510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

RX Imola

K052914510(k) TraditionalCleared Mar 31, 2006168-day reviewClass II21 CFR 862.1600Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Randox Laboratories, Ltd. · CEM Electrode, Ion Specific, Potassium

Cleared on Mar 31, 2006 after a 168-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 2 later devices descend from it. It has been recalled 4 times.

Decided
Mar 31, 2006
Review
168 days
Class
Class II
Product code
CEM
Recalls
4
Predicates
0

2 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2023: 1 descendant cleared2014: 1 descendant cleared20052010201520202025
  1. This deviceK052914 RX ImolaRandox Laboratories, Ltd. · 2006
  2. Descendant · cites it directly · 0 of its ownK230890 ISE ElectrodesRandox Laboratories, Ltd. · 2023
  3. Descendant · cites it directly · 0 of its ownK131554 RX Daytona Plus Chemistry Analyzer; RX Daytona Plus Aspartate…Randox Laboratories, Ltd. · 2014 · Recalled

Most influential descendants

Showing 2 of 2 · See all
  • K230890ISE ElectrodesRandox Laboratories, Ltd. · 2023 · cites it directly0descendants
  • K131554RX Daytona Plus Chemistry Analyzer; RX Daytona Plus Aspartate Aminotransferase (Ast)…Randox Laboratories, Ltd. · 2014 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K052914RX ImolaRandox Laboratories, Ltd.2006—
+1K230890ISE ElectrodesRandox Laboratories, Ltd.2023—
+1K131554RX Daytona Plus Chemistry Analyzer; RX Daytona Plus Aspartate Aminotransferase…Randox Laboratories, Ltd.20143

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
202310
FIG. 1 — Lineage of K052914, 2006–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CEM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CEM device

  1. #1 K253491 ISE indirect K for Gen.2; ISE indirect Na for…score 11.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K230890ISE ElectrodesRandox Laboratories, Ltd.20230
K131554RX Daytona Plus Chemistry Analyzer; RX Daytona Plus Aspartate Aminotransferase (Ast)…Randox Laboratories, Ltd.20140

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 6, 2023Class IIZ-0200-2024Under Investigation by firmRX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when…
Jun 24, 2019Class IIZ-2084-2019Under Investigation by firmThere was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument.
May 4, 2018Class IIZ-2059-2018Under Investigation by firmAbnormal premature termination of the software could affect the use of the analyser when running patient samples. The impact being a delay in reporting test results.
Apr 14, 2016Class IIZ-0725-2017Device DesignAccording to the firm, Carry over was observed when the amylase or pancreatic amylase test is run directly before or after the Magnesium assay. An update is required to the RX Imola analyser running order and an acid wash recommended when…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K230890 (Randox Laboratories, Ltd., 2023) has gone furthest, with 0 descendants of its own.

How does its category compare?

All CEM devices

CEM Electrode, Ion Specific, Potassium: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CEM

1in 2025

2007: 12009: 12010: 32011: 32012: 42020: 12023: 22025: 12026 (partial year): 1
2007high 4 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CEM

180in 2012

2010: 195 days median2011: 74 days median2012: 180 days median
2007high 195 (2010) · low 74 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CEM

4in 2025

2008: 12010: 122011: 62016: 12019: 12020: 32021: 12023: 22025: 4
2007high 12 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CEM

29in 2025

2007: 662008: 2452009: 792010: 1132011: 1262012: 452013: 372014: 42015: 72016: 62017: 322018: 132019: 252020: 272021: 182022: 162023: 302024: 422025: 29
2007high 245 (2008) · low 4 (2014)2025

MAUDE reports naming product code CEM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—066
20080—1245
20091—079
2010319512113
20113746126
20124180045
20130—037
20140—04
20150—07
20160—16
20170—032
20180—013
20190—125
20201—327
20210—118
20220—016
20232—230
20240—042
20251—429
20261—0—

What is it indicated for?

The RX Imola is a medium-sized desktop fully Automated Clinical Chemistry Analyzer complete with Ion Selective Electrode (ISE) Unit and dedicated analyzer software. An external PC operates the analyzer and results can be printed as required. The analyzer may be connected to a host computer, when required. The analyzer can be used to run tests such as magnesium in serum and plasma samples. Magnesium measurements are used in the diagnosis and treatment of…

Read the full statement (142 words)

The RX Imola is a medium-sized desktop fully Automated Clinical Chemistry Analyzer complete with Ion Selective Electrode (ISE) Unit and dedicated analyzer software. An external PC operates the analyzer and results can be printed as required. The analyzer may be connected to a host computer, when required.

The analyzer can be used to run tests such as magnesium in serum and plasma samples. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia and hypermagnesemia. Various other clinical chemistry assays are adaptable to the analyzer.

The ISE Unit on the RX Imola can be used for measurement of the electrolytes sodium, potassium and chloride in serum, plasma and urine and for use in diagnosis and treatment of electrolyte imbalance.

The RX Imola analyzer must only be used by suitably qualified personnel, under appropriate laboratory conditions.

For in vitro diagnostic use only.

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K052914: RX Imola.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K052914

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026