510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Randox Liquid Protein Calibrator

K061056510(k) TraditionalCleared Jul 20, 200694-day reviewClass II21 CFR 862.1150Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Randox Laboratories, Ltd. · JIX Calibrator, Multi-Analyte Mixture

Cleared on Jul 20, 2006 after a 94-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 2 later devices descend from it. It has been recalled 3 times.

Decided
Jul 20, 2006
Review
94 days
Class
Class II
Product code
JIX
Recalls
3
Predicates
0

2 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2011: 2 descendants cleared20052010201520202025
  1. This deviceK061056 Randox Liquid Protein CalibratorRandox Laboratories, Ltd. · 2006
  2. Descendant · cites it directly · 0 of its ownK110736 Advia Chemistry Ferritin (FRT) MethodSiemens Healthcare Diagnostics, Inc. · 2011 · Recalled
  3. Descendant · cites it directly · 0 of its ownK103701 Advia Chemistry Liquid Specific Protein CalibratorsSiemens Healthcare Diagnostics, Inc. · 2011

Most influential descendants

Showing 2 of 2 · See all
  • K110736Advia Chemistry Ferritin (FRT) MethodSiemens Healthcare Diagnostics, Inc. · 2011 · cites it directly0descendants
  • K103701Advia Chemistry Liquid Specific Protein CalibratorsSiemens Healthcare Diagnostics, Inc. · 2011 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K061056Randox Liquid Protein CalibratorRandox Laboratories, Ltd.2006—
+1K110736Advia Chemistry Ferritin (FRT) MethodSiemens Healthcare Diagnostics, Inc.20113
+1K103701Advia Chemistry Liquid Specific Protein CalibratorsSiemens Healthcare Diagnostics, Inc.2011—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201120
FIG. 1 — Lineage of K061056, 2006–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JIX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K110736Advia Chemistry Ferritin (FRT) MethodSiemens Healthcare Diagnostics, Inc.20110
K103701Advia Chemistry Liquid Specific Protein CalibratorsSiemens Healthcare Diagnostics, Inc.20110

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 1, 2024Class IIZ-1328-2024Under Investigation by firmThe firm has realigned C3 and Haptoglobin in Liquid Protein Calibrators, IT2691, to reference material ERM-DA470k/IFCC.
Jul 19, 2023Class IIZ-2448-2023Under Investigation by firmAs part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The calibrators lots…
Aug 9, 2019Class IIZ-0283-2020Nonconforming Material/ComponentThe assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British standard from NIBSC. Ref = 64/002. The calibrator values have been adjusted to re-align to the reference material.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K110736 (Siemens Healthcare Diagnostics, Inc., 2011) has gone furthest, with 0 descendants of its own.

How does its category compare?

All JIX devices

JIX Calibrator, Multi-Analyte Mixture: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JIX

1in 2016

2007: 62008: 52009: 42010: 52011: 72012: 22013: 42014: 12015: 32016: 1
2007high 7 (2011) · low 0 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JIX

29in 2015

2007: 53 days median2008: 69 days median2009: 121 days median2010: 100 days median2011: 77 days median2013: 50 days median2015: 29 days median
2007high 121 (2009) · low 29 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JIX

7in 2023

2007: 22008: 32009: 22010: 12011: 22012: 12014: 12016: 12017: 12018: 12019: 12020: 22023: 72026 (partial year): 1
2007high 7 (2023) · low 0 (2013)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JIX

18in 2025

2007: 1402008: 1732009: 392011: 92012: 32013: 242014: 352015: 42016: 62017: 172018: 162019: 122020: 222021: 92022: 112023: 132024: 152025: 18
2007high 173 (2008) · low 0 (2010)2025

MAUDE reports naming product code JIX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20076532140
20085693173
20094121239
2010510010
201177729
20122—13
2013450024
20141—135
201532904
20161—16
20170—117
20180—116
20190—112
20200—222
20210—09
20220—011
20230—713
20240—015
20250—018
20260—1—

What is it indicated for?

The Randox Laboratories Limited Liquid Protein Calibrator (for neat sample assays) is derived from normal human serum obtained from volunteer donors. It has been developed for the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, IgA, IgG, IgM, Prealbumin, Transferrin and Haptoglobin assays (all neat sample assays). Assignment was performed at Randox Laboratories by immunoturbidimetry with reference to International Reference Material CRM…

Read the full statement (181 words)

The Randox Laboratories Limited Liquid Protein Calibrator (for neat sample assays) is derived from normal human serum obtained from volunteer donors. It has been developed for the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, IgA, IgG, IgM, Prealbumin, Transferrin and Haptoglobin assays (all neat sample assays).

Assignment was performed at Randox Laboratories by immunoturbidimetry with reference to International Reference Material CRM 470. The constituent concentrations of these Calibrators are present at levels 1, 2, 3, 4 and 5.

This calibrator can be used on Abbott Spectrum, Abbott aeroset, Abbott Architect i2000, Architect i2000sr, Ace Analyser, Bayer Advia 1650, Advia 2400, Advia 1200, Dade Dimension RXL, Dimension AR, Hitachi 704, Hitachi 717, Hitachi 911, Hitachi 917, Hitachi 912, Hitachi 747, Kone progress, Olympus AU800, AU600, AU400, AU2700, AU5400, Selectra vitalab, Synchron CX4 Synchron CX5 Synchron CX7, Synchron LX20, ILAB300, ILAB900, ILAB1800, ILAB600, RX Daytona, RX Imola, Cobas Mira, Cobas Mira S, Cobas Mira plus systems

The Randox Laboratories Limited Liquid Protein Calibrators, should only be used by suitably qualified laboratory personnel under appropriate laboratory conditions.

For in vitro diagnostic use.

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K061056: Randox Liquid Protein Calibrator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K061056

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026