# K061655: Dimension Vista Flex Reagent Cartridges

- Pathway: 510(k) (Special)
- Applicant: Dade Behring, Inc.
- Decided: Jul 10, 2006 after 27 days
- Product code: LDP (Colorimetry, Acetaminophen)
- Device class: 2
- Regulation: 21 CFR 862.3030
- Recalls of this device: 7 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 4 generations back to K891093 (1989). 23 ancestors in all.
1 later device descends from it; 1 cites it directly.
0 of 25 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K963244: Acetaminophen Flex Reagent Cartridge, Dade Intl., Inc. (1996) Summary page 2.
- K942640: Dupont Dimension Amylase Method, Dupont Medical Products (1994) Summary page 2.
- K941584: Creatine Kinase Method, Dupont Medical Products (1994) Summary page 2.
- K861700: Dimension (TM) Clinical Chemistry System, E.I. Dupont DE Nemours & Co., Inc. (1986) Summary page 2.
- K031135: Glucose Flex Reagent Cartridge (Gluc), Dade Behring, Inc. (2003) Summary page 2.
- K053157: Automated HDL Cholesterol Flex Reagent Cartridge, Model DF48A, Dade Behring, Inc. (2005) Summary page 2.
- K020724: Dimension Automated LDL Cholesterol Flex Reagentcartridge Method, Model DF131, Dade Behring, Inc. (2002) Summary page 2.
- K033809: Dimension Lidocaine (Lido) Flex Reagent Cartridge Method, Dimension Drug Calibrator II, Models DF113. DC49D, Dade Behring, Inc. (2004) Summary page 3.
- K950180: Dupont Dimension (R) Pseudocholinesterase Method, E.I. Dupont DE Nemours & Co., Inc. (1995) Summary page 3.
- K032573: Dimension Procainamide (Proc) Flex Reagent Cartridge Method, Model DF110, Dade Behring, Inc. (2003) Summary page 3.
- K904301: DU Pont Dimension (R) Salicylate (Sal) Method, E.I. Dupont DE Nemours & Co., Inc. (1990) Summary page 3.
- K862359: Meth./Cali./Intru. Check for the Dimension System, E.I. Dupont DE Nemours & Co., Inc. (1986) Summary page 3.
- K962818: Tobramycin Flex Reagent Cartridge, Dade Intl., Inc. (1996) Summary page 3.
- K010871: TGL Flex Reagent Cartridge, Dade Behring, Inc. (2001) Summary page 3.
- K982880: Dimension (R) Valproic (Valp) Acid Method, Dade Behring, Inc. (1998) Summary page 3.
- K963267: Vancomycin Flex Reagent Cartridge, Dade Intl., Inc. (1996) Summary page 3.

## Most influential descendants
- K140790: Dimension Vista Chemistry 1 Calibrator (CHEM 1 CAL), Dimension Vista Magnesium Flex reagent cartridge (MG) (2015), 0 descendants of its own

## Recalls of this device
- Jun 29, 2017 · Class II · Z-1658-2018: Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely dep
- Jun 29, 2017 · Class II · Z-1660-2018: Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely dep
- Mar 17, 2016 · Class II · Z-1665-2016: Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depre
- Mar 17, 2016 · Class II · Z-1666-2016: Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depre
- Sep 30, 2015 · Class II · Z-0339-2016: Erroneous low results on a small number of reagent wells. There is the potential for under-recovery on certain wells for quality control (QC) and patient sample results. The under-recovery ranged from
- Jul 17, 2012 · Class III · Z-2175-2012: Siemens has received complaints of customers experiencing slopes >1.05 when calibrating Flex(R) reagent cartridge lot 12095AD. This results in a slope failure and prevents the VANC assay from being ru
- Feb 17, 2012 · Not classified · Z-1433-2012: Firm has confirmed a negative shift of approximately 0.2mg/dL [0.08 mmol/L] in patient and QC samples when using Vista Magnesium

