# K052640: Gemini Raptor

- Pathway: 510(k) (Traditional)
- Applicant: Philips Medical Systems (Cleveland), Inc.
- Decided: Oct 7, 2005 after 11 days
- Product code: KPS (System, Tomography, Computed, Emission)
- Device class: 2
- Regulation: 21 CFR 892.1200
- Recalls of this device: 36 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 8 generations back to K930090 (1993). 21 ancestors in all.
12 later devices descend from it; 6 cite it directly.
0 of 34 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K051170: Gemini GXL, Philips Medical Systems (Cleveland), Inc. (2005)

## Most influential descendants
- K123599: Ingenuity Digital Pet/CT (2013), 4 descendants of its own
- K210880: Vereos PET/CT (2021), 2 descendants of its own
- K140651: Celesteion, Pca-9000A/2 (2014), 2 descendants of its own
- K170490: Celesteion, PCA-9000A/3 (2017), 1 descendant of its own
- K251401: PennPET Explorer Positron Emission Tomograph (2025), 0 descendants of its own

## Recalls of this device
- May 1, 2021 · Class II · Z-0279-2022: Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, b
- May 1, 2021 · Class II · Z-0281-2022: Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, b
- May 1, 2021 · Class II · Z-1987-2021: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the
- Mar 28, 2018 · Class II · Z-2276-2018: A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or user
- Feb 15, 2018 · Class II · Z-1392-2018: Following a period of inactivity, the mass storage device may cause the acquisition console to become unresponsive. This may prevent an acquisition from proceeding.
- Nov 1, 2017 · Class II · Z-0522-2018: One of the two wire channels mounted inside the Gantry Separation Unit (GSU) that is also used to secure the top two GSU covers, was missing mounting hardware.
- Jul 1, 2015 · Class II · Z-2194-2015: Philips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.
- Jul 1, 2015 · Class II · Z-2196-2015: Philips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.
- Jul 1, 2015 · Class II · Z-2197-2015: Philips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.
- Jul 1, 2015 · Class II · Z-2195-2015: Philips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.
- Jul 1, 2015 · Class II · Z-2198-2015: Philips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.
- Mar 6, 2015 · Class II · Z-1787-2015: The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overh
- Mar 6, 2015 · Class II · Z-1785-2015: The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overh
- Mar 6, 2015 · Class II · Z-1783-2015: The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overh
- Mar 6, 2015 · Class II · Z-1788-2015: The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overh
- Mar 6, 2015 · Class II · Z-1782-2015: The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overh
- Mar 6, 2015 · Class II · Z-1786-2015: The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overh
- Jun 18, 2014 · Class II · Z-2037-2014: When beginning a procedure to start the gantry and CT Host, a Close Estop message may appear to the operator, but disappears before action can be taken. If this occurs it results in an inability to c
- Dec 4, 2013 · Class II · Z-0932-2014: The PET Reconstruction Server (PRS) database may randomly lock up after the low dose CT acquisition is complete and will not allow the PET acquisition to begin, resulting in an unexpected incomplete s
- Aug 16, 2013 · Class II · Z-0067-2014: Issues with the Ingenuity TF PET/CT Software Version 4.0 could pose a risk for patients in the event that a patient rescan were to require reinjection of PET radiopharmaceutical.
- Jan 21, 2013 · Class II · Z-0897-2013: The system may not default to the same gating trigger with which the data were acquired. This could cause a risk for patients.
- Dec 20, 2012 · Class II · Z-1029-2013: Issues have been detected in the Ingenuity TF PET/CT currently running software version 4.0.60605.0 that if they were to re-occur, could cause a risk for patients, users or service personnel, includin
- Oct 24, 2012 · Class II · Z-0882-2013: Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory 
- Aug 8, 2012 · Class II · Z-2239-2012: Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casing on one of the batteries. Supplier of the UPS, Chloride/Emerson, determined the
- Jan 26, 2012 · Not classified · Z-1369-2012: TumorLOC will save the incorrect contour when the user saves the contour while in the edit mode, and the software does not force the user to exit the edit mode before saving. In TumorLOC, when saving 
- Dec 13, 2011 · Not classified · Z-1056-2012: A number of non-conformances have been noted in the TruFlight Select PET/CT System. A number of these non-conformances have been corrected to improve the overall quality and functionality of the suspe
- Dec 2, 2011 · Not classified · Z-0805-2012: Philips received notification by a customer of an issue relating to unwanted auto delete occurrences resulting in raw data being deleted and, in some cases, raw data being deleted immediately after sc
- Oct 24, 2011 · Not classified · Z-0937-2012: Philips became aware of an issue relating to a UPS cabinet that contained a melted battery which caused damage to other batteries lower in the cabinet, leading to battery acid leaking on the floor.
- May 31, 2011 · Not classified · Z-3200-2011: The firm is conducting a recall/software correction to several of their GEMINI PET/CT X-ray system units in the field. The firm believes that if these software defects were to recur, it could impact t
- May 31, 2011 · Not classified · Z-3199-2011: The firm is conducting a recall/software correction to several of their GEMINI PET/CT X-ray system units in the field. The firm believes that if these software defects were to recur, it could impact t
- Feb 19, 2009 · Not classified · Z-1580-2009: A software anomaly in the TumorLoc software for the firm's Brilliance CT scanners was discovered. The same software anomaly was subsequently determined to be present in the firm's GEMINI PET/CT units.
- Jan 26, 2009 · Not classified · Z-0261-2010: Upper patient pallet component of the recalling firm's Gemini PET/CT system may be thinner than normally required which may reduce the product's load bearing capacity, resulting in cracking under less
- Nov 24, 2008 · Not classified · Z-0287-2010: A leak between the Tube Interface and the Outlet Tube of the transducer for the Respiratory Gating System component may result in a failure of the device to produce respiratory correlated images.
- May 1, 2008 · Class II · Z-0197-2016: The reason for this inspection is to determine if the Spindle Block Servo Controllers on the CT subsystem of potentially affected GEMINI PET/CT systems are susceptible to premature failure.
- Sep 21, 2007 · Not classified · Z-0462-2008: Artifacts: Artifacts may appear in some images from head scans on the CT subsystem of the GEMINI PET/CT system. Reading images with these artifacts may lead to misdiagnosis or improper treatment.
- Sep 21, 2007 · Not classified · Z-0461-2008: Artifacts: Artifacts may appear in some images from head scans on the CT subsystem of the GEMINI PET/CT system. Reading images with these artifacts may lead to misdiagnosis or improper treatment.

