510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

ILED

K061317510(k) TraditionalCleared Jun 22, 200642-day reviewClass II21 CFR 878.4580Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Trumpf Kreuzer Medizin Systeme GmbH + Co.Kg · FSY Light, Surgical, Ceiling Mounted

Cleared on Jun 22, 2006 as substantially equivalent to K024132 (Berchtold Holding GmbH, 2003) after a 42-day review. Its family tree reaches back 2 generations to 1997. 16 later devices descend from it. It has been recalled 7 times.

Decided
Jun 22, 2006
Review
42 days
Class
Class II
Product code
FSY
Recalls
7
Predicates
1

Its longest line of descent runs back 2 generations to K965130 (1997). 16 later devices descend from it, 7 citing it directly.

Descendants, by year cleared 2010: 5 descendants cleared2013: 3 descendants cleared2009: 2 descendants cleared2007: 1 descendant cleared2012: 1 descendant cleared2014: 4 descendants clearedK965130 · Chromophare · Berchtold Holding GmbH · 1997K024132 · Chromophare X 65 · Berchtold Holding GmbH · 20031995200020052010201520202025
Descendants, by generation 1 generation forward: 7 descendants2 generations forward: 6 descendants3 generations forward: 2 descendants4 generations forward: 1 descendant K024132 · Chromophare X 65 · Berchtold Holding GmbH · 2003K965130 · Chromophare · Berchtold Holding GmbH · 19970
  1. Root · 2 generations backK965130 ChromophareBerchtold Holding GmbH · 1997
  2. 1 generation backK024132 Chromophare X 65Berchtold Holding GmbH · 2003
  3. This deviceK061317 ILEDTrumpf Kreuzer Medizin Systeme GmbH + Co.Kg · 2006
  4. Descendant · cites it directly · 10 of its ownK072072 Harmony Led-1 Surgical Lighting SystemSTERIS Corporation · 2007 · Recalled
  5. Descendant · 2 generations on · 4 of its ownK093009 Mach Led SCDr. Mach GmbH & Co. KG · 2010
  6. Descendant · cites it directly · 2 of its ownK093010 Mach Led MCDr. Mach GmbH & Co. KG · 2010
  7. Descendant · 3 generations on · 1 of its ownK132307 Vida Led Surgical Lighting SystemConvida Healthcare & Systems Corporation · 2013
  8. Descendant · cites it directly · 1 of its ownK100388 21ST Century Centurion Exceled and System Two Led Lighting SystemsMedical Illumination International, Inc. · 2010

Showing 5 of 16 descendants. See all descendants

Descendants cleared per year

2007: 12009: 22010: 52012: 12013: 32014: 4
2007high 5 (2010) · low 0 (2008)2026

Most influential descendants

Showing 5 of 16 · See all
  • K072072Harmony Led-1 Surgical Lighting SystemSTERIS Corporation · 2007 · cites it directly10descendants
  • K093009Mach Led SCDr. Mach GmbH & Co. KG · 2010 · 2 generations on4descendants
  • K093010Mach Led MCDr. Mach GmbH & Co. KG · 2010 · cites it directly2descendants
  • K132307Vida Led Surgical Lighting SystemConvida Healthcare & Systems Corporation · 2013 · 3 generations on1descendant
  • K10038821ST Century Centurion Exceled and System Two Led Lighting SystemsMedical Illumination International, Inc. · 2010 · cites it directly1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K965130ChromophareBerchtold Holding GmbH1997—
−1K024132Chromophare X 65Berchtold Holding GmbH2003—
0 (this device)K061317ILEDTrumpf Kreuzer Medizin Systeme GmbH + Co.Kg2006—
+1K072072Harmony Led-1 Surgical Lighting SystemSTERIS Corporation20071
+2K093009Mach Led SCDr. Mach GmbH & Co. KG2010—
+1K093010Mach Led MCDr. Mach GmbH & Co. KG2010—
+3K132307Vida Led Surgical Lighting SystemConvida Healthcare & Systems Corporation2013—
+1K10038821ST Century Centurion Exceled and System Two Led Lighting SystemsMedical Illumination International, Inc.2010—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
200920
201050
201210
201330
201440
FIG. 1 — Lineage of K061317, 1997–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FSY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K024132Chromophare X 65
Summary page 2
…site on the patient's body with a high intensity, shadow fee, "cold" light, Predicate Device: CHROMOPHARE X 65 K#024132 Substantial Equivalence: The LED is substantially equivalent to the surgical light CHROMOPHARE X65.…
Berchtold Holding GmbH2003—

