510(k) · Cleared
Recalled · Class II · 2016Defogging Heated Endoscope Lens Protector (Dhelp); Defogging Endoscope Lens Protector (Delp)
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
New Wave Surgical Corporation · GCJ Laparoscope, General & Plastic Surgery
Cleared on Oct 5, 2006 as substantially equivalent to K993604 (Mectra Labs, Inc., 2000) and 9 other predicates after a 17-day review. Its family tree reaches back 2 generations to 1991. 23 later devices descend from it. It has been recalled 2 times.
- Decided
- Oct 5, 2006
- Review
- 17 days
- Class
- Class II
- Product code
- GCJ
- Recalls
- 2
- Predicates
- 10
Its longest line of descent runs back 2 generations to K904871 (1991). 15 ancestors in all. 23 later devices descend from it, 5 citing it directly.
- Root · 2 generations backK904871 Medi-KleerNorth American Marketing Corp. · 1991
- 1 generation backK974454 Clearfield Anti-Fog Sterile Wipe, Product Number 300-004Tucson Medical Corp. · 1997
- This deviceK062779 Defogging Heated Endoscope Lens Protector (Dhelp); Defogging…New Wave Surgical Corporation · 2006
- Descendant · cites it directly · 16 of its ownK080613 Clear-VU System, Models: CVL-01-1000, CVL-01-1030, CVL-01-1045Minimally Invasive Devices, LLC · 2008
- Descendant · 2 generations on · 8 of its ownK151117 Laparoscope Lens Shield DeviceMedeon Biodesign, Inc. · 2015
- Descendant · 3 generations on · 7 of its ownK160172 Laparoscope Lens Shield DeviceMedeon Biodesign, Inc. · 2016
- Descendant · 4 generations on · 6 of its ownK170103 Laparoscope Lens Shield Device (LENS)Medeon Biodesign, Inc. · 2017
- Descendant · 5 generations on · 4 of its ownK192891 Laparoscope Lens Shield Device (LENS)Medeon Biodesign, Inc. · 2019
Showing 5 of 23 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 23 · See all- K080613Clear-VU System, Models: CVL-01-1000, CVL-01-1030, CVL-01-1045Minimally Invasive Devices, LLC · 2008 · cites it directly16descendants
- K151117Laparoscope Lens Shield DeviceMedeon Biodesign, Inc. · 2015 · 2 generations on8descendants
- K160172Laparoscope Lens Shield DeviceMedeon Biodesign, Inc. · 2016 · 3 generations on7descendants
- K170103Laparoscope Lens Shield Device (LENS)Medeon Biodesign, Inc. · 2017 · 4 generations on6descendants
- K192891Laparoscope Lens Shield Device (LENS)Medeon Biodesign, Inc. · 2019 · 5 generations on4descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K904871 | Medi-Kleer | North American Marketing Corp. | 1991 | — |
| −1 | K974454 | Clearfield Anti-Fog Sterile Wipe, Product Number 300-004 | Tucson Medical Corp. | 1997 | — |
| 0 (this device) | K062779 | Defogging Heated Endoscope Lens Protector (Dhelp); Defogging Endoscope Lens… | New Wave Surgical Corporation | 2006 | — |
| +1 | K080613 | Clear-VU System, Models: CVL-01-1000, CVL-01-1030, CVL-01-1045 | Minimally Invasive Devices, LLC | 2008 | — |
| +2 | K151117 | Laparoscope Lens Shield Device | Medeon Biodesign, Inc. | 2015 | — |
| +3 | K160172 | Laparoscope Lens Shield Device | Medeon Biodesign, Inc. | 2016 | — |
| +4 | K170103 | Laparoscope Lens Shield Device (LENS) | Medeon Biodesign, Inc. | 2017 | — |
| +5 | K192891 | Laparoscope Lens Shield Device (LENS) | Medeon Biodesign, Inc. | 2019 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1993 | 1 | 0 |
| 1994 | 2 | 0 |
| 1996 | 1 | 0 |
| 1998 | 1 | 0 |
| 1999 | 2 | 0 |
| 2000 | 1 | 0 |
| 2002 | 1 | 0 |
| 2004 | 1 | 0 |
| 2005 | 1 | 0 |
| 2006 | 2 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2008 | 1 | 0 |
| 2010 | 1 | 0 |
| 2015 | 2 | 0 |
| 2016 | 2 | 0 |
| 2017 | 2 | 0 |
| 2018 | 2 | 0 |
| 2019 | 1 | 0 |
| 2020 | 2 | 0 |
| 2022 | 3 | 0 |
| 2023 | 1 | 0 |
| 2024 | 2 | 0 |
| 2025 | 1 | 0 |
| 2026 | 3 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in GCJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new GCJ device
- #1 K214046 780 nm 1688 4K Camera System, 780 nm L11 LED…score 21.4 · 21 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (10)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K993604 | Endoscopic Anti Fog DeviceSummary page 1…upplies and devices: Predicate devices using sterile antifog solutions in laparoscopic procedures: Endoscopic Anti Fog Device (Mister Clear) 510k Number K993604 Clear-It Anti-Fog Solution 510k Number K022826 ClearField Anti-Fog Sterile Wipe… | 2000 |
| * | K022826 | Clear-IT Anti-Fog, Model PS003Summary page 1…aroscopic procedures: Endoscopic Anti Fog Device (Mister Clear) 510k Number K993604 Clear-It Anti-Fog Solution 510k Number K022826 ClearField Anti-Fog Sterile Wipe 510k Number K974454 Predicate devices that defog the laparoscope by wa… | 2002 |
