510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Smith & Nephew 6.5MM and 8.0MM Cannulated Screws

K060736510(k) SpecialCleared Apr 18, 200629-day reviewClass II21 CFR 888.3040Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Smith & Nephew, Inc. · HWC Screw, Fixation, Bone

Cleared on Apr 18, 2006 as substantially equivalent to K993106 (Smith & Nephew, Inc., 1999) and 3 other predicates after a 29-day review. Its family tree reaches back 2 generations to 1990. 1 of the 537 devices in its family, counting itself, has a Class I recall. 531 later devices descend from it. It has been recalled once.

Decided
Apr 18, 2006
Review
29 days
Class
Class II
Product code
HWC
Recalls
1
Predicates
4

Its longest line of descent runs back 2 generations to K903810 (1990). 5 ancestors in all. 531 later devices descend from it, 10 citing it directly. 1 of the 537 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor1999: 1 other ancestor2002: 1 other ancestor2009: 4 descendants cleared2025: 51 descendants cleared2007: 1 descendant cleared2022: 35 descendants cleared2008: 2 descendants cleared2010: 4 descendants cleared2014: 23 descendants cleared2021: 44 descendants cleared, 1 with a Class I recall12013: 8 descendants cleared2020: 36 descendants cleared2018: 24 descendants cleared2011: 5 descendants cleared2016: 16 descendants cleared2017: 39 descendants cleared2023: 45 descendants cleared2015: 26 descendants cleared2012: 11 descendants cleared2019: 29 descendants cleared2026: 67 descendants cleared2024: 61 descendants clearedK903810 · Ace Cannulated Self Tapping Cancellous Bone Screw · Ace Medical Co. · 1990K051296 · Sijf Cannulated Screw System · Depuy Spine, Inc. · 20051990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 1 other ancestor 1 generation forward: 10 descendants2 generations forward: 47 descendants3 generations forward: 105 descendants4 generations forward: 141 descendants5 generations forward: 115 descendants, 1 with a Class I recall16 generations forward: 51 descendants7 generations forward: 36 descendants8 generations forward: 16 descendants9 generations forward: 7 descendants10 generations forward: 2 descendants11 generations forward: 1 descendant K051296 · Sijf Cannulated Screw System · Depuy Spine, Inc. · 2005K903810 · Ace Cannulated Self Tapping Cancellous Bone Screw · Ace Medical Co. · 19900+5+10
  1. Root · 2 generations backK903810 Ace Cannulated Self Tapping Cancellous Bone ScrewAce Medical Co. · 1990
  2. 1 generation backK051296 Sijf Cannulated Screw SystemDepuy Spine, Inc. · 2005
  3. This deviceK060736 Smith & Nephew 6.5MM and 8.0MM Cannulated ScrewsSmith & Nephew, Inc. · 2006
  4. Descendant · cites it directly · 487 of its ownK090675 Smith & Nephew, Inc. VLP Foot Plating, Screw System and AccessoriesSmith & Nephew, Inc. · 2009
  5. Descendant · 2 generations on · 228 of its ownK100776 Synthes 2.4 MM/2.7 MM Variable Angle LCP Forefoot/Midfoot SystemSynthes (Usa) · 2010
  6. Descendant · 2 generations on · 161 of its ownK111994 Smith & Nephew Cannulated Screws and WashersSmith & Nephew, Inc. · 2011
  7. Descendant · 3 generations on · 118 of its ownK112583 Synthes Cortical ScrewsSynthes (USA) Products, LLC · 2011
  8. Descendant · cites it directly · 93 of its ownK072818 Peri-Loc Periartricular LCK Pit Sys-Prox Fem Plates/Screws.Calbe…Smith & Nephew, Inc. · 2007

