510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Lma Stonebreaker Device; Probe Size 1.0MM, 1.6MM, 2.0MM

K062475510(k) TraditionalCleared May 9, 2007258-day reviewClass II21 CFR 876.4480Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

The Laryngeal Mask Co., Ltd. · FFK Lithotriptor, Electro-Hydraulic

Cleared on May 9, 2007 as substantially equivalent to K951531 (Electro Medical Systems SA, 1995) and 2 other predicates after a 258-day review. Its family tree reaches back 3 generations to 1986. 1 later device descends from it. It has been recalled 7 times.

Decided
May 9, 2007
Review
258 days
Class
Class II
Product code
FFK
Recalls
7
Predicates
3

Its longest line of descent runs back 3 generations to K863642 (1986). 6 ancestors in all. 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1995: 1 other ancestor1992: 1 other ancestor1989: 1 other ancestor2013: 1 descendant clearedK863642 · Candela Model MDL-1 Laser Lithotriptor · Candela Corp. · 1986K963285 · Ems Swiss Lithoclast Lithotripter · Electro Medical Systems SA · 1997K974164 · Swiss Lithoclast Master Handpiece and 3.2 MM Probe · Medical Device Consultants, Inc. · 19971990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 2 other ancestors 1 generation forward: 1 descendant K974164 · Swiss Lithoclast Master Handpiece and 3.2 MM Probe · Medical Device Consultants, Inc. · 1997K963285 · Ems Swiss Lithoclast Lithotripter · Electro Medical Systems SA · 1997K863642 · Candela Model MDL-1 Laser Lithotriptor · Candela Corp. · 19860
  1. Root · 3 generations backK863642 Candela Model MDL-1 Laser LithotriptorCandela Corp. · 1986
  2. 2 generations backK963285 Ems Swiss Lithoclast LithotripterElectro Medical Systems SA · 1997
  3. 1 generation backK974164 Swiss Lithoclast Master Handpiece and 3.2 MM ProbeMedical Device Consultants, Inc. · 1997
  4. This deviceK062475 Lma Stonebreaker Device; Probe Size 1.0MM, 1.6MM, 2.0MMThe Laryngeal Mask Co., Ltd. · 2007
  5. Descendant · cites it directly · 0 of its ownK123038 Swiss LithobreakerE.M.S Electro Medical Systems S.A · 2013

Most influential descendants

Showing 1 of 1 · See all
  • K123038Swiss LithobreakerE.M.S Electro Medical Systems S.A · 2013 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K863642Candela Model MDL-1 Laser LithotriptorCandela Corp.1986—
−2K963285Ems Swiss Lithoclast LithotripterElectro Medical Systems SA1997—
−1K974164Swiss Lithoclast Master Handpiece and 3.2 MM ProbeMedical Device Consultants, Inc.1997—
0 (this device)K062475Lma Stonebreaker Device; Probe Size 1.0MM, 1.6MM, 2.0MMThe Laryngeal Mask Co., Ltd.2007—
+1K123038Swiss LithobreakerE.M.S Electro Medical Systems S.A2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199210
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
FIG. 1 — Lineage of K062475, 1986–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FFK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FFK device

  1. #1 K230488 EL27-Compact; Sterile EHL-Probesscore 12.2 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K951531Swiss Lithoclast Lithotripter
Summary page 2
…Predicate 1: Swiss LithoClast® (for fragmentation of ureteral and bladder stones) already marketed in the USA under K951531 Predicate 2: Swiss LithoClast® (for fragmentation Of ureteral, bladder and kidney stones…
Electro Medical Systems SA1995—
*K963285Ems Swiss Lithoclast Lithotripter
Summary page 2
…e 2: Swiss LithoClast® (for fragmentation Of ureteral, bladder and kidney stones) already marketed in the USA under K963285 Predicate 3: Swiss LithoClast®~Master already marketed in the USA under K974164 3.…
Electro Medical Systems SA1997—
*K974164Swiss Lithoclast Master Handpiece and 3.2 MM Probe
Summary page 2
…eted in the USA under K963285 Predicate 3: Swiss LithoClast®~Master already marketed in the USA under K974164 3. Device Description The LNA.A Stone~AreakerTMV isAporrtable nnn-eiectricAl, pneuimatfic contnc~t intracorpor…
Medical Device Consultants, Inc.1997—

Also cleared against K951531: 7 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K123038Swiss LithobreakerE.M.S Electro Medical Systems S.A20130

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 31, 2018Class IIZ-1593-2018Error in labelingThe interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
Jan 31, 2018Class IIZ-1592-2018Error in labelingThe interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
Jan 31, 2018Class IIZ-1594-2018Error in labelingThe interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
Jan 31, 2018Class IIZ-1595-2018Error in labelingThe interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
Jan 31, 2018Class IIZ-1596-2018Error in labelingThe interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
Jan 31, 2018Class IIZ-1590-2018Error in labelingThe interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
Jan 31, 2018Class IIZ-1591-2018Error in labelingThe interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 4 companies; a new generation was cleared every 9.4 years on median.
  • None of the 6 ancestors has a recall.
  • Of its descendants, K123038 (E.M.S Electro Medical Systems S.A, 2013) has gone furthest, with 0 descendants of its own.
  • 7 other devices also cite its predicate K951531.

How does its category compare?

All FFK devices

FFK Lithotriptor, Electro-Hydraulic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · FFK

2in 2025

2007: 22010: 22012: 22013: 32016: 22019: 12021: 12023: 12025: 2
2007high 3 (2013) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FFK

110in 2013

2013: 110 days median
2007high 110 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FFK

1in 2021

2007: 12009: 12017: 12021: 1
2007high 1 (2007) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FFK

6in 2025

2007: 62008: 122009: 352010: 162011: 62012: 22013: 102014: 92015: 72016: 72017: 92018: 182019: 102020: 392021: 222022: 132023: 102024: 82025: 6
2007high 39 (2020) · low 2 (2012)2025

MAUDE reports naming product code FFK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—16
20080—012
20090—135
20102—016
20110—06
20122—02
20133110010
20140—09
20150—07
20162—07
20170—19
20180—018
20191—010
20200—039
20211—122
20220—013
20231—010
20240—08
20252—06
20260—0—

What is it indicated for?

The LMA StoneBreakerTM is indicated for use via a rigid or semi-rigid endoscope with a straight working channel, for the fragmentation of urinary tract (i.e. kidney, ureter and bladder) stones.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K062475: Lma Stonebreaker Device; Probe Size 1.0MM, 1.6MM, 2.0MM.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K062475

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026