTA devicesBTA Pump, Portable, Aspiration (Manual Or Powered): FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · BTA
7so far in 2026, the most since 2007
2026 to Oct 25, outlined.
Median 510(k) review, days · BTA
87in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · BTA
1in 2021
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · BTA
13in 2025
MAUDE reports naming product code BTA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 1 | — | 1 | 13 |
| 2008 | 3 | 135 | 3 | 5 |
| 2009 | 2 | — | 0 | 14 |
| 2010 | 2 | — | 3 | 178 |
| 2011 | 1 | — | 0 | 62 |
| 2012 | 1 | — | 0 | 47 |
| 2013 | 3 | 45 | 0 | 18 |
| 2014 | 3 | 91 | 0 | 399 |
| 2015 | 0 | — | 0 | 15 |
| 2016 | 3 | 149 | 0 | 25 |
| 2017 | 1 | — | 0 | 4 |
| 2018 | 4 | 125 | 0 | 17 |
| 2019 | 0 | — | 0 | 14 |
| 2020 | 1 | — | 0 | 10 |
| 2021 | 1 | — | 1 | 9 |
| 2022 | 3 | 58 | 0 | 7 |
| 2023 | 4 | 265 | 0 | 8 |
| 2024 | 5 | 131 | 0 | 13 |
| 2025 | 6 | 87 | 0 | 13 |
| 2026 | 7 | 89 | 0 | — |
What is it indicated for?
The SmartSuction HARMONY device is an AC-powered suction device intended to be used as a suction source to replace wall suction in the hospital operating room to remove fluids and debris from the surgical field during surgical procedures.
The SmartSuction HARMONY device is not intended for endotracheal suction. The SmartSuction HARMONY device automatically adjusts suction up to 150 mmHg negative pressure when the suction tip is occluded. Do not use SmartSuction HARMONY device as a suction source in any procedure where 150 mmHg of suction could damage underlying tissue.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K052626 · printed Nov 6, 2026
Cite this page
510(k) Report. “K052626: Smartsuction Harmony Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K052626
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026