Showing 5 of 531 descendants. See all descendants

Descendants cleared per year

2007: 12008: 22009: 42010: 42011: 52012: 112013: 82014: 232015: 262016: 162017: 392018: 242019: 292020: 362021: 442022: 352023: 452024: 612025: 512026: 67
2007high 67 (2026) · low 1 (2007)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 531 · See all
  • K090675Smith & Nephew, Inc. VLP Foot Plating, Screw System and AccessoriesSmith & Nephew, Inc. · 2009 · cites it directly487descendants
  • K100776Synthes 2.4 MM/2.7 MM Variable Angle LCP Forefoot/Midfoot SystemSynthes (Usa) · 2010 · 2 generations on228descendants
  • K111994Smith & Nephew Cannulated Screws and WashersSmith & Nephew, Inc. · 2011 · 2 generations on161descendants
  • K112583Synthes Cortical ScrewsSynthes (USA) Products, LLC · 2011 · 3 generations on118descendants
  • K072818Peri-Loc Periartricular LCK Pit Sys-Prox Fem Plates/Screws.Calbe AccessSmith & Nephew, Inc. · 2007 · cites it directly93descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K903810Ace Cannulated Self Tapping Cancellous Bone ScrewAce Medical Co.1990—
−1K051296Sijf Cannulated Screw SystemDepuy Spine, Inc.2005—
0 (this device)K060736Smith & Nephew 6.5MM and 8.0MM Cannulated ScrewsSmith & Nephew, Inc.2006—
+1K090675Smith & Nephew, Inc. VLP Foot Plating, Screw System and AccessoriesSmith & Nephew, Inc.2009—
+2K100776Synthes 2.4 MM/2.7 MM Variable Angle LCP Forefoot/Midfoot SystemSynthes (Usa)2010—
+2K111994Smith & Nephew Cannulated Screws and WashersSmith & Nephew, Inc.2011—
+3K112583Synthes Cortical ScrewsSynthes (USA) Products, LLC2011—
+1K072818Peri-Loc Periartricular LCK Pit Sys-Prox Fem Plates/Screws.Calbe AccessSmith & Nephew, Inc.2007—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199910
200010
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
200820
200940
201040
201150
2012110
201380
2014230
2015260
2016160
2017390
2018240
2019290
2020360
2021441
2022350
2023450
2024610
2025510
2026670
FIG. 1 — Lineage of K060736, 1990–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HWC device

  1. #1 K203011 Monster Screw Systemscore 22.2 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K993106Smith & Nephew Bone Plate System (Bone Plates, Bone Screws and Accessories)
Summary page 1
…n indications for use, design features, operational principles, and material composition when compared to the predicate devices cleared under the following submissions: K993106, Smith & Nephew Bone Plate System; K903810, Ace Medical Cannulated Screw; K000080, Howmedica Osteonics ASNIS III Cannulated Screw…
Smith & Nephew, Inc.1999—
*K903810Ace Cannulated Self Tapping Cancellous Bone Screw
Summary page 1
…rational principles, and material composition when compared to the predicate devices cleared under the following submissions: K993106, Smith & Nephew Bone Plate System; K903810, Ace Medical Cannulated Screw; K000080, Howmedica Osteonics ASNIS III Cannulated Screw System; and KO0i1296, DePuy Spine Inc. SIJ…
Ace Medical Co.1990—
*K000080Asnis Iii Cannulated Screw System
Summary page 1
…sition when compared to the predicate devices cleared under the following submissions: K993106, Smith & Nephew Bone Plate System; K903810, Ace Medical Cannulated Screw; K000080, Howmedica Osteonics ASNIS III Cannulated Screw System; and KO0i1296, DePuy Spine Inc. SIJF Cannulated Screw System.…
Howmedica Osteonics Corp.20001
*K051296Sijf Cannulated Screw System
Our reading of the summary gave K001296, FDA’s number for Spinal Cord and Immobilization Traction System (Cavendish Scott, Inc.). Corrected to the number one digit away whose FDA name carries the summary’s words for the device, “DePuy Spine Inc. SIJF Cannulated Screw System”.
Depuy Spine, Inc.2005—

Also cleared against K993106: 35 other devices →

What descends from it? (531)

