510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

PDS Plus Antibacterial Suture

K061037510(k) TraditionalCleared Jul 14, 200691-day reviewClass II21 CFR 878.4840Recalls 9

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

ETHICON, Inc. · NEW Suture, Surgical, Absorbable, Polydioxanone

Cleared on Jul 14, 2006 as substantially equivalent to K032420 (ETHICON, Inc., 2003) and 1 other predicate after a 91-day review. Its family tree reaches back 2 generations to 1996. 21 later devices descend from it. It has been recalled 9 times.

Decided
Jul 14, 2006
Review
91 days
Class
Class II
Product code
NEW
Recalls
9
Predicates
3

Its longest line of descent runs back 2 generations to K960653 (1996). 4 ancestors in all. 21 later devices descend from it, 5 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor1996: 1 other ancestor2022: 2 descendants cleared2010: 1 descendant cleared2019: 4 descendants cleared2011: 1 descendant cleared2015: 1 descendant cleared2014: 2 descendants cleared2013: 1 descendant cleared2025: 4 descendants cleared2024: 2 descendants cleared2020: 1 descendant cleared2026: 1 descendant cleared2018: 1 descendant clearedK960653 · Monocryl (Poliglecaprone 25) Suture, Dyed (Modified) · ETHICON, Inc. · 1996 · recalledK050845 · Monocryl Plus Antibacterial Suture · ETHICON, Inc. · 2005 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 5 descendants2 generations forward: 6 descendants3 generations forward: 8 descendants4 generations forward: 2 descendants K050845 · Monocryl Plus Antibacterial Suture · ETHICON, Inc. · 2005 · recalledK960653 · Monocryl (Poliglecaprone 25) Suture, Dyed (Modified) · ETHICON, Inc. · 1996 · recalled0
  1. Root · 2 generations backK960653 Monocryl (Poliglecaprone 25) Suture, Dyed (Modified)ETHICON, Inc. · 1996 · Recalled
  2. 1 generation backK050845 Monocryl Plus Antibacterial SutureETHICON, Inc. · 2005 · Recalled
  3. This deviceK061037 PDS Plus Antibacterial SutureETHICON, Inc. · 2006
  4. Descendant · cites it directly · 8 of its ownK150670 STRATAFIX Spiral PDS Plus Knotless Tissue Control DeviceETHICON, Inc. · 2015
  5. Descendant · 2 generations on · 7 of its ownK182873 STRATAFIX Symmetric PDS Plus Knotless Tissue Control Devices…ETHICON, Inc. · 2019 · Recalled
  6. Descendant · cites it directly · 5 of its ownK100461 PDM, Vilet, Vilet Quick, Mono Q, Plain Gut:# 6 Chromic GutRiverpoint Medical · 2010 · Recalled
  7. Descendant · 3 generations on · 2 of its ownK221744 STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control DeviceETHICON, Inc. · 2022
  8. Descendant · cites it directly · 2 of its ownK212380 PDS Plus Antibacterial (Polydioxanone) Sterile Synthetic Absorable…ETHICON, Inc. · 2022 · Recalled

Showing 5 of 21 descendants. See all descendants

Descendants cleared per year

2010: 12011: 12013: 12014: 22015: 12018: 12019: 42020: 12022: 22024: 22025: 42026: 1
2010high 4 (2019) · low 0 (2012)2026

Most influential descendants

Showing 5 of 21 · See all
  • K150670STRATAFIX Spiral PDS Plus Knotless Tissue Control DeviceETHICON, Inc. · 2015 · cites it directly8descendants
  • K182873STRATAFIX Symmetric PDS Plus Knotless Tissue Control Devices, STRATAFIX Spiral PDS Plus…ETHICON, Inc. · 2019 · 2 generations on7descendants
  • K100461PDM, Vilet, Vilet Quick, Mono Q, Plain Gut:# 6 Chromic GutRiverpoint Medical · 2010 · cites it directly5descendants
  • K221744STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control DeviceETHICON, Inc. · 2022 · 3 generations on2descendants
  • K212380PDS Plus Antibacterial (Polydioxanone) Sterile Synthetic Absorable Surgical SutureETHICON, Inc. · 2022 · cites it directly2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K960653Monocryl (Poliglecaprone 25) Suture, Dyed (Modified)ETHICON, Inc.19962
−1K050845Monocryl Plus Antibacterial SutureETHICON, Inc.20051
0 (this device)K061037PDS Plus Antibacterial SutureETHICON, Inc.2006—
+1K150670STRATAFIX Spiral PDS Plus Knotless Tissue Control DeviceETHICON, Inc.2015—
+2K182873STRATAFIX Symmetric PDS Plus Knotless Tissue Control Devices, STRATAFIX Spiral…ETHICON, Inc.20195
+1K100461PDM, Vilet, Vilet Quick, Mono Q, Plain Gut:# 6 Chromic GutRiverpoint Medical20101
+3K221744STRATAFIX™ Spiral MONOCRYL™ Plus Knotless Tissue Control DeviceETHICON, Inc.2022—
+1K212380PDS Plus Antibacterial (Polydioxanone) Sterile Synthetic Absorable Surgical…ETHICON, Inc.20221

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201110
201310
201420
201510
201810
201940
202010
202220
202420
202540
202610
FIG. 1 — Lineage of K061037, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NEW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NEW device

