510(k) Report

510(k) · Cleared

Cleared

Sherlock Tip Location System (TLS) Detector and Accessories

K061240510(k) SpecialCleared Jun 2, 200630-day reviewClass II21 CFR 880.5970

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Lucent Medical Systems · LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days

Cleared on Jun 2, 2006 after a 30-day review. The summary cites one number, which could not be confirmed as its predicate (see the table below). 6 of the 44 devices in its family, counting itself, have a Class I recall. 43 later devices descend from it.

Decided
Jun 2, 2006
Review
30 days
Class
Class II
Product code
LJS
Recalls
None
Predicates
1

43 later devices descend from it, 2 citing it directly. 6 of the 44 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2009: 1 descendant cleared2006: 1 descendant cleared2008: 2 descendants cleared2010: 2 descendants cleared2012: 3 descendants cleared, 1 with a Class I recall12014: 4 descendants cleared2017: 3 descendants cleared2011: 3 descendants cleared2015: 5 descendants cleared2022: 2 descendants cleared2013: 4 descendants cleared, 2 with a Class I recall22018: 2 descendants cleared2025: 4 descendants cleared2021: 1 descendant cleared2016: 4 descendants cleared, 3 with a Class I recall32024: 1 descendant cleared2026: 1 descendant cleared20052010201520202025
  1. This deviceK061240 Sherlock Tip Location System (TLS) Detector and AccessoriesLucent Medical Systems · 2006
  2. Descendant · cites it directly · 42 of its ownK063240 Sherlock TLS Stylet, Sherlock TLS DetectorC.R. Bard, Inc. · 2006 · Recalled
  3. Descendant · 2 generations on · 40 of its ownK081625 Flowpicc StyletVasonova, Inc. · 2008
  4. Descendant · 2 generations on · 39 of its ownK081626 Flowpicc ConsoleVasonova, Inc. · 2008
  5. Descendant · cites it directly · 38 of its ownK091324 Sherlock 3CG Tip Positioning System, Sherlock 3CG Tip Positioning…C.R. Bard, Inc. · 2009 · Recalled
  6. Descendant · 2 generations on · 34 of its ownK093775 Evguide Tip Location SystemRomedex International Srl · 2010

Showing 5 of 43 descendants. See all descendants

Descendants cleared per year

2006: 12008: 22009: 12010: 22011: 32012: 32013: 42014: 42015: 52016: 42017: 32018: 22021: 12022: 22024: 12025: 42026: 1
2006high 5 (2015) · low 0 (2007)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 43 · See all
  • K063240Sherlock TLS Stylet, Sherlock TLS DetectorC.R. Bard, Inc. · 2006 · cites it directly42descendants
  • K081625Flowpicc StyletVasonova, Inc. · 2008 · 2 generations on40descendants
  • K081626Flowpicc ConsoleVasonova, Inc. · 2008 · 2 generations on39descendants
  • K091324Sherlock 3CG Tip Positioning System, Sherlock 3CG Tip Positioning System StyletC.R. Bard, Inc. · 2009 · cites it directly38descendants
  • K093775Evguide Tip Location SystemRomedex International Srl · 2010 · 2 generations on34descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K061240Sherlock Tip Location System (TLS) Detector and AccessoriesLucent Medical Systems2006—
+1K063240Sherlock TLS Stylet, Sherlock TLS DetectorC.R. Bard, Inc.200618
+2K081625Flowpicc StyletVasonova, Inc.2008—
+2K081626Flowpicc ConsoleVasonova, Inc.2008—
+1K091324Sherlock 3CG Tip Positioning System, Sherlock 3CG Tip Positioning System StyletC.R. Bard, Inc.200911
+2K093775Evguide Tip Location SystemRomedex International Srl2010—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200820
200910
201020
201130
201231
201342
201440
201550
201643
201730
201820
202110
202220
202410
202540
202610
FIG. 1 — Lineage of K061240, 2006–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LJS device

