# K062372: Freelite Human Kappa and Lambda Free Kits  for Use on the Spaplus Analyzer

- Pathway: 510(k) (Traditional)
- Applicant: The Binding Site, Ltd.
- Decided: Jun 21, 2007 after 311 days
- Product code: DFH (Kappa, Antigen, Antiserum, Control)
- Device class: 2
- Regulation: 21 CFR 866.5550
- Recalls of this device: 2 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
No later device cites it as a predicate yet.
0 of 1 device in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. FDA’s file for it is a 510(k) statement, which names no predicate.

## Recalls of this device
- Oct 28, 2016 · Class II · Z-1053-2017: A change in the calibration curve causing an increase in false prozone flags.
- Nov 28, 2012 · Class II · Z-1050-2017: Calibration curve activity has increased over time in the kit lots listed.

## Indications for use
This kit is intended for the quantitation of kappa free light chains in serum on The Binding Site SpaPlus. Measurement of the various amounts of the different types of light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenstrom's macroglobulinemia, amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus in conjunction with other laboratory and clinical findings. This kit is intended for the quantitation of lambda free light chains in serum on The Binding Site SpaPlus. Measurement of the various amounts of the different types of light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenstrom's macroglobulinemia, amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus in conjunction with other laboratory and clinical findings.