Cool-Tip RF Generator, Cool-Tip RF System Accessories

K053290510(k) TraditionalCleared Mar 15, 2006110-day reviewClass II21 CFR 878.4400

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Valleylab · GEI Electrosurgical, Cutting & Coagulation & Accessories

Cleared on Mar 15, 2006 as substantially equivalent to K042216 (Valleylab, 2004) after a 110-day review. Its family tree reaches back 5 generations to 1987. 50 later devices descend from it.

Decided
Mar 15, 2006
Review
110 days
Class
Class II
Product code
GEI
Recalls
None
Predicates
1

Its longest line of descent runs back 5 generations to K870028 (1987). 16 ancestors in all. 50 later devices descend from it, 6 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 2 other ancestors2001: 2 other ancestors2000: 2 other ancestors1999: 3 other ancestors1998: 1 other ancestor1997: 1 other ancestor2010: 1 descendant cleared2012: 3 descendants cleared2007: 1 descendant cleared2016: 3 descendants cleared2008: 1 descendant cleared2015: 3 descendants cleared2022: 2 descendants cleared2020: 7 descendants cleared2014: 1 descendant cleared2024: 8 descendants cleared2013: 1 descendant cleared2017: 2 descendants cleared2019: 2 descendants cleared2023: 6 descendants cleared2018: 2 descendants cleared2021: 3 descendants cleared2025: 2 descendants cleared2026: 2 descendants clearedK870028 · SMK Sluyter-Mehta Cannulae, RF Pole, Pole,Flexible · Radionics, Inc. · 1987K963577 · Radionics Sluijter-Mehta Cannula · Radionics, Inc. · 1996K980430 · Radionics Disposable RF Cannulae (SC-C, RFK-DB, RFK-DS) · Radionics, Inc. · 1998K021329 · Rita System · Rita Medical Systems · 2002K042216 · Cool-Tip RF Ablation System · Valleylab · 20041990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 6 other ancestors3 generations back: 5 other ancestors 1 generation forward: 6 descendants2 generations forward: 11 descendants3 generations forward: 11 descendants4 generations forward: 9 descendants5 generations forward: 10 descendants6 generations forward: 3 descendants K042216 · Cool-Tip RF Ablation System · Valleylab · 2004K021329 · Rita System · Rita Medical Systems · 2002K980430 · Radionics Disposable RF Cannulae (SC-C, RFK-DB, RFK-DS) · Radionics, Inc. · 1998K963577 · Radionics Sluijter-Mehta Cannula · Radionics, Inc. · 1996K870028 · SMK Sluyter-Mehta Cannulae, RF Pole, Pole,Flexible · Radionics, Inc. · 1987−50+5
  1. Root · 5 generations backK870028 SMK Sluyter-Mehta Cannulae, RF Pole, Pole,FlexibleRadionics, Inc. · 1987
  2. 4 generations backK963577 Radionics Sluijter-Mehta CannulaRadionics, Inc. · 1996
  3. 3 generations backK980430 Radionics Disposable RF Cannulae (SC-C, RFK-DB, RFK-DS)Radionics, Inc. · 1998
  4. 2 generations backK021329 Rita SystemRita Medical Systems · 2002
  5. 1 generation backK042216 Cool-Tip RF Ablation SystemValleylab · 2004
  6. This deviceK053290 Cool-Tip RF Generator, Cool-Tip RF System AccessoriesValleylab · 2006
  7. Descendant · cites it directly · 42 of its ownK072687 Valleylab Microwave Ablation GeneratorCovidien Lp, Formerly Know AS Valleylab, A Divison · 2008
  8. Descendant · cites it directly · 39 of its ownK100744 Certus 140 Microwave Ablation System and Accessories, Model Certus 140Neuwave Medical, Inc. · 2010
  9. Descendant · 2 generations on · 38 of its ownK113237 Certus 140 2.45GHZ Ablation SystemNeuwave Medical, Inc. · 2012
  10. Descendant · 2 generations on · 29 of its ownK133821 Emprint Ablation SystemCovidien, LLC · 2014 · Recalled
  11. Descendant · 3 generations on · 13 of its ownK163105 Emprint Ablation SystemCovidien, LLC · 2016 · Recalled

Showing 5 of 50 descendants. See all descendants

Descendants cleared per year

2007: 12008: 12010: 12012: 32013: 12014: 12015: 32016: 32017: 22018: 22019: 22020: 72021: 32022: 22023: 62024: 82025: 22026: 2
2007high 8 (2024) · low 0 (2009)2026