  1. #1 K241910 Neonav ECG Tip Location Systemscore 18.0 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
Kept off the family treeK000099FDA’s K000099 is Camlog Screw Implant System
Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree
Altatec Biotechnologies2000—

What descends from it? (43)

See all 43

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K063240Sherlock TLS Stylet, Sherlock TLS DetectorC.R. Bard, Inc.200642
K091324Sherlock 3CG Tip Positioning System, Sherlock 3CG Tip Positioning System StyletC.R. Bard, Inc.200938

Has anything in its family been recalled? (6 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K122545CG+ Arrow Picc Powered by Arrow VPS StyletArrow International (Subsidiary of Teleflex Inc.) · 2012 · 4 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K132133CG+ Arrow Jacc Powered by Arrow VPS StyletArrow International (Subsidiary of Teleflex Inc.) · 2013 · 4 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K130876CG+ Arrow Picc Powered by Arrow VPS StyletArrow International, Inc. · 2013 · 5 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K153423CG+ Arrow JACC powered by Arrow VPS Stylet, CG+ Arrow JACCArrow International, Inc. (Subsidiary of Teleflex, Inc.) · 2016 · 5 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall
  • K153487CG+ Arrow PICC powered by Arrow VPS Stylet PLUS and Arrow VPS Stylet PLUSArrow International, Inc. (Subsidiary of Teleflex, Inc.) · 2016 · 5 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall
  • K160925VPS Rhythm Device with TipTracker TechnologyArrow International, Inc. (A Subsidiary of Teleflex, Inc.) · 2016 · 5 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K063240 (C.R. Bard, Inc., 2006) has gone furthest, with 42 descendants of its own.

How does its category compare?

All LJS devices

LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LJS

8in 2025

2007: 142008: 42009: 162010: 92011: 62012: 142013: 142014: 82015: 72016: 92017: 72018: 102020: 22021: 32022: 32023: 42024: 12025: 82026 (partial year): 2
2007high 16 (2009) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LJS

171in 2025

2007: 92 days median2008: 97 days median2009: 70 days median2010: 114 days median2011: 55 days median2012: 77 days median2013: 31 days median2014: 47 days median2015: 71 days median2016: 128 days median2017: 101 days median2018: 148 days median2021: 284 days median2022: 177 days median2023: 47 days median2025: 171 days median
2007high 284 (2021) · low 31 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LJS

5in 2023

2008: 12009: 12010: 122011: 42012: 402014: 152015: 22016: 182017: 92018: 32021: 12023: 5, 2 Class I2026 (partial year): 8, 1 Class I
2007high 40 (2012) · low 0 (2007)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · LJS

4,372in 2025

2007: 3922008: 3692009: 3792010: 4952011: 6962012: 5642013: 6802014: 8422015: 1,3062016: 1,1232017: 2,1002018: 2,3242019: 2,0522020: 2,5392021: 3,0842022: 3,2562023: 2,7992024: 4,2432025: 4,372
2007high 4,372 (2025) · low 369 (2008)2025

MAUDE reports naming product code LJS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714920392
20084971369
200916701379
2010911412495
20116554696
2012147740564
201314310680
201484715842
201577121,306
20169128181,123
2017710192,100
20181014832,324
20190—02,052
20202—02,539
2021328413,084
2022317703,256
20234475 (2)2,799
20241—04,243
2025817104,372
20262—8 (1)—

What is it indicated for?

The SherlockTM Tip Location System (TLS) Detector quickly locates and confirms the position of specially designed, magnet-tipped Peripherally-Inserted Central Catheters (PICCs) and Central Venous Catheters (CVCs) during initial placement. This device may be used by appropriate caregivers in hospitals, long-term care facilities or home-care settings.

The SherlockTM Tip Location System (TLS) Detector provides rapid feedback to the caregiver, but was not designed to replace conventional methods of placement verification. Users are urged to confirm correct placement according to their established institutional protocol and clinical judgment.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K061240: Sherlock Tip Location System (TLS) Detector and Accessories.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K061240

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026