  1. #1 K252047 STRATAFIX™ Symmetric PDS™ Plus Knotless Tissue…score 15.3 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
Kept off the family treeK018331PDS II (Polydioxanone) Suture
Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree
*K032420Modified Coated Vicryl Plus Antibacterial (Polyglactin 910) Synthetic Absorbable Suture
Summary page 1
…dioxanone) Antibacterial Suture Predicate Devices: PDS II*(Polydioxanone) Suture (N18331) VICRYL* Plus Antibacterial Suture (K032420) MONOCRYL* Plus Antibacterial Suture (K050845) Statement of Intended Use: PDS* Plus A…
ETHICON, Inc.20031
*K050845Monocryl Plus Antibacterial Suture
Summary page 1
…PDS II*(Polydioxanone) Suture (N18331) VICRYL* Plus Antibacterial Suture (K032420) MONOCRYL* Plus Antibacterial Suture (K050845) Statement of Intended Use: PDS* Plus Antibacterial monofilament synthetic absorbable sutures are indic…
ETHICON, Inc.20051

Also cleared against K032420: 2 other devices →

What descends from it? (21)

See all 21

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K150670STRATAFIX Spiral PDS Plus Knotless Tissue Control DeviceETHICON, Inc.20158
K100461PDM, Vilet, Vilet Quick, Mono Q, Plain Gut:# 6 Chromic GutRiverpoint Medical20105
K212380PDS Plus Antibacterial (Polydioxanone) Sterile Synthetic Absorable Surgical SutureETHICON, Inc.20222
K113004PDS (TM) Barbed SuturesETHICON, Inc.20111
K183183VICRYL Polyglactin 910 Sterile Synthetic Absorbable Surgical Suture, PDS II…ETHICON, Inc.20190

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 10, 2024Class IIZ-2147-2024Under Investigation by firmIssue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
Mar 31, 2023Class IIZ-1571-2023Under Investigation by firmSuture breakage during intra-operative use could result in poor performance of the impacted product because the intended benefit of tissue approximation and/or ligation may not be achieved
Mar 12, 2019Class IIZ-1135-2019Packaging process controlVICRYL Suture, VICRYL Plus Suture, PDS Suture- may contain product that does not match the sales unit carton and foil label.
Jun 8, 2018Class IIZ-2533-2018Under Investigation by firmProducts do not meet a tensile strength specification
Jun 8, 2018Class IIZ-2532-2018Under Investigation by firmProducts do not meet a tensile strength specification
Jan 10, 2018Class IIZ-1340-2018Process controlThree (3) lots were distributed with the incorrect expiry date on the label. The lots were labeled with an expiry date of five (5) years from date of manufacture instead of the correct expiry date of two (2) years from date of manufacture.
Jan 10, 2018Class IIZ-1338-2018Process controlThree (3) lots were distributed with the incorrect expiry date on the label. The lots were labeled with an expiry date of five (5) years from date of manufacture instead of the correct expiry date of two (2) years from date of manufacture.
Jan 10, 2018Class IIZ-1339-2018Process controlThree (3) lots were distributed with the incorrect expiry date on the label. The lots were labeled with an expiry date of five (5) years from date of manufacture instead of the correct expiry date of two (2) years from date of manufacture.
Feb 2, 2009Not classifiedZ-1314-2009Process controlSterility Issue: The firm decided to remove the product because it cannot assure the sterility of the products due to a failure in the sterilization process.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 1 company; a new generation was cleared every 9.3 years on median.
  • 4 of the 4 ancestors have a recall of any class.
  • Of its descendants, K150670 (ETHICON, Inc., 2015) has gone furthest, with 8 descendants of its own.
  • 2 other devices also cite its predicate K032420.

How does its category compare?

All NEW devices

NEW Suture, Surgical, Absorbable, Polydioxanone: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · NEW

6in 2025

2007: 22008: 52010: 32011: 12012: 32013: 52014: 12015: 22016: 32017: 12018: 32019: 42020: 32021: 12022: 12023: 12024: 32025: 62026 (partial year): 5
2007high 6 (2025) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NEW

146in 2025

2008: 22 days median2010: 270 days median2012: 113 days median2013: 32 days median2016: 254 days median2018: 441 days median2019: 144 days median2020: 230 days median2024: 48 days median2025: 146 days median2026 (partial year): 264 days median
2007high 441 (2018) · low 22 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NEW

1in 2022

2008: 12009: 42012: 12013: 12014: 12019: 22022: 1
2007high 4 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NEW

2,504in 2025

2007: 72008: 292009: 422010: 1152011: 592012: 822013: 1852014: 2742015: 2802016: 3212017: 4612018: 9002019: 1,1222020: 1,0102021: 1,4662022: 1,5372023: 1,9752024: 2,1142025: 2,504
2007high 2,504 (2025) · low 7 (2007)2025

MAUDE reports naming product code NEW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—07
2008522129
20090—442
201032700115
20111—059
20123113182
20135321185
20141—1274
20152—0280
201632540321
20171—0461
201834410900
2019414421,122
2020323001,010
20211—01,466
20221—11,537
20231—01,975
202434802,114
2025614602,504
202652640—

What is it indicated for?

PDS* Plus Antibacterial monofilament synthetic absorbable sutures are indicated for use in soft tissue approximation, including use in pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery (other than contact with cornea and sclera). PDS Plus is not indicated in adult cardiovascular tissue, microsurgery, and neural tissue. These sutures are particularly useful where the combination of an absorbable suture and extended wound support (up to six weeks) is desirable.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K061037: PDS Plus Antibacterial Suture.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K061037

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026