# K052792: Anthology Hip Stem

- Pathway: 510(k) (Special)
- Applicant: Smith & Nephew, Inc.
- Decided: Oct 7, 2005 after 4 days
- Product code: JDI (Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented)
- Device class: 2
- Regulation: 21 CFR 888.3350
- Recalls of this device: 1 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
129 later devices descend from it; 19 cite it directly.
0 of 130 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. The FDA summary names its predicate without a number.
- (no number) SYNERGY Hip Stem: named in the summary without a submission number

## Most influential descendants
- K112802: Pipeline Total Hip System (2012), 53 descendants of its own
- K072417: Smith & Nephew Mis Hip Stem (2008), 44 descendants of its own
- K090982: Smith & Nephew Hip Systems with HA (2010), 24 descendants of its own
- K101072: Tritanium Peri-Apatitie Acetabular Solid-Backed Shell, Tritanium Peri-Apatite (2011), 22 descendants of its own
- K122158: PBP Total Hip System (2012), 21 descendants of its own

## Recalls of this device
- May 2, 2018 · Class II · Z-1921-2018: Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.

## Indications for use
The ANTHOLOGY Hip Stem is indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant. Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, arthritis secondary to a variety of diseases and anomalies, and congenital dysplasia; old, remote osteomyetitis with an extended drainage-free period, in which case, the patient should be warned of an above normal danger of infection postoperatively; treatments of nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques: endoprosthesis, femoral osteotomy, or Girdlestone resection; fracture-dislocation of the hip; and correction of deformity. The ANTHOLOGY Hip Stem is for single use only.