# K063206: Dimension Vista Hcys Flex Reagent Cartridge; Protein 1 Calibrator; Protein 1 Control Low, Medium, High

- Pathway: 510(k) (Traditional)
- Applicant: Dade Behring, Inc.
- Decided: Dec 28, 2006 after 66 days
- Product code: LPS (Urinary Homocystine (Nonquantitative) Test System)
- Device class: 2
- Regulation: 21 CFR 862.1377
- Recalls of this device: 3 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 2 generations back to K002852 (2000). 6 ancestors in all.
1 later device descends from it; 1 cites it directly.
0 of 8 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K052788: N Latex Hcy, N Protein Standard SL, N/T Protein Control L/M/H, Dade Behring, Inc. (2006) Summary page 1.
- K012470: Modification to N Protein Standard SL, Dade Behring, Inc. (2001) Summary page 1.
- K012468: Modification to N/T Protein Control SL, Dade Behring, Inc. (2001) Summary page 1.

## Most influential descendants
- K062105: Diazyme Creatinine Liquid Reagents Assay (2007), 0 descendants of its own

## Recalls of this device
- May 17, 2016 · Class II · Z-2241-2016: Reagent lots # 15243MA and 15243MB shows a reduced stability once opened that does not meet the Instructions For Use claim of seven (7) days. A maximum bias of -37% was observed.
- Sep 18, 2013 · Class II · Z-0541-2014: Dimension Vista HCYS lot 13064MA shows erratic recovery of QC and patient samples.
- Aug 3, 2012 · Class III · Z-0253-2013: Cartridges are leaking. The leaking is observed when opening the wrapper that holds the reagent cartridge.

## Indications for use
The HCYS method is an in vitro diagnostic test for the quantitative determination of total homocysteine in human serum, heparinized and EDTA plasma on the Dimension Vista® System. Measurements of homocysteine aid in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria. PROT1 CAL is an in vitro diagnostic product for the calibration of the C3 Complement (C3), C4 Complement (C4), Homocysteine (HCYS), Immunoglobulin A (IGA), Immunoglobulin G (IGG), Immunoglobulin M(jGM) and Prealbumin / Transthyretin (PREALB) methods on the Dimension Vista System. PROT1 CON L, M and H are assayed intralaboratory quality controls for assessment of precision and analytical bias in the determination of C3 Complement (C3), C4 Complement (C4), Homocysteine (HCYS), Immunoglobulin A (IGA), Immunoglobulin G (IGG), Immunoglobulin M (IGM) and Prealbumin / Transthyretin (PREALB) on the Dimension Vista®.