## Indications for use
The Dimension VistaTM Acetaminophen (ACTM) Flex® reagent cartridge is a device intended to measure acetaminophen, an analgesic and antipyretic (fever reducing) drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of acetaminophen overdose. The Dimension VistaTM Amylase (AMY) Flex® reagent cartridge is a device intended to measure the activity of the enzyme amylase in serum, plasma and urine. Amylase measurements are used primarily for the diagnosis and treatment of pancreatitis (inflammation of the pancreas). The Dimension VistaTM Creatine Kinase (CK) Flex® reagent cartridge is a device intended to measure the activity of the enzyme creatine kinase in serum and plasma. Measurements of creatine kinase are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive Duchenne-type muscular dystrophy. The Dimension VistaTM Cholesterol (CHOL) Flex® reagent cartridge is a device intended to measure cholesterol in serum and plasma. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. The Dimension VistaTM Gamma-glutamyl transferase (GGT) Flex® reagent cartridge is a device intended to measure gamma-glutamyl transferase in human serum and plasma. Gamma-glutamyl transferase measurements are used in the diagnosis and treatment of liver diseases such as alcoholic cirrhosis and primary and secondary liver tumors. The Dimension VistaTM Glucose (GLU) Flex® reagent cartridge is a device intended to measure glucose in human serum, plasma, urine and cerebrospinal fluid. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal and idiopathic hypoglycemia, and pancreatic islet cell carcinoma. The Dimension VistaTM High-Density Lipoprotein Cholesterol (HDLC) Flex® reagent cartridge is intended to measure high-density lipoprotein cholesterol in serum and plasma. Measurements of high-density lipoprotein cholesterol are used in the diagnosis of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. The Dimension VistaTM Low-Density Lipoprotein Cholesterol (LDLC) Flex® reagent cartridge is intended to measure low-density lipoprotein cholesterol in serum and plasma. Measurements of low-density lipoprotein cholesterol are used in the diagnosis of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. The Dimension VistaTM Lidocaine (LIDO) Flex® reagent cartridge is a device intended to measure lidocaine, an antiarrythmic and anticonvulsant drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of lidocaine overdose or in monitoring levels of lidocaine to ensure appropriate therapy. The Dimension VistaTM Magnesium (MG) Flex® reagent cartridge is intended for the measurement of magnesium levels in serum and plasma. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypermagnesemia (abnormally high plasma levels of magnesium). The Dimension VistaTM Pseudocholinesterase (PCHE) Flex® reagent cartridge is a device intended to measure pseudocholinesterase activity in human serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of cholinesterase inhibition disorders (e.g., insecticide poisoning and succinylcholine poisoning). The Dimension VistaTM Phosphorus (PHOS) Flex® reagent cartridge is a device intended to measure inorganic phosphorus in serum, plasma, and urine. Measurements of phosphorus (inorganic) are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases, and vitamin D imbalance. The Dimension VistaTM Procainamide (PROC) Flex® reagent cartridge is a device intended to measure procainamide in serum and plasma. Measurements obtained may be used in the diagnosis and treatment of procainamide overdose and in monitoring levels of procainamide to ensure appropriate therapy. The Dimension VistaTM Salicylate (SAL) Flex® reagent cartridge is a device intended to measure salicylates, a class of analgesic, antipyretic and anti-inflammatory drugs that includes aspirin, in human serum. Measurements obtained by this device are used in the diagnosis and treatment of salicylate overdose and in monitoring salicylate levels to ensure appropriate therapy. The Dimension VistaTM Thyroxine (T4) Flex®reagent cartridge is a device intended to measure total (free and protein bound) thyroxine (thyroid hormone) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases. The Dimension VistaTM Tobramycin (TOBR) Flex® reagent cartridge is a device intended to measure tobramycin, an aminoglycoside antibiotic drug, in palsma and serum. Measurements obtained by this device are used in the diagnosis and treatment of tobramycin overdose and in monitoring levels of tobramycin to ensure appropriate therapy. The Dimension VistaTM Triglyceride (TRIG) Flex® reagent cartridge is a device intended to measure triglyceride (neutral fat) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders. The Dimension VistaTM Uric Acid (URCA) Flex® reagent cartridge is a device intended to measure uric acid in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs. The Dimension VistaTM Valproic Acid (VALP) Flex® reagent cartridge is a device intended to measure valproic acid, an anti-convulsant drug in serum and plasma. Measurements obtained may be used in the diagnosis and treatment of valproic acid overdose and in monitoring levels of valproic acid to ensure appropriate therapy. The Dimension VistaTM Vancomycin (VANC) Flex® reagent cartridge is a device intended to measure vancomycin, an antibiotic drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of vancomycin overdose and in monitoring the level of vancomycin to ensure appropriate therapy.

Source page: https://510kreport.com/device/K061655