510(k) Report

510(k) · Cleared

Cleared

Modification to DBB-05 Hemodialysis Delivery System

K061519510(k) SpecialCleared Sep 28, 2007484-day reviewClass II21 CFR 876.5860

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Nikkiso CO., LTD. · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Cleared on Sep 28, 2007 as substantially equivalent to K023509 (Davita Clinical Research (Trri), 2003) after a 484-day review. Its family tree reaches back 3 generations to 1994. 3 later devices descend from it.

Decided
Sep 28, 2007
Review
484 days
Class
Class II
Product code
KDI
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K921456 (1994). 3 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2010: 1 descendant cleared2016: 1 descendant cleared2025: 1 descendant clearedK921456 · Fresenius 3008/FDS-08 Dialysate Delivery System · Fresenius USA, Inc. · 1994 · recalledK994267 · Fresenius 2008K · Fresenius Medical Care · 2000 · recalledK023509 · DBB-05 Hemodialysis Delivery System · Davita Clinical Research (Trri) · 20031990200020102020
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant K023509 · DBB-05 Hemodialysis Delivery System · Davita Clinical Research (Trri) · 2003K994267 · Fresenius 2008K · Fresenius Medical Care · 2000 · recalledK921456 · Fresenius 3008/FDS-08 Dialysate Delivery System · Fresenius USA, Inc. · 1994 · recalled0
  1. Root · 3 generations backK921456 Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc. · 1994 · Recalled
  2. 2 generations backK994267 Fresenius 2008KFresenius Medical Care · 2000 · Recalled
  3. 1 generation backK023509 DBB-05 Hemodialysis Delivery SystemDavita Clinical Research (Trri) · 2003
  4. This deviceK061519 Modification to DBB-05 Hemodialysis Delivery SystemNikkiso CO., LTD. · 2007
  5. Descendant · cites it directly · 2 of its ownK091978 DBB-06 Hemodialysis Delivery SystemNikkiso CO., LTD. · 2010 · Recalled
  6. Descendant · 2 generations on · 1 of its ownK152938 DBB-06 Hemodialysis Delivery SystemNikkiso CO., LTD. · 2016 · Recalled
  7. Descendant · 3 generations on · 0 of its ownK242155 DBB-06 PRO Hemodialysis Delivery System (DBB-06 PRO)Nikkiso CO., LTD. · 2025

Most influential descendants

Showing 3 of 3 · See all
  • K091978DBB-06 Hemodialysis Delivery SystemNikkiso CO., LTD. · 2010 · cites it directly2descendants
  • K152938DBB-06 Hemodialysis Delivery SystemNikkiso CO., LTD. · 2016 · 2 generations on1descendant
  • K242155DBB-06 PRO Hemodialysis Delivery System (DBB-06 PRO)Nikkiso CO., LTD. · 2025 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K921456Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc.19943
−2K994267Fresenius 2008KFresenius Medical Care20005
−1K023509DBB-05 Hemodialysis Delivery SystemDavita Clinical Research (Trri)2003—
0 (this device)K061519Modification to DBB-05 Hemodialysis Delivery SystemNikkiso CO., LTD.2007—
+1K091978DBB-06 Hemodialysis Delivery SystemNikkiso CO., LTD.20101
+2K152938DBB-06 Hemodialysis Delivery SystemNikkiso CO., LTD.20161
+3K242155DBB-06 PRO Hemodialysis Delivery System (DBB-06 PRO)Nikkiso CO., LTD.2025—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201610
202510
FIG. 1 — Lineage of K061519, 1994–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KDI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KDI device

  1. #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K023509DBB-05 Hemodialysis Delivery System
Summary page 2
…quipped with the necessary protective systems for patient safety and correct operation. C. Predicate Device Information DBB-05 Hemodialysis Delivery System [K023509] (10/30/03) D. Indications for Use/Intended Use The DBB-05 Hemodialysis Delivery System is indicated for hemodialysi…
Davita Clinical Research (Trri)2003—

Also cleared against K023509: 1 other device →

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K091978DBB-06 Hemodialysis Delivery SystemNikkiso CO., LTD.20102

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 3.9 years on median.
  • 2 of the 3 ancestors have a recall of any class.
  • Of its descendants, K091978 (Nikkiso CO., LTD., 2010) has gone furthest, with 2 descendants of its own.
  • 1 other device also cites its predicate K023509.

How does its category compare?

All KDI devices

KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · KDI

9in 2025

2007: 112008: 62009: 42010: 92011: 52012: 62013: 112014: 82015: 52016: 52017: 82018: 22019: 42020: 92021: 22022: 72023: 72024: 72025: 92026 (partial year): 6
2007high 11 (2007) · low 2 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KDI

199in 2025

2007: 91 days median2008: 148 days median2009: 236 days median2010: 186 days median2011: 85 days median2012: 153 days median2013: 247 days median2014: 164 days median2015: 69 days median2016: 230 days median2017: 188 days median2019: 263 days median2020: 266 days median2022: 259 days median2023: 157 days median2024: 235 days median2025: 199 days median2026 (partial year): 154 days median
2007high 266 (2020) · low 69 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KDI

2in 2025

2007: 42009: 32010: 12011: 42012: 12013: 22014: 32015: 52016: 42017: 12018: 12019: 62021: 22022: 172023: 8, 8 Class I2024: 4, 2 Class I2025: 22026 (partial year): 2
2007high 17 (2022) · low 0 (2008)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · KDI

2,786in 2025

2007: 9162008: 1,0422009: 5222010: 8232011: 2,2702012: 1,7652013: 2,3602014: 6,1542015: 4,1822016: 7662017: 9632018: 1,7162019: 1,5872020: 1,6332021: 1,9442022: 2,1802023: 2,6922024: 2,4412025: 2,786
2007high 6,154 (2014) · low 522 (2009)2025

MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711914916
2008614801,042
200942363522
201091861823
201158542,270
2012615311,765
20131124722,360
2014816436,154
201556954,182
201652304766
201781881963
20182—11,716
2019426361,587
2020926601,633
20212—21,944
20227259172,180
202371578 (8)2,692
202472354 (2)2,441
2025919922,786
202661542—

What is it indicated for?

The DBB-05 Hemodialysis Delivery System is indicated for hemodialysis

prescribed by physicians for adult and pediatric patients with acute or

chronic renal failure. The DBB-05 is intended for hemodialysis performed

in hospitals and dialysis clinics by a qualified operator.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K061519: Modification to DBB-05 Hemodialysis Delivery System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K061519

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026