510(k) Report

510(k) · Cleared

Recalled · 2007

CPS Direct SL Outer Catheter

K053277510(k) TraditionalCleared Feb 24, 200693-day reviewClass II21 CFR 870.1250Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

St Jude Medical · DQY Catheter, Percutaneous

Cleared on Feb 24, 2006 as substantially equivalent to K021587 (Medtronic Vascular, 2002) and 3 other predicates after a 93-day review. Its family tree reaches back 6 generations to 1989. It has been recalled once.

Decided
Feb 24, 2006
Review
93 days
Class
Class II
Product code
DQY
Recalls
1
Predicates
4

Its longest line of descent runs back 6 generations to K893658 (1989). 48 ancestors in all, the oldest K760385 (1976). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2003: 1 other ancestor2002: 4 other ancestors2001: 2 other ancestors1999: 2 other ancestors1996: 2 other ancestors1995: 5 other ancestors1994: 4 other ancestors1992: 4 other ancestors1990: 5 other ancestors1980: 1 other ancestor1976: 1 other ancestor1998: 1 other ancestor1993: 1 other ancestor1991: 1 other ancestor1988: 1 other ancestor1987: 3 other ancestors1986: 1 other ancestor1985: 2 other ancestors1977: 1 other ancestorK893658 · Percutaneous Introducer · Medamicus, Inc. · 1989K965167 · Percutaneous Venous Introducer 10264 · Medamicus, Inc. · 1997K012130 · Attain LDS 6216 Left-Heart Delivery System · Medtronic Vascular · 2001K013517 · 10600 Deflectable Catheter System, Model 10600 · Medtronic Vascular · 2002K031211 · Attain Prevail Steerable Catheter Set, Model 6228CTH · Medtronic Vascular · 2003K042194 · Attain Select 6238TEL Guide Catheter Set for Left-Heart Delivery · Medtronic Vascular · 20041975198019851990199520002005
  1. Root · 6 generations backK893658 Percutaneous IntroducerMedamicus, Inc. · 1989
  2. 5 generations backK965167 Percutaneous Venous Introducer 10264Medamicus, Inc. · 1997
  3. 4 generations backK012130 Attain LDS 6216 Left-Heart Delivery SystemMedtronic Vascular · 2001
  4. 3 generations backK013517 10600 Deflectable Catheter System, Model 10600Medtronic Vascular · 2002
  5. 2 generations backK031211 Attain Prevail Steerable Catheter Set, Model 6228CTHMedtronic Vascular · 2003
  6. 1 generation backK042194 Attain Select 6238TEL Guide Catheter Set for Left-Heart DeliveryMedtronic Vascular · 2004
  7. This deviceK053277 CPS Direct SL Outer CatheterSt Jude Medical · 2006
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K893658Percutaneous IntroducerMedamicus, Inc.1989—
−5K965167Percutaneous Venous Introducer 10264Medamicus, Inc.1997—
−4K012130Attain LDS 6216 Left-Heart Delivery SystemMedtronic Vascular2001—
−3K01351710600 Deflectable Catheter System, Model 10600Medtronic Vascular2002—
−2K031211Attain Prevail Steerable Catheter Set, Model 6228CTHMedtronic Vascular2003—
−1K042194Attain Select 6238TEL Guide Catheter Set for Left-Heart DeliveryMedtronic Vascular2004—
0 (this device)K053277CPS Direct SL Outer CatheterSt Jude Medical2006—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197610
197710
198010
198520
198610
198730
198810
199050
199110
199240
199310
199440
199550
199620
199810
199920
200120
200240
200310
FIG. 1 — Lineage of K053277, 1989–2006Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DQY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQY device

