510(k) Report

510(k) · Cleared

Recalled · 2010

Amplatzer Sizing Balloon II

K053440510(k) SpecialCleared Dec 30, 200521-day reviewClass II21 CFR 870.4450Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Aga Medical Corp. · MJN Catheter, Intravascular Occluding, Temporary

Cleared on Dec 30, 2005 as substantially equivalent to K993248 (Aga Medical Corp., 2000) and 1 other predicate after a 21-day review. Its family tree reaches back 3 generations to 1994. It has been recalled 2 times.

Decided
Dec 30, 2005
Review
21 days
Class
Class II
Product code
MJN
Recalls
2
Predicates
2

Its longest line of descent runs back 3 generations to K941269 (1994). 6 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 3 generations backK941269 Gore Smoother II and Gore BPTBW. L. Gore & Associates, Inc. · 1994
  2. 2 generations backK962765 Cordis Marker Wire Steerable Guidewire (Stabilizer with Radiopaque…Cordis Corp. · 1997
  3. 1 generation backK993248 Amplatzer Sizing BalloonAga Medical Corp. · 2000
  4. This deviceK053440 Amplatzer Sizing Balloon IIAga Medical Corp. · 2005
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K941269Gore Smoother II and Gore BPTBW. L. Gore & Associates, Inc.1994—
−2K962765Cordis Marker Wire Steerable Guidewire (Stabilizer with Radiopaque Markers)…Cordis Corp.1997—
−1K993248Amplatzer Sizing BalloonAga Medical Corp.2000—
0 (this device)K053440Amplatzer Sizing Balloon IIAga Medical Corp.2005—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
199610
200110
FIG. 1 — Lineage of K053440, 1994–2005Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MJN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MJN device

  1. #1 K243795 pREBOA-PRO Catheterscore 12.1 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K993248Amplatzer Sizing Balloon
Summary page 2
…to currently marketed balloon catheters for temporary vessel occlusion: - AMPLATZER Sizing Balloon (K993248) - Numed PTA-OS Sizing Balloon (K003320) Performance Standards: No performance st…
Aga Medical Corp.2000—
*K003320Pta-OS Sizing Balloon, Model 360
Summary page 2
…for temporary vessel occlusion: - AMPLATZER Sizing Balloon (K993248) - Numed PTA-OS Sizing Balloon (K003320) Performance Standards: No performance standards have been developed under sectio…
NuMED, Inc.2001—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 17, 2010Not classifiedZ-1595-2011Employee errorAGA Medical is initiating a voluntary recall of a specific lot of mislabeled AMPLATZER Sizing Balloon II. AGA Medical received 2 reports from customers who received sizing balloons that were labeled as 34 mm; however, the package…
Apr 25, 2008Not classifiedZ-2193-2008Packaging process controlAGA Medical Corporation has identified an issue with outer pouch seal integrity within one production lot of 10 French 45/80 delivery systems and one lot of 24 mm Sizing Balloon. We have not observed incomplete seals with the inner pouch.…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 3 companies; a new generation was cleared every 3.4 years on median.
  • None of the 6 ancestors has a recall.

How does its category compare?

All MJN devices

MJN Catheter, Intravascular Occluding, Temporary: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MJN

2in 2025

2007: 22008: 32009: 52010: 52011: 42012: 22013: 42014: 22015: 42016: 32017: 32018: 32019: 32020: 42021: 42022: 22023: 12025: 2
2007high 5 (2009) · low 0 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MJN

116in 2021

2008: 153 days median2009: 39 days median2010: 30 days median2011: 69 days median2013: 87 days median2015: 245 days median2016: 110 days median2017: 54 days median2018: 63 days median2019: 117 days median2020: 64 days median2021: 116 days median
2007high 245 (2015) · low 30 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MJN

1in 2020

2008: 22010: 12011: 12013: 82015: 12016: 22018: 12020: 1
2007high 8 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MJN

194in 2025

2007: 252008: 262009: 252010: 352011: 922012: 532013: 762014: 1002015: 902016: 712017: 1572018: 1072019: 1392020: 1572021: 1582022: 1482023: 1812024: 2242025: 194
2007high 224 (2024) · low 25 (2007)2025

MAUDE reports naming product code MJN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—025
20083153226
2009539025
2010530135
2011469192
20122—053
2013487876
20142—0100
20154245190
20163110271
20173540157
20183631107
201931170139
20204641157
202141160158
20222—0148
20231—0181
20240—0224
20252—0194
20260—0—

What is it indicated for?

AMPLATZER® Sizing Balloon II is indicated for use in those patients with cardiovascular defects wherein accurate measurement of the defect is important to select the appropriately sized occluder device.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K053440: Amplatzer Sizing Balloon II.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K053440

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026