| * | K974454 | Clearfield Anti-Fog Sterile Wipe, Product Number 300-004Summary page 1…510k Number K993604 Clear-It Anti-Fog Solution 510k Number K022826 ClearField Anti-Fog Sterile Wipe 510k Number K974454 Predicate devices that defog the laparoscope by warming the scope in a heated solution: Applied Medical's Scope Warmer… | 1997 |
| * | K931895 | Endoscope AccessorySummary page 1…umber K974454 Predicate devices that defog the laparoscope by warming the scope in a heated solution: Applied Medical's Scope Warmer 510k Number K931895 Stryker's Scope Warmer 510k Number K053311 Deroyal Industries, Inc Defogger… | 1993 |
| * | K053311 | Stryker Scope WarmerSummary page 1…scope in a heated solution: Applied Medical's Scope Warmer 510k Number K931895 Stryker's Scope Warmer 510k Number K053311 Deroyal Industries, Inc Defogger 510k Number K982465 OR Solution Scope Holder… | 2006 |
| * | K982465 | Deroyal Industries, Inc. DefoggerSummary page 1…510k Number K931895 Stryker's Scope Warmer 510k Number K053311 Deroyal Industries, Inc Defogger 510k Number K982465 OR Solution Scope Holder 510k Number K051979 Predicate devices that are disposable and use batteri… | 1998 |
| * | K051979 | Ors Model 6000 Endoscope HolderSummary page 1…510k Number K053311 Deroyal Industries, Inc Defogger 510k Number K982465 OR Solution Scope Holder 510k Number K051979 Predicate devices that are disposable and use batteries to power an electrical resistance heating mechanism: Surgical Medical De… | 2005 |
| * | K023506 | Surgical Medical Device CauterySummary page 1…dicate devices that are disposable and use batteries to power an electrical resistance heating mechanism: Surgical Medical Device Cautery 510K Number K023506 Predicate devices that are battery operated and disposable: The following battery-operated device is used for laparoscopic surge… | 2004 |
| * | K052797 | Lapevac, Filtration Device for the PeriotoneumSummary page 2…device containing batteries that is disposed after every laparoscopic surgery. Buffalo Filter's LapEvac 51 Ok Number K052797 Circon Surgiflex Wave Suction-Irrigator System 51 Ok Number K992126 Comparison Table: Characteristic D.t.E.L.P… | 2006 |
| * | K992126 | Circon Surgiflex Wave Tsunami Suction-Irrigation SystemSummary page 2…= Page 2 === Buffalo Filter's LapEvac 51 Ok Number K052797 Circon Surgiflex Wave Suction-Irrigator System 51 Ok Number K992126 Comparison Table: Characteristic D.t.E.L.P Clear-It Anti- Deroyal OR Stryker's and Material… | 1999 |
What descends from it? (23)
See all 235 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 4, 2016 | Class II | Z-0889-2017 | Employee error | Product sterility is compromised due to breach of the sterile barrier |
| Oct 4, 2016 | Class II | Z-0890-2017 | Employee error | Product sterility is compromised due to breach of the sterile barrier |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 15 years and passes through 3 companies; a new generation was cleared every 8.8 years on median.
- 4 of the 15 ancestors have a recall of any class.
- Of its descendants, K080613 (Minimally Invasive Devices, LLC, 2008) has gone furthest, with 16 descendants of its own.
How does its category compare?
All GCJ devicesGCJ Laparoscope, General & Plastic Surgery: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · GCJ
21in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · GCJ
108in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · GCJ
5in 2025
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · GCJ
5,860in 2025
MAUDE reports naming product code GCJ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 18 | 42 | 12 | 1,639 |
| 2008 | 19 | 73 | 16 | 1,319 |
| 2009 | 21 | 85 | 6 | 1,995 |
| 2010 | 24 | 122 | 30 | 2,657 |
| 2011 | 15 | 113 | 9 | 2,133 |
| 2012 | 17 | 88 | 1 | 1,526 |
| 2013 | 15 | 93 | 7 (1) | 1,565 |
| 2014 | 19 | 90 | 2 | 1,714 |
| 2015 | 26 | 76 | 0 | 1,710 |
| 2016 | 27 | 99 | 5 | 2,080 |
| 2017 | 20 | 77 | 1 | 3,468 |
| 2018 | 21 | 88 | 1 | 2,562 |
| 2019 | 15 | 60 | 0 | 2,578 |
| 2020 | 28 | 61 | 0 | 2,618 |
| 2021 | 18 | 116 | 0 | 3,431 |
| 2022 | 20 | 103 | 1 | 3,945 |
| 2023 | 31 | 109 | 1 | 6,002 |
| 2024 | 29 | 90 | 0 | 6,047 |
| 2025 | 21 | 108 | 5 | 5,860 |
| 2026 | 26 | 122 | 1 | — |
What is it indicated for?
DHELP is intended to be used prior to and during endoscopic and laparoscopic procedures to prevent fogging of the scope lens.
DELP is intended to be used prior to and during endoscopic and laparoscopic procedures to prevent fogging of the scope lens.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K062779 · printed Nov 4, 2026
Cite this page
510(k) Report. “K062779: Defogging Heated Endoscope Lens Protector (Dhelp); Defogging Endoscope Lens Protector (Delp).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K062779
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026