510(k) Report

510(k) · Cleared

Cleared

Bis Eeg Monitor View

K062613510(k) SpecialCleared Jun 18, 2007286-day reviewClass II21 CFR 882.1400

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Aspect Medical Systems, Inc. · OLW Index-Generating Electroencephalograph Software

Cleared on Jun 18, 2007 as substantially equivalent to K052362 (Aspect Medical Systems, Inc., 2005) after a 286-day review. Its family tree reaches back 7 generations to 1996. 9 later devices descend from it.

Decided
Jun 18, 2007
Review
286 days
Class
Class II
Product code
OLW
Recalls
None
Predicates
1

Its longest line of descent runs back 7 generations to K963644 (1996). 11 ancestors in all, the oldest K923043 (1992). 9 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 2 other ancestors1992: 1 other ancestor2000: 1 other ancestor2007: 1 descendant cleared2014: 1 descendant cleared2024: 1 descendant cleared2021: 1 descendant cleared2023: 1 descendant cleared2026: 2 descendants cleared2019: 1 descendant cleared2020: 1 descendant clearedK963644 · A-1000 Eeg Monitor and a-1050 Eeg Monitor · Aspect Medical Systems, Inc. · 1996K974496 · A-2000 Eeg Monitor with Bis · Aspect Medical Systems, Inc. · 1998K012466 · Bis/Eeg Module · General Electric Medical Systems Information Techn · 2001×3: 3 generations cleared close together, K031694 (2003), K030267 (2004), K040183 (2004) K052362 · Bis Eeg Monitor, Model a-3000 · Aspect Medical Systems, Inc. · 20051990200020102020
  1. Root · 7 generations backK963644 A-1000 Eeg Monitor and a-1050 Eeg MonitorAspect Medical Systems, Inc. · 1996
  2. 6 generations backK974496 A-2000 Eeg Monitor with BisAspect Medical Systems, Inc. · 1998
  3. 5 generations backK012466 Bis/Eeg ModuleGeneral Electric Medical Systems Information Techn · 2001
  4. 4 generations backK031694 Modification to Eeg Monitor with Bis, Model a-2000Aspect Medical Systems, Inc. · 2003
  5. 3 generations backK030267 Eeg Monitor with Bis, Model a-2000Aspect Medical Systems, Inc. · 2004
  6. 2 generations backK040183 BISXAspect Medical Systems, Inc. · 2004
  7. 1 generation backK052362 Bis Eeg Monitor, Model a-3000Aspect Medical Systems, Inc. · 2005
  8. This deviceK062613 Bis Eeg Monitor ViewAspect Medical Systems, Inc. · 2007
  9. Descendant · cites it directly · 8 of its ownK072286 Bis Eeg Vista Monitor System and BisxAspect Medical Systems, Inc. · 2007 · Recalled
  10. Descendant · 2 generations on · 2 of its ownK131898 A Series and Q Series Patient MonitorsGuangdong Biolight Meditech Co., Ltd. · 2014
  11. Descendant · 2 generations on · 1 of its ownK230693 BIS™ Advance Monitoring SystemCovidien, LLC · 2024
  12. Descendant · 3 generations on · 1 of its ownK183687 LifeWindow LW8 LiteDigicare Biomedical Technology, Inc. · 2019
  13. Descendant · 2 generations on · 0 of its ownK254213 Pascall Sensor Unit (PSU1)Pascall Systems, Inc. · 2026

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2007: 12014: 12019: 12020: 12021: 12023: 12024: 12026: 2
2007high 2 (2026) · low 0 (2008)2026