## Indications for use
The device is a diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem produces cross-sectional images of the body by computer reconstruction of x-ray transmission data. The PET subsystem produces images of the distribution of PET radiopharmaceuticals in the patient body (specific radiopharmaceuticals are used for whole body, brain, heart and other organ imaging). Attenuation correction is accomplished by CTAC. The device also provides for list mode, dynamic, and gated acquisitions. Image processing and display workstations provide software applications to process, analyze, display, quantify and interpret medical images/data. The PET and CT images may be registered and displayed in a "fused" (overlaid in the same spatial orientation) format to provide combined metabolic and anatomical data at different angles. Trained professionals use the images in: o The evaluation, detection and diagnosis of lesions, disease and organ function such as but not limited to cancer, cardiovascular disease, and neurological disorders. o The detection, localization, and staging of tumors and diagnosing cancer patients. o Treatment planning and interventional radiology procedures. The device includes software that provides a quantified analysis of regional cerebral activity from PET images. Cardiac imaging software provides functionality for the quantification of cardiology images and datasets including but not limited to myocardial perfusion for the display of wall motion and quantification of left-ventricular function parameters from gated myocardial perfusion studies and for the 3D alignment of coronary artery images from CT coronary angiography onto the myocardium. Both subsystems (PET and CT) can also be operated independently as fully functional, diagnostic imaging systems including application of the CT scanner as a radiation therapy simulation scanner.