510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Microscan Microstrep Plus Panel Ampicillin (0.013 - 16 MCG/ML)

K062927510(k) TraditionalCleared Oct 12, 200614-day reviewClass II21 CFR 866.1640Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Dade Behring, Inc. · LRG Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems

Cleared on Oct 12, 2006 after a 14-day review. The FDA summary names its predicate without a number. It has been recalled 2 times.

Decided
Oct 12, 2006
Review
14 days
Class
Class II
Product code
LRG
Recalls
2
Predicates
0

The FDA summary names its predicate without a number.

FIG. 1 — Lineage of K062927Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—MicroScan MICroSTREP plus Panel
Named without a submission number
—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 10, 2015Class IIZ-0108-2016Equipment maintenanceBeckman Coulter is recalling MICroSTREP plus 1 and MICroSTREP plus 2 panels because a manufacturing error may lead to antimicrobic degradation and the potential for elevated or false resistant results.
Sep 10, 2015Class IIZ-0109-2016Equipment maintenanceBeckman Coulter is recalling MICroSTREP plus 1 and MICroSTREP plus 2 panels because a manufacturing error may lead to antimicrobic degradation and the potential for elevated or false resistant results.

No ancestor or descendant has a Class I recall.

How does its category compare?

All LRG devices

LRG Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LRG

2in 2019

2007: 12008: 12010: 12011: 42019: 2
2007high 4 (2011) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LRG

144in 2011

2011: 144 days median
2007high 144 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LRG

1in 2020

2020: 1
2007high 1 (2020) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LRG

1in 2025

2008: 132009: 112010: 122011: 112012: 82013: 62014: 312015: 302016: 12017: 12019: 32020: 12024: 12025: 1
2007high 31 (2014) · low 0 (2007)2025

MAUDE reports naming product code LRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20081—013
20090—011
20101—012
20114144011
20120—08
20130—06
20140—031
20150—030
20160—01
20170—01
20180—00
20192—03
20200—11
20210—00
20220—00
20230—00
20240—01
20250—01
20260—0—

What is it indicated for?

The MicroScan MICroSTREP plus Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcuspneumoniae. After inoculation, panels are incubated for 20 - 24 hours at 35°C +/- i°C in a non-CO2 incubator, and read visually according to the Package Insert. Additionally, the panels may be incubated in and read by a MicroScan® WalkAway instrument.…

Read the full statement (112 words)

The MicroScan MICroSTREP plus Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcuspneumoniae. After inoculation, panels are incubated for 20 - 24 hours at 35°C +/- i°C in a non-CO2 incubator, and read visually according to the Package Insert. Additionally, the panels may be incubated in and read by a MicroScan® WalkAway instrument.

This particular submission is for the addition of instrument read capability of the antimicrobial Ampicillin, at concentrations of 0.03- 16 mcg/ml on the MicroScan MICroSTREP plus®Panel.

The organisms which may be used for Ampicillin susceptibility testing on this panel are:

Streptococcus spp. other than Streptococcus pneumoniae

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K062927: Microscan Microstrep Plus Panel Ampicillin (0.013 - 16 MCG/ML).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K062927

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026