Most influential descendants

Showing 5 of 9 · See all
  • K072286Bis Eeg Vista Monitor System and BisxAspect Medical Systems, Inc. · 2007 · cites it directly8descendants
  • K131898A Series and Q Series Patient MonitorsGuangdong Biolight Meditech Co., Ltd. · 2014 · 2 generations on2descendants
  • K230693BIS™ Advance Monitoring SystemCovidien, LLC · 2024 · 2 generations on1descendant
  • K183687LifeWindow LW8 LiteDigicare Biomedical Technology, Inc. · 2019 · 3 generations on1descendant
  • K254213Pascall Sensor Unit (PSU1)Pascall Systems, Inc. · 2026 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K963644A-1000 Eeg Monitor and a-1050 Eeg MonitorAspect Medical Systems, Inc.1996—
−6K974496A-2000 Eeg Monitor with BisAspect Medical Systems, Inc.1998—
−5K012466Bis/Eeg ModuleGeneral Electric Medical Systems Information Techn2001—
−4K031694Modification to Eeg Monitor with Bis, Model a-2000Aspect Medical Systems, Inc.2003—
−3K030267Eeg Monitor with Bis, Model a-2000Aspect Medical Systems, Inc.2004—
−2K040183BISXAspect Medical Systems, Inc.2004—
−1K052362Bis Eeg Monitor, Model a-3000Aspect Medical Systems, Inc.2005—
0 (this device)K062613Bis Eeg Monitor ViewAspect Medical Systems, Inc.2007—
+1K072286Bis Eeg Vista Monitor System and BisxAspect Medical Systems, Inc.20071
+2K131898A Series and Q Series Patient MonitorsGuangdong Biolight Meditech Co., Ltd.2014—
+2K230693BIS™ Advance Monitoring SystemCovidien, LLC2024—
+3K183687LifeWindow LW8 LiteDigicare Biomedical Technology, Inc.2019—
+2K254213Pascall Sensor Unit (PSU1)Pascall Systems, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199210
200010
200120

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
201410
201910
202010
202110
202310
202410
202620
FIG. 1 — Lineage of K062613, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OLW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OLW device

  1. #1 K260880 SedLine Sedation Monitorscore 17.0 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052362Bis Eeg Monitor, Model a-3000
Summary page 1
…onitors have been classified by the Neurological Devices Panel as Class I devices (21 CFR 882.1400) Predicate Device Aspect Medical Systems A-3000 EEG Monitor with BIS (K052362) Device Description The BIS EEG Monitor, VIEW, is an EEG Monitor that displays EEG, as well as reports and graphs the BIS valuc…
Aspect Medical Systems, Inc.2005—

Also cleared against K052362: 1 other device →

What descends from it? (9)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K072286Bis Eeg Vista Monitor System and BisxAspect Medical Systems, Inc.20078

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 2 companies; a new generation was cleared every 1.5 years on median.
  • None of the 11 ancestors has a recall.
  • Of its descendants, K072286 (Aspect Medical Systems, Inc., 2007) has gone furthest, with 8 descendants of its own.
  • 1 other device also cites its predicate K052362.

How does its category compare?

All OLW devices

OLW Index-Generating Electroencephalograph Software: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · OLW

1in 2024

2007: 32008: 22015: 12018: 12021: 12022: 12023: 12024: 12026 (partial year): 2
2007high 3 (2007) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OLW

105in 2007

2007: 105 days median
2007high 105 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OLW

No recall names OLW since 2007.

Adverse event reports per year · OLW

73in 2025

2017: 32018: 32019: 62020: 152021: 152022: 272023: 182024: 932025: 73
2007high 93 (2024) · low 0 (2007)2025

MAUDE reports naming product code OLW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007310500
20082—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20151—00
20160—00
20170—03
20181—03
20190—06
20200—015
20211—015
20221—027
20231—018
20241—093
20250—073
20262—0—

What is it indicated for?

Intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The BIS Monitor is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents; and its usage with certain anesthetic agents…

Read the full statement (123 words)

Intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The BIS Monitor is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals.

The BIS may be used as an aid in monitoring the effects of certain anesthetic agents; and its usage with certain anesthetic agents may be associated with a reduction in primary anesthetic use and a reduction in emergence and recovery time.

Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of incidence of awareness with recall in adults during general anesthesia and sedation.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K062613: Bis Eeg Monitor View.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K062613

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026