See all 531

10 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K090675Smith & Nephew, Inc. VLP Foot Plating, Screw System and AccessoriesSmith & Nephew, Inc.2009487
K072818Peri-Loc Periartricular LCK Pit Sys-Prox Fem Plates/Screws.Calbe AccessSmith & Nephew, Inc.200793
K132886Smith & Nephew Variable-Angle Locking Mini-Fragment Plating SystemSmith & Nephew, Inc.201446
K124027Monster Screw SystemApogee Orthosolutions201341
K123055Smith & Nephew, Inc Screw and Plate System InstrumentationSmith & Nephew, Inc.201333
K080649Pogo ScrewFxdevices200815
K093676Omnitech Large, Trimed Compression ScrewTriMed, Inc.201013
K210837Smith & Nephew Plates and ScrewsSmith & Nephew, Inc.20217
K213126Smith & Nephew, Inc. Plates and Screws Systems: EVOS, Peri-Loc, D-Rad, TC-100, VLP…Smith & Nephew, Inc.20222
K241666EVOS Pelvic and Acetabular SystemSmith & Nephew20250

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 20, 2021Class IIZ-0143-2022Process controlThe product label incorrectly indicated that the screws are fully threaded, but the screws contained in the package were partially threaded.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K211616ExcelsiusHubGlobus Medical, Inc. · 2021 · 5 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 16 years and passes through 3 companies; a new generation was cleared every 14.9 years on median.
  • 1 of the 5 ancestors has a recall of any class.
  • Of its descendants, K090675 (Smith & Nephew, Inc., 2009) has gone furthest, with 487 descendants of its own.
  • 35 other devices also cite its predicate K993106.

How does its category compare?

All HWC devices

HWC Screw, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · HWC

15in 2025

2007: 352008: 332009: 272010: 232011: 352012: 202013: 252014: 312015: 412016: 262017: 402018: 202019: 302020: 282021: 262022: 262023: 242024: 172025: 152026 (partial year): 21
2007high 41 (2015) · low 15 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HWC

111in 2025

2007: 77 days median2008: 76 days median2009: 113 days median2010: 110 days median2011: 88 days median2012: 99 days median2013: 90 days median2014: 88 days median2015: 85 days median2016: 131 days median2017: 127 days median2018: 106 days median2019: 108 days median2020: 75 days median2021: 84 days median2022: 167 days median2023: 104 days median2024: 90 days median2025: 111 days median2026 (partial year): 125 days median
2007high 167 (2022) · low 75 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HWC

1in 2025

2007: 252008: 332009: 292010: 72011: 122012: 302013: 142014: 22015: 52016: 42017: 42018: 12019: 52020: 42021: 22022: 132025: 12026 (partial year): 2
2007high 33 (2008) · low 0 (2023)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HWC

2,893in 2025

2007: 6692008: 6602009: 6512010: 7652011: 1,1042012: 3,3392013: 3,8502014: 2,5492015: 2,7112016: 2,2852017: 2,5612018: 3,6622019: 5,4632020: 4,8202021: 4,4462022: 3,9342023: 3,7182024: 3,5752025: 2,893
2007high 5,463 (2019) · low 651 (2009)2025

MAUDE reports naming product code HWC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007357725669
2008337633660
20092711329651
2010231107765
20113588121,104
20122099303,339
20132590143,850
2014318822,549
2015418552,711
20162613142,285
20174012742,561
20182010613,662
20193010855,463
2020287544,820
2021268424,446
202226167133,934
20232410403,718
2024179003,575
20251511112,893
2026211252—

What is it indicated for?

The Smith & Nephew Bone Plate System is used for adult and pediatric patients as indicated for pelvic, small and long bone fracture fixation. Indications for use include fractures of the tibia, fibula, femoral condyle, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, middle hand and middle foot bones; treatment of the calcaneal; hip arthrodesis, and provisional bone fixation.

Smith & Nephew 6.5mm and 8.0mm Cannulated Screws are for single use only

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K060736: Smith & Nephew 6.5MM and 8.0MM Cannulated Screws.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K060736

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026