Also cleared against K024132: 2 other devices →

What descends from it? (16)

See all 16

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K072072Harmony Led-1 Surgical Lighting SystemSTERIS Corporation200710
K093010Mach Led MCDr. Mach GmbH & Co. KG20102
K10038821ST Century Centurion Exceled and System Two Led Lighting SystemsMedical Illumination International, Inc.20101
K123776Polaris 100/200Drager Medical GmbH20130
K102758Trulight 5520 and Trulight 5320Trumpf Kreuzer Medizin Systeme GmbH + Co.Kg20100
K091246Iled 3 Mobile (I300M), Trulight 5500 Mobile (T500M), Trulight 5300 Mobile (T300M)Trumpf Medizin Systeme GmbH + Co. KG20090
K083323Nuvo Verde Led Lighting SystemNuvo, Inc.20090

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 2, 2015Class IIZ-0563-2016OtherThe firm has received 15 complaints over ten years related to the connection of the spring arm to the top horizontal arm of the lighting system that while positioning the light head enabled the spring arm and light head to fall.
Jul 27, 2015Class IIZ-2353-2015Process controlComplaints received of the front joint of the spring arm AC 2000 could break at the interface to the monitor.
Jul 27, 2015Class IIZ-2354-2015Process controlComplaints received of the front joint of the spring arm AC 2000 could break at the interface to the monitor.
Jul 27, 2015Class IIZ-2356-2015Process controlComplaints received of the front joint of the spring arm AC 2000 could break at the interface to the monitor.
Jul 27, 2015Class IIZ-2357-2015Process controlComplaints received of the front joint of the spring arm AC 2000 could break at the interface to the monitor.
May 29, 2012Class IIZ-2311-2012Process controlCracks can arise on the welded seam on the rear joint.
Feb 27, 2009Not classifiedZ-1122-2009OtherCracks may form around the screw connections of the cover plates causing brittleness, which may lead to fragments breaking off and falling into the operating area.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 2 companies; a new generation was cleared every 5.7 years on median.
  • None of the 2 ancestors has a recall.
  • Of its descendants, K072072 (STERIS Corporation, 2007) has gone furthest, with 10 descendants of its own.
  • 2 other devices also cite its predicate K024132.

How does its category compare?

All FSY devices

FSY Light, Surgical, Ceiling Mounted: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FSY

7in 2014

2007: 32008: 22009: 22010: 52012: 22013: 32014: 7
2007high 7 (2014) · low 0 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FSY

59in 2014

2007: 29 days median2010: 130 days median2013: 118 days median2014: 59 days median
2007high 130 (2010) · low 29 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FSY

11in 2023

2009: 42010: 32011: 42012: 42013: 12015: 2282018: 22019: 42023: 112026 (partial year): 4
2007high 228 (2015) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FSY

286in 2025

2007: 702008: 532009: 402010: 722011: 702012: 1712013: 1372014: 1392015: 1752016: 1822017: 1622018: 3162019: 4442020: 4392021: 3802022: 4302023: 6632024: 5412025: 286
2007high 663 (2023) · low 40 (2009)2025

MAUDE reports naming product code FSY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007329070
20082—053
20092—440
20105130372
20110—470
20122—4171
201331181137
20147590139
20150—228175
20160—0182
20170—0162
20180—2316
20190—4444
20200—0439
20210—0380
20220—0430
20230—11663
20240—0541
20250—0286
20260—4—

What is it indicated for?

The surgical light iLED is intended to illuminate locally the operating site on the patient's body with a high intensity, shadow fee, "cold" light,

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K061317: ILED.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K061317

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026