Most influential descendants

Showing 5 of 50 · See all
  • K072687Valleylab Microwave Ablation GeneratorCovidien Lp, Formerly Know AS Valleylab, A Divison · 2008 · cites it directly42descendants
  • K100744Certus 140 Microwave Ablation System and Accessories, Model Certus 140Neuwave Medical, Inc. · 2010 · cites it directly39descendants
  • K113237Certus 140 2.45GHZ Ablation SystemNeuwave Medical, Inc. · 2012 · 2 generations on38descendants
  • K133821Emprint Ablation SystemCovidien, LLC · 2014 · 2 generations on29descendants
  • K163105Emprint Ablation SystemCovidien, LLC · 2016 · 3 generations on13descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K870028SMK Sluyter-Mehta Cannulae, RF Pole, Pole,FlexibleRadionics, Inc.1987—
−4K963577Radionics Sluijter-Mehta CannulaRadionics, Inc.1996—
−3K980430Radionics Disposable RF Cannulae (SC-C, RFK-DB, RFK-DS)Radionics, Inc.1998—
−2K021329Rita SystemRita Medical Systems2002—
−1K042216Cool-Tip RF Ablation SystemValleylab2004—
0 (this device)K053290Cool-Tip RF Generator, Cool-Tip RF System AccessoriesValleylab2006—
+1K072687Valleylab Microwave Ablation GeneratorCovidien Lp, Formerly Know AS Valleylab, A Divison2008—
+1K100744Certus 140 Microwave Ablation System and Accessories, Model Certus 140Neuwave Medical, Inc.2010—
+2K113237Certus 140 2.45GHZ Ablation SystemNeuwave Medical, Inc.2012—
+2K133821Emprint Ablation SystemCovidien, LLC20147
+3K163105Emprint Ablation SystemCovidien, LLC20163

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199710
199810
199930
200020
200120
200320

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
200810
201010
201230
201310
201410
201530
201630
201720
201820
201920
202070
202130
202220
202360
202480
202520
202620
FIG. 1 — Lineage of K053290, 1987–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GEI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEI device

  1. #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042216Cool-Tip RF Ablation System
Summary page 1
…edicate Devices The Cool-tip T" RF System of this submission is substantially equivalent to the Cool-tip T" RF System covered by 510(k) notification K042216 dated 11/12/04. 4. Device Description The Cool-tip-' RF System consists of a microprocessor based RF generator.…
Valleylab2004—

Also cleared against K042216: 5 other devices →

What descends from it? (50)

See all 50

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K072687Valleylab Microwave Ablation GeneratorCovidien Lp, Formerly Know AS Valleylab, A Divison200842
K100744Certus 140 Microwave Ablation System and Accessories, Model Certus 140Neuwave Medical, Inc.201039
K150112HS Amica Devices FamilyH.S Hospital Service S.P.A20152
K070446Cool-Tip Switching Controller, CTSW ControlValleylab,A Division of Tyco Healthcare Group LP20072
K150019TheraVision Ultrasound Ablation System and ACOUSTx ApplicatorsAcoustic Medsystems, Inc.20160
K120116Serf Ablation System Serf Ablation NeedleThermedical20120

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 19 years and passes through 3 companies; a new generation was cleared every 2.1 years on median.
  • 4 of the 16 ancestors have a recall of any class.
  • Of its descendants, K072687 (Covidien Lp, Formerly Know AS Valleylab, A Divison, 2008) has gone furthest, with 42 descendants of its own.
  • 5 other devices also cite its predicate K042216.

How does its category compare?

All GEI devices

GEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GEI

52in 2025

2007: 502008: 612009: 422010: 462011: 352012: 392013: 482014: 492015: 472016: 642017: 522018: 472019: 482020: 592021: 462022: 552023: 652024: 592025: 522026 (partial year): 43
2007high 65 (2023) · low 35 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEI

113in 2025

2007: 51 days median2008: 71 days median2009: 130 days median2010: 109 days median2011: 135 days median2012: 133 days median2013: 136 days median2014: 103 days median2015: 123 days median2016: 108 days median2017: 72 days median2018: 93 days median2019: 93 days median2020: 105 days median2021: 146 days median2022: 150 days median2023: 104 days median2024: 125 days median2025: 113 days median2026 (partial year): 183 days median
2007high 183 (2026) · low 51 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEI

4in 2025

2007: 212008: 192009: 272010: 512011: 322012: 302013: 132014: 32015: 42016: 32017: 22018: 62019: 52020: 12021: 62022: 492023: 27, 4 Class I2024: 22, 10 Class I2025: 42026 (partial year): 8
2007high 51 (2010) · low 1 (2020)2026

Class I (14 since 2007) Class II/III not classified

Adverse event reports per year · GEI

7,592in 2025

2007: 2,7952008: 2,5562009: 3,5912010: 3,3772011: 4,4632012: 5,2162013: 4,8972014: 4,9082015: 4,9362016: 5,6162017: 6,0732018: 6,5452019: 7,0672020: 6,4902021: 6,3542022: 6,1032023: 7,1762024: 7,8972025: 7,592
2007high 7,897 (2024) · low 2,556 (2008)2025

MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075051212,795
20086171192,556
200942130273,591
201046109513,377
201135135324,463
201239133305,216
201348136134,897
20144910334,908
20154712344,936
20166410835,616
2017527226,073
2018479366,545
2019489357,067
20205910516,490
20214614666,354
202255150496,103
20236510427 (4)7,176
20245912522 (10)7,897
20255211347,592
2026431838—

What is it indicated for?

The Cool-tipTM RF Ablation System (Generator and accessories) is intended for use in percutaneous, laparoscopic and intraoperative coagulation and ablation of tissue, such as partial or complete ablation of non-resectable liver tumors.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K053290: Cool-Tip RF Generator, Cool-Tip RF System Accessories.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K053290

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026