  1. #1 K234115 APRO 55 Catheterscore 17.5 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K021587Attain LDS 6216A Left-Heart Delivery System, Model 6216A
Summary page 1
…Percuatenous Classification: Class 11, 21 CFR 870.1250 Legally marketed device to which your firm is claiming equivalence: Medtronic AttainTM LDS (K021587) Attain Access (K021589), Guidant RapidoTM Cut-Away/Advance (K031505) Device Description: The out…
Medtronic Vascular20021
*K021589Attain Access 6218A Left-Heart Delivery System, Model 6218
Summary page 1
…ion: Class 11, 21 CFR 870.1250 Legally marketed device to which your firm is claiming equivalence: Medtronic AttainTM LDS (K021587) Attain Access (K021589), Guidant RapidoTM Cut-Away/Advance (K031505) Device Description: The outer catheter is an introd…
Medtronic Vascular20021
*K031505Rapido Cut-Away Guiding Catheter
Summary page 1
…ich your firm is claiming equivalence: Medtronic AttainTM LDS (K021587) Attain Access (K021589), Guidant RapidoTM Cut-Away/Advance (K031505) Device Description: The outer catheter is an introducer that is used to cannulate the Coronary Sinus (CS). The CPS Direct SL…
Guidant Corporation20031
*K042194Attain Select 6238TEL Guide Catheter Set for Left-Heart Delivery
Summary page 2
…ilar technology, have similar intended uses, functions, materials and methods of operation as the following predicate devices. · Medtronic AttainTM LDS/Access, K042194, K021589 * Guidant RapidoTM Cut-Away/Advance, K031505 Summary of Studies: Device comparison testing was performed to su…
Medtronic Vascular2004—

Also cleared against K021587: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 20, 2007Not classifiedZ-0933-2007OtherDevice Separation- product may separate at the mapping collar transition joint located at the proximal (light blue) marker band end of the tool during the implant procedure.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 3 companies; a new generation was cleared every 1.4 years on median.
  • 10 of the 48 ancestors have a recall of any class.
  • 4 other devices also cite its predicate K021587.

How does its category compare?

All DQY devices

DQY Catheter, Percutaneous: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DQY

21in 2025

2007: 162008: 312009: 372010: 222011: 362012: 302013: 322014: 192015: 272016: 222017: 262018: 252019: 302020: 202021: 182022: 182023: 182024: 162025: 212026 (partial year): 11
2007high 37 (2009) · low 16 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQY

115in 2025

2007: 85 days median2008: 66 days median2009: 73 days median2010: 74 days median2011: 71 days median2012: 87 days median2013: 92 days median2014: 127 days median2015: 118 days median2016: 117 days median2017: 113 days median2018: 106 days median2019: 111 days median2020: 120 days median2021: 81 days median2022: 74 days median2023: 114 days median2024: 87 days median2025: 115 days median2026 (partial year): 30 days median
2007high 127 (2014) · low 66 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQY

1in 2025

2007: 62008: 122009: 172010: 322011: 62012: 22013: 36, 1 Class I2014: 32015: 32016: 22, 7 Class I2017: 82018: 22019: 12020: 22021: 22023: 172024: 12025: 1, 1 Class I2026 (partial year): 2, 1 Class I
2007high 36 (2013) · low 0 (2022)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · DQY

3,056in 2025

2007: 3132008: 4962009: 8192010: 1,2932011: 1,2112012: 1,1912013: 1,3522014: 1,4972015: 2,3372016: 2,2522017: 2,4892018: 2,8552019: 2,9022020: 2,8892021: 3,0062022: 2,6432023: 3,0492024: 3,1902025: 3,056
2007high 3,190 (2024) · low 313 (2007)2025

MAUDE reports naming product code DQY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200716856313
2008316612496
2009377317819
20102274321,293
2011367161,211
2012308721,191
2013329236 (1)1,352
20141912731,497
20152711832,337
20162211722 (7)2,252
20172611382,489
20182510622,855
20193011112,902
20202012022,889
2021188123,006
2022187402,643
202318114173,049
2024168713,190
2025211151 (1)3,056
202611302 (1)—

What is it indicated for?

The St. Jude Medical Slittable DirectTM Outer Guide Catheter is designed for intracardiac access of the venous system of the heart and to serve as a conduit during implantation for the delivery of contrast medium and St. Jude Medical devices, including implantable left heart leads and delivery tools, and support of fluids where minimizing blood loss is essential. In addition, the outer guide catheters can work with inner catheters as a system.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K053277: CPS Direct SL Outer Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K053277

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026