# K062077: Vitros Chemistry Products Amph Reagent; Calibrator Kit 26; FS Calibrator 1; Dat Performance Verifiers

- Pathway: 510(k) (Traditional)
- Applicant: Ortho-Clinical Diagnostics, Inc.
- Decided: Nov 22, 2006 after 124 days
- Product code: DKZ (Enzyme Immunoassay, Amphetamine)
- Device class: 2
- Regulation: 21 CFR 862.3100
- Recalls of this device: 5 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 2 generations back to K993982 (2000). 3 ancestors in all.
1 later device descends from it; 1 cites it directly.
0 of 5 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K031004: Syva Emit II Plus Amphetamines Assay, Dade Behring, Inc. (2003) Summary page 3.
- K022707: Liquichek Urine Toxicology Control Level S1E, Model 438; Level S2E, Model 439, Bio-Rad (2002) Summary page 3.

## Most influential descendants
- K143500: Immunalysis Amphetamine Urine Enzyme Immunoassay, Immunalysis Amphetamine Urine Calibrator, Immunalysis Amphetamine Urine Control Set (2015), 0 descendants of its own

## Recalls of this device
- Sep 22, 2017 · Class II · Z-0911-2018: This product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers causing customers to obtain lower than expected VITROS BENZ results when using this control.
- Sep 22, 2017 · Class II · Z-0913-2018: This product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers causing customers to obtain lower than expected VITROS BENZ results when using this control.
- Sep 22, 2017 · Class II · Z-0912-2018: This product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers causing customers to obtain lower than expected VITROS BENZ results when using this control.
- Sep 22, 2017 · Class II · Z-0914-2018: This product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers causing customers to obtain lower than expected VITROS BENZ results when using this control.
- Sep 22, 2017 · Class II · Z-0915-2018: This product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers causing customers to obtain lower than expected VITROS BENZ results when using this control.

## Indications for use
For in vitro diagnostic use only. VITROS Chemistry Products AMPH Reagent is used on VITROS 5,1 FS Chemistry Systems for the semi-quantitative or qualitative determination of amphetamines (AMPH) in human urine using a cutoff of either 500 or 1000 ng/mL. Measurements obtained with the VITROS AMPH method are used in the diagnosis and treatment of amphetamines use or overdose. The VITROS Chemistry Products AMPH assay is intended for use by professional laboratory personnel. It provides only a preliminary test result. A more specific alternative chemical method must be used to confirm a result obtained with this assay. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when evaluating a preliminary positive result. For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 26 is used to calibrate VITROS 5,1 FS Chemistry Systems for the qualitative or semi-quantitative measurement of drugs of abuse. For in vitro diagnostic use only. VITROS Chemistry Products FS Calibrator 1 is used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate VITROS 5,1 FS Chemistry Systems. For in vitro diagnostic use only. VITROS Chemistry Products DAT Performance Verifiers are assayed controls used to monitor performance of urine drugs of abuse screening assays on VITROS 5,1 FS Chemistry Systems.